(Circulation. 1999;100:387-392.)
© 1999 American Heart Association, Inc.
Clinical Investigation and Reports |
From the Krannert Institute of Cardiology, Indiana University, Indianapolis, Ind (W.J.G., S.A.B., E.D.E., D.P.Z.); Methodist Hospital, Indianapolis, Ind (P.R.F., B.J.C.); and the Southwest Florida Heart Group, Fort Myers, Fla (W.M.M., M.E.B.).
Correspondence to Douglas P. Zipes, MD, Krannert Institute of Cardiology, Indiana University School of Medicine, 1111 W 10th St, Indianapolis, IN 46202-4800. E-mail dzipes{at}iupui.edu
| Abstract |
|---|
|
|
|---|
Methods and ResultsWe evaluated the response in ICD function in 170 subjects during a 10- to 15-second midgate walk-through of and during extreme (2 minutes within 6 in of the gate) exposure to 3 common article surveillance systems. Complete testing was done in 169 subjects. During a 10- to 15-second (very slow) walk-through of the 3 surveillance systems, no interactions were observed that would negatively affect ICD function. During extreme exposure (169 subjects) and during extreme exposure and pacing via the ICD (126 subjects), interactions between the ICD and the article surveillance systems were observed in 19 subjects. In 7 subjects, this interaction was clinically relevant and would have likely (3 subjects) and possibly (4 subjects) resulted in ICD shocks. In 12 subjects, the interaction was minor.
ConclusionsIt is safe for a patient with an ICD to walk through electronic article surveillance systems. Lingering in a surveillance system may result in an inappropriate ICD shock.
Key Words: heart-assist device cardioversion defibrillation electromagnetic fields arrhythmia
| Introduction |
|---|
|
|
|---|
Electronic article surveillance systems are in common use in retail and public places; one estimate is that as many as 800 000 systems are in use worldwide. ICDs also are common, with estimates of 400 000 devices implanted worldwide. Exposures to article surveillance systems in those with ICDs occur and will continue to occur frequently; thus, an understanding of possible negative interactions is mandatory. We assessed the effects of electronic article surveillance systems on ICD function in 170 individuals.
| Methods |
|---|
|
|
|---|
The study design was approved by the Institutional Review boards at each center, and subjects gave informed consent before participation. Three electronic article surveillance systems similar to those used in antitheft applications were set up at each location. The 3 systems evaluated were manufactured by Sensormatic Electronics Corporation (Boca Raton, Fla): a pulsed acoustomagnetic system (Ultra-Max) and 2 different electromagnetic systems, Aislekeeper and P-Magnetic. These systems were chosen for 3 reasons: (1) case reports implicating electronic article surveillance equipment as a cause of inappropriate defibrillator shocks have occurred with 2 of these systems (2 with Ultra-Max and 1 with Aislekeeper); (2) other studies have shown interactions with pacemakers and these systems; and (3) these electronic article surveillance systems are commonly used worldwide.1 2 3 4 5 6 7 8
ICD sensing parameters for tachyarrhythmia detection were left as clinically programmed but with the actual delivery of therapies inactivated. Continuous ECG monitoring in the presence of a standby external defibrillator was used during the time period that therapies were inactivated. Exposure during 3 gate transits with each surveillance system was tested.
1. Routine exposure. Subjects were asked to walk through the middle of the gates over a 10- to 15-second period. This very slow transit was intended to replicate the exposure that the absolute slowest person might experience routinely.
2. Extreme exposure. Subjects stood in the surveillance system within 6 in of the gate transmitter for 2 minutes. Subjects slowly rotated near (or leaning on) the gate transmitter to assure exposure despite directional field effects.
3. Extreme exposure with pacing. We tested exposure during pacing because in many ICDs, the maximal sensitivity for detecting cardiac events (or noise) occurs during pacing. In addition, pacing at a fixed rate allows determination of whether any oversensing (of noise) by the ICD occurs, because such a response would delay pacing. Subjects who agreed to pacing for the time of the testing were paced at 20 bpm faster than their resting heart rates. For dual-chamber ICDs, pacing was done in both the atrium and ventricle.
A continuous ECG was recorded during gate transit. After each gate transit, the ICD was interrogated to determine whether any events were detected by the device or whether defibrillator reprogramming occurred.
In subjects who had a separate pacemaker, no changes were made in pacemaker programming. The pacemaker was interrogated before and after gate transits to ensure that no reprogramming occurred.
Data Analysis
Subject and ICD characteristics were recorded. The ICD
sensing circuit was categorized into 1 of 3 types: (1) endocardial tip
ring (true bipolar), (2) endocardial tip coil (integrated bipolar), and
(3) epicardial. Any interaction between the ICD function and the
specific surveillance systems was noted. An interaction that could
possibly result in an inappropriate defibrillator therapy was defined
as clinically relevant. Continuous variables are given as mean±SD.
Statistical tests included
2 analysis
with Pearson's formulation and Fisher's exact test. Student's
t test was used for continuous variables.
Multivariate analysis was conducted by means of
a Cox regression analysis. All tests were conducted with a
2-sided
risk level of 0.05.
| Results |
|---|
|
|
|---|
|
Routine Exposure
No alteration in baseline cardiac rhythm was observed with
gate transit. After a slow walk through the middle of each surveillance
system, interrogation of the ICDs revealed no interactions or any
reprogramming. Defibrillators manufactured by Ventritex (St Jude
Medical) do not provide information regarding
tachyarrhythmia detection unless therapies are
activated. In addition, Ventritex defibrillators manifested a
noise reversion mode in which noise sensing may appropriately
deactivate therapies temporarily. In newer Ventritex models,
(V-115, V-145, and V-180), an internal ECG recording was
activated to determine device sensing during transit. In these
15 defibrillators, no spurious sensing was detected. In older Ventritex
devices (V-100 and V-110), there was no capability to determine device
sensing except during pacing.
Extreme Exposure
In 3 subjects undergoing extreme exposure to the 3 surveillance
systems, the ICDs showed inappropriate tachyarrhythmia
detection (Table 2
). All occurred with
the Ultra-Max surveillance system, and all would likely have resulted
in shocks based on the ICD therapies. The first of these occurred in a
subject with a Medtronic model 7219 ICD. Sensing revealed continuous
noise with gate exposure interpreted by the ICD as
ventricular fibrillation (Figure 1
). Both the second and third
subjects had model 1746 defibrillators from Cardiac Pacemaker, Inc
(CPI). In both subjects, sensing revealed continuous noise with gate
exposure interpreted by the ICD as ventricular fibrillation
(Figure 2
). No ICD reprogramming
was observed in any of the ICDs.
|
|
|
Extreme Exposure With Pacing
The pacing effect on ICDarticle surveillance system
interaction was evaluated in 126 subjects. Five subjects were
characterized as dependent on their ICDs for pacing, and in 121
subjects, pacing at 20 bpm faster than resting heart rate was done.
Pacing resulted in oversensing of the T wave in 6 individuals with
Medtronic ICDs before surveillance system exposure. In these subjects,
ICD sensitivity was decreased by 0.15 mV sequentially until T-wave
oversensing was not observed, similar to what might be done during
normal clinical practice.
During extreme surveillance system exposure with pacing, ICDs in
19 of 126 subjects showed evidence of oversensing. No ICD reprogramming
was observed. The results showing ICD oversensing are summarized in
Table 2
. In 12 subjects, the interaction was not clinically
relevant. In these subjects, only an intermittent delay in pacing
caused primarily by noise-augmented T-wave oversensing was observed
(Figure 3
). In 7 subjects, the
interactions were clinically relevant. In 5 subjects, complete pacing
inhibition occurred with surveillance system exposure. Three of these
subjects had inappropriate ICD detection during extreme exposure.
Similar to the inappropriate ICD detections, pacing inhibition occurred
in the Ultra-Max surveillance system. The other 2 cases of complete
pacing inhibition occurred in subjects with Ventritex model V-110 ICDs.
In both subjects, pacing inhibition occurred with exposure to the
Aislekeeper surveillance system. In 2 additional subjects with
Ventritex model V-100 ICDs, prolonged pacing inhibition of
2
consecutive beats was observed with extreme exposure to the Aislekeeper
surveillance system. In 1 subject with a Ventritex model V-110 ICD in
whom complete pacing inhibition was observed on exposure to the
Aislekeeper surveillance system, pacing inhibition of 2 consecutive
beats also was observed during exposure to the Ultra-Max surveillance
system. Whether the pacing inhibition observed in these Ventritex
defibrillators would have led to inappropriate defibrillator therapies
or would have resulted in temporary noise reversion is unknown.
|
In 4 of the 5 subjects characterized as dependent on their ICDs for pacing, no changes in pacing were observed with extreme gate exposure. One subject who had inappropriate tachyarrhythmia detection and complete pacing inhibition on extreme exposure to the Ultra-Max surveillance system remained asymptomatic, with an escape heart rate in response to underlying atrial fibrillation of 40 bpm.
Evaluation in Subjects With Separate Pacemakers
Four subjects had separate pacemakers. All of these
individuals had pacing in the atrium, ventricle, or both. No
abnormalities in pacemaker function or abnormal pacemaker-ICD
interactions were observed with any gate exposure. Interrogation of the
pacemakers after gate transit showed no reprogramming.
Correlates of DefibrillatorSurveillance System
Interference
Subject and ICD characteristics were analyzed to
determine factors increasing the likelihood of surveillance
systemdefibrillator interaction. Characteristics found to predict an
interaction through univariate analysis are
summarized in Tables 3
and 4
. With multivariate
analysis, a diminished R-wave amplitude and a Ventritex ICD
were the only characteristics that independently predicted an increased
likelihood of any interaction (R wave, P=0.01; Ventritex,
P<0.01) or a clinically relevant interaction (R wave,
P=0.05; Ventritex, P=0.02).
|
|
| Discussion |
|---|
|
|
|---|
Previous Reports
Three case reports have described defibrillator shocks related to
surveillance system exposure.1 2 3 As in our study, the
shocks in these 3 patients occurred with close or protracted exposure
to the surveillance system, with abdominal implants (2 of 3), and with
exposure to a pulsed acoustomagnetic (Ultra-Max) or an electromagnetic
(Aislekeeper) surveillance system. A single study has evaluated
interactions in patients with ICDs and surveillance
systems.8 In this study, no interference was found in 25
subjects despite close and protracted exposure. No pacing was done in
this study; therefore, minor interference might not have been
recognized.
Study Limitations
Our study evaluated only 1 manufacturer's surveillance systems.
It is likely that other systems using similar electromagnetic fields
would interact similarly.8 No testing of
swept-radiofrequency surveillance systems was done. A single previous
study evaluating pacemakers and article surveillance systems has not
shown any interference with the swept-radiofrequency
systems.8 Because of the presence of a noise reversion
mode in Ventritex ICDs, what was considered a clinically relevant
interaction might only result in pacing output inhibition and not lead
to an inappropriate ICD shock. ICDs manufactured by CPI also have a
noise-sensing window that could influence electromagnetic
interference.
Conclusions and Recommendations
On the basis of our study, it does not appear that
electronic article surveillance systems pose a threat to patients with
ICDs if exposure is kept to a 10- to 15-second (ie, very slow)
walk-through. More intense exposure in both time and proximity may lead
to inappropriate shocks. The already minimal risk of an interaction
should diminish even more as older and abdominal ICDs are replaced with
newer pectoral defibrillators.
The Food and Drug Administration (FDA Safety Notification, September 28, 1998) and the American Heart Association (AHA Science Advisory, November 17, 1998) have reviewed the available data and issued statements agreeing that significant interactions are unlikely and that the public should be informed but not alarmed. Our study supports this message. What would seem most prudent is to ensure that physicians who manage patients with ICDs (and pacemakers) understand and inform their patients of the potential interactions associated with electronic article surveillance systems. Recognition of surveillance systems by the public should be possible to avoid protracted or close exposure. Surveillance systems that are not in plain view of the public should have signage indicating their presence. This policy should allow patients with ICDs (or pacemakers) to walk through such systems without lingering and therefore be seemingly at no risk of any interaction.
| Acknowledgments |
|---|
Received March 5, 1999; revision received April 21, 1999; accepted April 30, 1999.
| References |
|---|
|
|
|---|
2. McIvor ME. Environmental electromagnetic interference from electronic article surveillance devices: interactions with an ICD. Pacing Clin Electrophysiol. 1995;18:22292230.[Medline] [Order article via Infotrieve]
3.
Santucci PA, Haw J, Trohman RG, Pinski SL.
Interference with an implantable defibrillator by an electronic
antitheft surveillance device. N Engl J Med. 1998;339:13711374.
4. Beaugeard D, Kacet S, Bricout M, Camblin J. Interference between cardiac pacemaker and electromagnetic anti-theft devices in stores. Arch Mal Coeur Vaiss. 1992;85:14571461.[Medline] [Order article via Infotrieve]
5. Dodinot B, Godenir JP, Costa AB. Electronic article surveillance: a possible danger for pacemaker patients. Pacing Clin Electrophysiol. 1993;16:4653.[Medline] [Order article via Infotrieve]
6. Lucas EH, Johnson D, McElroy BP. The effects of electronic article surveillance systems on permanent cardiac pacemakers: an in vitro study. Pacing Clin Electrophysiol. 1994;17:20212026.[Medline] [Order article via Infotrieve]
7. Mugica J, Henry L. Interferences entre les stimulateurs cardiaques et les systems de detections antival. Stimucoeur. 1997;25:287288.
8. McIvor ME, Reddinger J, Floden E, Sheppard RC. Study of Pacemaker and Implantable Cardioverter Defibrillator Triggering by Electronic Article Surveillance Devices (SPICED TEAS). Pacing Clin Electrophysiol. 1998;21:18471861.This study evaluated the risk to patients with implantable cardioverter-defibrillators (ICDs) encountered during exposure to electronic article surveillance systems. In 169 subjects with ICDs, no evidence of any interaction was observed during a slow walk-through of 3 common article surveillance systems. With close, protracted exposure to the surveillance systems, an interaction was observed in 19 subjects: 12 were minor and 7 clinically relevant. We conclude that individuals with ICDs can safely pass through electronic article surveillance systems if they do not linger.[Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
J. R. Gimbel and J. W. Cox Jr Electronic Article Surveillance Systems and Interactions With Implantable Cardiac Devices: Risk of Adverse Interactions in Public and Commercial Spaces Mayo Clin. Proc., March 1, 2007; 82(3): 318 - 322. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. H. Schoenfeld Contemporary Pacemaker and Defibrillator Device Therapy: Challenges Confronting the General Cardiologist Circulation, February 6, 2007; 115(5): 638 - 653. [Full Text] [PDF] |
||||
![]() |
W. G. Stevenson, B. R. Chaitman, K. A. Ellenbogen, A. E. Epstein, W. L. Gross, D. L. Hayes, S. A. Strickberger, M. O. Sweeney, and for the Subcommittee on Electrocardiography and Ar Clinical Assessment and Management of Patients With Implanted Cardioverter-Defibrillators Presenting to Nonelectrophysiologists Circulation, December 21, 2004; 110(25): 3866 - 3869. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. L. Tri, J. M. Trusty, and D. L. Hayes Potential for Personal Digital Assistant Interference With Implantable Cardiac Devices Mayo Clin. Proc., December 1, 2004; 79(12): 1527 - 1530. [Abstract] [PDF] |
||||
![]() |
C. Kolb, S. Schmieder, G.u. Lehmann, B. Zrenner, M. R. Karch, A. Plewan, and C. Schmitt Do airport metal detectors interfere with implantable pacemakers or cardioverter-defibrillators? J. Am. Coll. Cardiol., June 4, 2003; 41(11): 2054 - 2059. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. NIEHAUS and J. TEBBENJOHANNS Electromagnetic interference in patients with implanted pacemakers or cardioverter-defibrillators Heart, September 1, 2001; 86(3): 246 - 248. [Full Text] [PDF] |
||||
![]() |
M. Santini Digital cellular telephones and ICDs Eur. Heart J., August 1, 2001; 22(15): 1251 - 1252. [Abstract] [PDF] |
||||
![]() |
J.A Chiladakis, P Davlouros, G Agelopoulos, and A.S Manolis In-vivo testing of digital cellular telephones in patients with implantable cardioverter-defibrillators Eur. Heart J., August 1, 2001; 22(15): 1337 - 1342. [Abstract] [PDF] |
||||
![]() |
N. Goldschlager, A. Epstein, P. Friedman, E. Gang, R. Krol, B. Olshansky, and for the North American Society of Pacing and Elect Environmental and Drug Effects on Patients With Pacemakers and Implantable Cardioverter/Defibrillators: A Practical Guide to Patient Treatment Arch Intern Med, March 12, 2001; 161(5): 649 - 655. [Full Text] [PDF] |
||||
![]() |
Walking Through an Electronic Article Surveillance System Is OK--but Don't Linger Journal Watch Cardiology, September 16, 1999; 1999(916): 11 - 11. [Full Text] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1999 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |