Circulation. 1999;100:2284-2301

Free Article
This Article
Free upon publication Free Article
Right arrow Extract
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Yucel, E. K.
Right arrow Articles by Pearce, W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Yucel, E. K.
Right arrow Articles by Pearce, W.
Related Collections
Right arrow AHA Statements and Guidelines
Right arrow Coronary imaging: angiography/ultrasound/Doppler/CC
Right arrow CT and MRI

(Circulation. 1999;100:2284.)
© 1999 American Heart Association, Inc.


AHA Scientific Statement

Magnetic Resonance Angiography

Update on Applications for Extracranial Arteries

E. Kent Yucel, MD, Chair; Charles M. Anderson, MD, PhD; Robert R. Edelman, MD; Thomas M. Grist, MD; Richard A. Baum, MD; Warren J. Manning, MD; Antonio Culebras, MD, Ex Officio; William Pearce, MD, Ex Officio


Key Words: AHA Scientific Statement • angiography • magnetic resonance imaging • imaging • arteries

Magnetic resonance angiography (MRA) has excited the interest of many physicians working in cardiovascular disease because of its ability to noninvasively visualize vascular disease. Its potential to replace conventional x-ray angiography (CA) that uses iodinated contrast has been recognized for many years, and this interest has been stimulated by the current emphasis on cost containment, outpatient evaluation, and minimally invasive diagnosis and therapy. In addition, recent advances in magnetic resonance (MR) technology resulting from fast gradients and use of contrast agents have allowed MRA to make substantial advances in many arterial beds of clinical interest. The goal of this scientific statement is to present the current state of MRA of the extracranial arteries and to suggest current as well as possible future clinical applications for MRA. For the purposes of this statement, MRA is defined as MR techniques that provide cross-sectional or projectional images of normal and diseased arterial anatomy. It does not deal with the equally important area of quantitative flow measurement with MR. The first section deals with the technical basis of MRA. Subsequent sections deal with individual vascular beds in which MRA has shown clinical promise.

MRA: Technical Considerations

The "gold standard" for many manifestations of vascular disease, especially arterial occlusive disease, is CA, an invasive, costly, and potentially hazardous procedure. MRA could represent an alternative, noninvasive approach. Rather than a single technique, MRA actually represents a family of different techniques. As outlined below, contrast between blood and soft tissues is derived from completely different MR mechanisms in the various MR techniques. We will consider the basic principles underlying the MR imaging (MRI) appearance of flowing blood and the techniques used to image blood flow and render angiogram-like MRI scans.

Depending on the imaging technique used, blood may appear bright or dark. On traditional spin-echo MR images, blood vessels usually appear dark. With spin-echo pulse sequences, a pair (90o and 180o) of section-selective radiofrequency (RF) pulses is used to produce an MR signal. If blood flows out of the plane of the section in the time interval between successive RF pulses, the result is an absence of signal, called a flow void.1 The flow void can be emphasized by use of a thin section or a long echo time. In a fast spin-echo sequence, a long train of echoes is acquired by use of a series of 180o RF pulses; as a result, washout effects are even more substantial than with conventional spin-echo techniques. Other methods for creating a flow void include presaturation, which involves application of an additional RF pulse outside the plane of the section to suppress the signal intensity of inflowing blood; dephasing gradients; and preinversion pulses to nullify the blood signal.2

To create bright blood, gradient-echo pulse sequences are used. In a gradient-echo sequence, only a single RF pulse is applied during each sequence repetition, so no signal is lost because of washout effects. Data for images on which blood is bright can be acquired as a series of overlapping thin sections (sequential 2-dimensional [2D]) or as 1 or more thick (3-dimensional [3D]) volumes. Each sequence has advantages, as discussed below. Bright-blood techniques can be subcategorized into time-of-flight (TOF)3 and phase-contrast techniques.4 The basis of TOF techniques is that positive flow contrast is generated by inflow effects, whereas the background (stationary tissue) is saturated by the rapid repeated application of RF pulses. Saturation pulses based on flow geometry can be used to eliminate unwanted vessels (eg, leg veins for lower-extremity arteriography). Use of a segmented gradient-echo sequence with cardiac triggering is helpful to eliminate arterial pulsation artifacts.5 The basis for phase-contrast MRA is that the flow of blood along a magnetic field gradient causes a shift in the phase of the MR signal. With phase contrast, pairs of images are acquired that have different sensitivities to flow. These are then subtracted to cancel background signal, leaving only signal from flowing blood. Phase contrast also permits flow (velocity) quantification,6 because the phase shift is proportional to the velocity.

The core of MRA is the ability to portray blood vessels in a projective format similar to CA. Currently, projection images are created by postprocessing of images acquired by a 2D or 3D gradient-echo sequence. Although image processing can be postponed until after the patient has left the MR suite, it is best done while the patient is still within the magnet so that additional scans can be obtained if needed. Most commonly, the images are processed by use of a maximum-intensity projection (MIP) algorithm.7 8 With an MIP algorithm, the brightest pixels along a user-defined direction are extracted to create a projection image. Areas with poor flow contrast, including the edges of blood vessels and small vessels with slow flow, may be obscured by overlap with brighter stationary tissue.9 Through reduction of the pixel size and suppression of the signal of stationary tissues, the quality of the MIP can be improved substantially. The reduction in pixel size is accomplished by use of a large (eg, 256x512) acquisition matrix along with a small field of view.

A variety of artifacts caused by phase or magnitude variations in the MR signal afflict MRA. Within a single image volume element (voxel), blood protons flowing at different velocities accumulate a range or dispersion of phase shifts. Complex flow can produce signal loss due to intraview phase dispersion (ie, it occurs during each repetition of the pulse sequence), ghost artifacts from view-to-view signal variations (ie, those that occur over multiple sequence repetitions), and flow-displacement errors relating to the time delay between RF excitation and frequency encoding or between phase and frequency encoding. These effects tend to falsely exaggerate the severity of a stenosis and are most pronounced with 2D MRA. They can also limit flow-based MRA imaging in areas of slow flow, such as aneurysms. A short echo time (TE) minimizes flow-displacement and phase-dispersion artifacts; phase dispersion is further decreased when the voxel size is minimized (eg, by use of thin sections). Small voxels and short TEs are most easily obtained with 3D TOF methods. The biggest drawback of the thick volumes used with 3D is that slow or recirculating flow can become saturated. The advantages of both 2D and 3D techniques are gained with a series of thin-slab 3D acquisitions. The sequential 3D (or multiple overlapping thin-slab acquisition [MOTSA]) technique gives better flow enhancement than single-slab 3D techniques and less dephasing than 2D techniques.10 The method has some drawbacks as well. For instance, the nonrectangular profile of the 3D slabs necessitates the use of substantial overlap (up to 50%) of adjacent slabs, so that total scan time is increased compared with single-slab 3D. Moreover, signal-intensity variations within the individual slabs due to saturation effects cause an annoying "venetian blind" artifact. With sequential 2D or 3D acquisitions, slight patient motion can generate discontinuities in the vessel contour that can be mistaken for focal stenosis or fibromuscular dysplasia.

Recently, an increasingly preferred method for chest and abdominal MRA is the combination of breath-hold 3D gradient-echo sequences with short repetition time (TR) and short TE during administration of a gadolinium chelate, often in a double dose.11 12 The contrast agent shortens the T1, or longitudinal relaxation time, of blood (to as low as 50 ms, compared with normal blood T1 of {approx}1200 ms), so that the blood appears bright irrespective of flow patterns or velocities. Correct timing of the injection is important to ensure synchronization between the transit of contrast material and scanning. Methods that can be used to achieve correct timing include empirical estimation of transit time, a small test bolus of contrast agent to determine the time delay between injection and arrival of the contrast agent bolus in the target vessel,13 automated detection of contrast bolus passage,14 and MR fluoroscopy to observe contrast passage.15

Contrast-enhanced MRA (CEMRA) is advantageous in displaying detailed vessel anatomy and in reducing artifacts. MIP images can be reliably rendered in multiple projections. If the patient is incapable of holding his or her breath, then a slower acquisition can be done over several minutes with a longer/slower infusion of contrast agent. However, branch vessel detail may be degraded by respiratory motion. Subtraction of a precontrast from a postcontrast scan eliminates the signal from background tissue (assuming no patient motion). Additionally, it is possible to perform time-resolved 3D MRA studies to better differentiate arteries and veins.16 Recently, testing has begun in humans of intravascular MR contrast agents that are retained within blood vessels and that selectively enhance the blood pool on T1-weighted MR images.17 18 19 20 21 AMI-227 (Advanced Magnetics), an ultrasmall iron particle, has been shown to enhance vessels within the abdomen and chest in the steady state.17 18 Another iron particle, NC100150 (Nycomed-Amersham), has recently entered human trials.19 MS-325 (EPIX Medical, Inc and Mallinckrodt Inc), a gadolinium-chelate that binds noncovalently to plasma albumin, has been shown to enhance the lower extremity and carotid vessels in the steady state as well as to enhance the arteries of the lower extremity during the first pass in a manner comparable to the currently available extracellular gadolinium chelates.22 The potential clinical applications of these agents await the results of large-scale multicenter clinical trials.

Coronary MRA is perhaps the most technically challenging area of MRA. To apply MRA techniques to the coronary arteries, additional technical obstacles had to be overcome, including compensation for respiratory and cardiac motion. Standard electrocardiographically (ECG) gated spin-echo and gradient-echo cine images only occasionally show portions of the coronary arteries, and these images are not adequate for detailed evaluation.23 Even a 1-second acquisition is too long to freeze cardiac motion. Instead, the acquisition can be "segmented" into blocks of phase-encoding steps, which are then interleaved to create an image.24 Segmentation within a single breath-hold of 10 to 20 seconds can reduce the time for data acquisition within each cardiac cycle to 100 ms or less, which is adequate to minimize cardiac motion artifacts, particularly in mid diastole or end systole. This strategy is known as segmented gradient-echo. Because the proximal coronary arteries are embedded in epicardial fat, chemical shift-selective fat saturation pulses are applied to reduce the fat signal. With this combination of techniques, the proximal portions of the left main, left anterior descending (LAD), and right coronary arteries (RCA) are routinely seen.

Recently, navigator echo methods have been applied to eliminate the need for breath-holding. Navigator echo is a motion-compensation technique that relies on tracking of an MR-detectable interface by the MR scanner. In the case of coronary MRA, the diaphragm-lung interface is typically used. With navigator gating for coronary MRA, MRI data are only accepted for image reconstruction when the navigatorecho indicates that the diaphragm or lung is within a certain operator-defined range (usually 3 to 5 mm). In this respect, navigator echoes perform a function similar to the bellows used for respiratory gating but provide more consistent results. Because many data are rejected, acquisition times are typically increased by a factor of 2 or more, depending on the patient’s breathing patterns. However, if a fast imaging sequence such as segmented gradient echo is used, scan times are still reasonable, on the order of a few minutes at most. Image sharpness is as good as with breath-holding.25 Because navigator echoes permit signal averaging, the signal-to-noise ratio is increased, and higher in-plane spatial resolution can be obtained (eg, 0.5 mm). Moreover, navigator echoes ensure a consistent cardiac position from image to image, thereby minimizing misregistration artifact if one wishes to process a projection image.

The navigator gating technique discards data that are acquired when the diaphragm is incorrectly positioned and thus increases scan time. Adaptive correction of image location by use of real-time navigator measurement of diaphragm position is a potential method for further reducing slice registration errors.26 27 This method dynamically repositions the slice to follow the movement of the diaphragm and thus keeps the slice in a constant relation to the tissue of interest despite its motion. This technique, called navigator correction, can be combined with navigator gating to improve the efficiency of data acquisition. Navigator methods appear promising for both 2D and 3D acquisitions. Nevertheless, achievement of adequate spatial resolution, especially of small vessels, remains a challenge for MRA. Resolution below 0.5 mm may be required to achieve comparability with CA, especially in smaller (<3 mm) vessels. Motion makes this goal even more difficult to achieve, both by lengthening the scan time to compensate for respiratory motion (through navigators) and by narrowing the scanning window to a portion of the cardiac cycle. In the heart, beat-to-beat variations in cardiac position may exceed 1 mm, making submillimeter resolution even more challenging.

MRA of the Carotid Artery

MRA of the carotid bifurcation may be accomplished by several different MRI techniques, including 2D TOF,28 3D TOF,29 and CEMRA.30 When the merits of these sequences are compared, several generalizations can be made. 2D TOF provides a strong vascular signal, even when the arterial velocity is low. 2D TOF should be used to differentiate near and complete internal carotid artery (ICA) occlusion. 3D TOF provides superior, submillimeter resolution but at the expense of sensitivity to flow. The weak vascular signal of 3D TOF in slow-flow situations may be improved by the use of MOTSA.31 The original slices from a 3D TOF acquisition may allow one to see some features of plaque directly. CEMRA is quick and robust and is not impaired by slow-flow situations. CEMRA is a good choice for those patients who cannot maintain a position for prolonged periods. Unfortunately, it can only be effectively performed on MRI systems with enhanced gradient hardware.

There is currently no agreement as to which of these techniques is best. In part, this results from the fact that MRI instruments of different age and from different vendors have slightly different imaging characteristics. However, most experts use the MOTSA sequence for their primary interpretation. CEMRA is a very promising technique, but its diagnostic role awaits further validation. The aortic arch is best imaged by CEMRA, which has proven effective for the study of stenoses, aneurysms, and dissections.32 33 The carotid siphons and intracranial arteries are best imaged with MOTSA.

As a result of time constraints, only the cervical carotid artery is studied usually. A full study of the brain, intracranial vessels, bifurcation, and aortic arch requires one to position the patient 3 times (in the head, neck, and body coils) and necessitates that the patient spend more than 1 hour in the magnet. Such a comprehensive examination should be reserved for those with suspected tandem pathology on the basis of a prior duplex ultrasound (DUS) or MRA examination.

A typical study might begin with a low-resolution 2D TOF to locate the bifurcation. Sections should be acquired as high as the base of the skull so that patency of the ICA can be positively established in the carotid canal of the temporal bone. Lack of flow at this level implies an occlusion at the bifurcation, because the ICA lacks branches between these points. If a vertebral artery or other vessel is absent on the 2D images, the sequence should be repeated without presaturation bands to exclude reversal of flow. Then, a MOTSA acquisition can be acquired of the bifurcation or at any other level identified as stenotic on the 2D series. A MOTSA acquisition of the bifurcation takes {approx}10 minutes, whereas 2D TOF of the entire length of the carotid artery takes {approx}6 minutes. Technical factors that may affect MRA are excessive patient motion and the presence of ferromagnetic metal in the neck.

When carotid MRA is interpreted, special care should be taken to avoid overestimation of the stenosis severity in an area of turbulent flow.34 This phenomenon is encountered within the decelerating jet distal to a critical lesion. The tendency to overestimate stenosis severity is greatly reduced if one interprets the study from source images or reformations rather than from calculated projections.35 36 37 Overestimation is also reduced in 3D acquisitions, presumably because the more gradient structure and submillimeter voxels of 3D acquisition result in less phase dispersion, although many practitioners of 2D TOF have shown an excellent ability to properly grade severe lesions. Finally, the tendency to overestimate stenoses is reduced if one quantitatively measures the vessel diameters formally with a calibrated jeweler’s loupe rather than by "eyeball estimation."

Table 1Down summarizes many of the recent prospective comparisons of MRA and CA for the evaluation of the carotid bifurcation.37 38 39 40 41 42 43 44 45 46 47 48 49 The median sensitivity for a high-grade lesion was 93%, whereas the median specificity was 88%. These studies assumed CA to be the gold standard. A problem with this assumption is that many CA errors are considered errors in MRA. Because the reproducibility of CA itself is no better than 94%,48 50 51 one can conclude that the actual sensitivities and specificities may be better than those reported in these studies. In fact, preliminary data suggest that noninvasive imaging may be more sensitive than CA in some instances.40 A comparison of CA, MRA, and DUS, with surgical specimens rather than CA used as the gold standard,52 found that DUS and MRA each correlate better with the endarterectomy specimen than does CA. This discrepancy can be attributed to the fact that the smallest diameter is often not appreciated by CA when the stenosis is elliptical or complex in shape. When CA and DUS are compared, it is particularly important to remember that the stenotic lumen is usually not circular, because Doppler velocities are determined by cross-sectional area rather than by diameter.53 Potentially important information about the shape of the plaque, available by MRA but not by CA, was not considered in the validation studies. The role of CA as a modality for assessing carotid stenosis has been established by its use in the large, multicenter clinical trials that have established endarterectomy as a successful stroke-prevention intervention. Multicenter clinical trials like the North American Symptomatic Carotid Endarterectomy Trial (NASCET) have never been undertaken with MRA rather than CA used to quantify stenosis. Therefore, MRA and DUS are assumed to be effective only to the degree that they agree with CA diameter measurements.


View this table:
[in this window]
[in a new window]
 
Table 1. Summary of Carotid MRA Accuracy Studies

Kent et al45 examined the cost-effectiveness of various imaging strategies for a population of patients with symptomatic carotid artery disease. Use of DUS alone resulted in a quality-adjusted life expectancy (QALE) of 9.619 years. CA alone resulted in a QALE of 9.632 years, with an excessive incremental cost-effectiveness ratio of $99 200 per quality-adjusted life year (QALY). The combination of DUS and MRA, followed by CA in the event of disparate results, maximized clinical outcome to 9.639 years at an incremental cost-effectiveness ratio of just $22 400 and was considered the optimum strategy. Obuchowski et al54 assumed a 20% prevalence of stenosis requiring surgery in patients who presented with a neck bruit. Three imaging strategies were examined: DUS followed in selected cases by MRA, DUS followed in selected cases by CA, and MRA alone. The QALE of the 3 strategies was virtually identical, whereas the incremental cost per QALY was $2922, $7470, and $7700, respectively. DUS alone was not examined. These data again argue for the use of a combined noninvasive approach.55 When asymptomatic patients are being screened, the advantages of noninvasive imaging are even greater. Kuntz et al56 concluded that use of CA in this population actually resulted in a greater incidence of stroke (7.12% in 5 years) than did DUS alone (6.35%), MRA alone (6.17%), or a combination of DUS and MRA followed by CA if necessary (6.34%).

Currently, it appears that the workup of suspected carotid artery stenosis should begin with DUS, performed by an experienced and preferably accredited laboratory. Although it has been suggested that the detection of >70% stenosis on DUS alone might be sufficient to recommend carotid endarterectomy, it may be appropriate to confirm this diagnosis with an additional imaging study to minimize the possibility of erroneous management. If so, a technically adequate MRA at an imaging facility with substantial experience in the performance and interpretation of carotid MRA can fulfill this role at the lowest cost and risk to the patient.57 58 CA should be considered when the results of DUS and MRA are discrepant, in cases of possible hairline patency (angiographic string sign), or where lesions are so atypical that they can only be understood by use of the high-resolution, selective vascular contrast of a catheter study.59 Supplementary imaging is especially advisable when results of DUS are technically limited. Typical limitations that necessitate additional imaging include the presence of a shadowing plaque, a deep course of the ICA, discordant gray-scale and Doppler measurements, and evidence of tandem lesions in cases in which these would alter patient management. Tandem lesions are suggested by the presence of a dampened and delayed waveform60 or unusually low diastolic velocities. To detect disease outside the ultrasonic window, MRA should also be considered in patients with atypical symptoms of cerebral ischemia and a DUS showing no disease or only mild disease. In some situations, the evaluation might begin with MRA (for example, when the MRA is performed as part of a brain MRI in patients with neurological symptoms).

MRA of the Thoracic and Abdominal Aorta

Standard T1-weighted MR techniques provide detailed anatomic information about the dimensions of the thoracic and abdominal aorta in multiple planes, can identify the presence of intraluminal thrombus within aneurysms or intimal flaps associated with aortic dissection, and can locate the main aortic branches, such as the iliac, main renal, and brachiocephalic arteries, in relation to aortic pathology. However, for projectional angiographic imaging and for detection of branch vessels and branch vessel disease, MRA is essential.

Abdominal aortic aneurysms (AAAs) have been an important focus of studies with MRA because of the desirability of replacing CA for preoperative evaluation. AAAs are typically diagnosed by physical examination, ultrasound, or computed tomography (CT), but angiographic imaging is often desired before preoperative evaluation. The important information that angiography must provide includes the upper extent of the aneurysm (including the relationship of the aneurysm to the renal arteries), the number and patency of the renal arteries, the patency of the mesenteric arteries, the inferior extent of the aneurysm (including the distance from the aneurysm to the aortic bifurcation), and aneurysmal or atherosclerotic disease of the iliac arteries. CA has been the standard approach for such preoperative evaluation. However, several studies of MRA for preoperative evaluation of AAAs have been published. Initial studies used noncontrast TOF MRA, often in conjunction with T1-weighted MRI. Although classification of aneurysms as suprarenal or infrarenal was good, detection of accessory renal arteries, renal artery stenosis, and iliac vascular disease was not adequate to justify replacement of CA. Limitations included the poor resolution in the slice-selection direction obtainable with 2D techniques and poor signal associated with slow flow in the ectatic aorta.61 62 63

More recently, interest has focused on CEMRA for AAA evaluation.64 65 66 67 The use of contrast enhancement both enables higher-resolution 3D imaging and provides high blood signal without the need for fast flow. In an initial study of 27 patients,66 CEMRA in combination with spin-echo and precontrast 3D TOF detected 7 of 9 accessory renal arteries, 8 of 9 renal artery stenoses, 4 of 4 celiac stenoses, and all iliac aneurysmal and stenotic disease. Interestingly, MRA correlated with surgical findings as well as CA for defining the proximal extent of the aneurysm. In a subsequent publication by the same group,67 MRA accurately defined the proximal extent of aneurysm in 87% of 38 patients. Sensitivities for iliofemoral occlusive and aneurysmal disease were 83% and 79%, respectively. Sensitivities for renal artery stenosis and accessory renal arteries were each 71%. In terms of surgical planning, MRA correctly predicted the cross-clamp site in 87%, the proximal anastomotic site in 95%, the need for renal revascularization in 91%, and the use of a bifurcated aortic prosthesis in 75%.

In a study of 43 subjects64 that used CEMRA in combination with noncontrast techniques, MRA correctly defined the maximum aneurysm diameter, as well as its proximal and distal extent, in all patients. For detection of aortic branch artery stenosis involving the celiac, superior mesenteric, renal, or iliac arteries, sensitivity was 94%, with specificity of 98%. Potential limitations of the CEMRA technique included the inability to define the severity of branch vessel stenosis and inadequate visualization of the inferior mesenteric artery. The latter can often be adequately evaluated at the time of surgery. The results suggest that performance of CEMRA is adequate to allow the treatment decision to be made between aortic tube graft and aorto-bifemoral graft. With the imminent introduction of aortic stent-graft devices, additional clinical study will be needed to determine whether MRA can substitute for CA in the decision making process with regard to their use. Current resolution appears adequate to screen the renal arteries for disease. However, additional study is indicated before MRA can be relied on for evaluation of the mesenteric circulation. Recent improvements in scanner speed that allow contrast-enhanced aortography to be performed in <30 seconds, or within a breath-hold interval,13 68 69 70 and new blood-pool contrast agents currently under investigation21 may allow additional improvements in this area.

For the thoracic aorta, evaluation of aortic dissection in addition to aneurysmal disease has been another important focus of thoracic MRA. Conventional spin-echo and gradient-echo cine MRI have been shown to be quite effective modalities for evaluation of thoracic aortic disease.71 72 73 74 75 Hartnell et al,76 using noncontrast MRA, suggested that MRA allowed better demonstration of branch vessel stenoses, intimal flaps, and communications into false aneurysms than standard MRI of the aorta. CEMRA was used for evaluation in a range of disease of the thoracic aorta in another study77 ; in 30 patients with angiographic or surgical correlation, MRA successfully demonstrated dissection in 8 (including the correct type), coarctation in 3, and aneurysm in 10, as well as identifying 1 aberrant subclavian artery. Sensitivity for branch vessel stenosis in that study was 90%. MRA must be combined with conventional MRI to detect intramural hematoma or extravascular fluid collections.72

Renal MRA

Recent developments in renal MRA have led to a significant improvement in technical success rate and diagnostic accuracy. Specifically, 3D CEMRA methods have been shown to provide a more reliable depiction of renal artery morphology than noncontrast MRA techniques. In addition, preliminary work with adjunctive MRA techniques suggests that it is possible to determine the hemodynamic significance of renal artery disease by use of MRA and MRI. MRA methods will therefore likely play an important clinical role in the evaluation of patients with renovascular disease.

MRA was initially advocated as a method for imaging the renal arteries as early as 1990.78 The early approaches to renal MRA included 2 different methods, TOF and phase-contrast MRA. These methods are capable of evaluating the aorta and proximal renal arteries, but the more distal branches of the renal artery are obscured by motion artifact and the limited spatial resolution of these techniques. Early reports on the accuracy of renal artery MRA used acquisition methods that did not require the administration of intravenous or intra-arterial contrast; therefore, they were entirely noninvasive. However, the diagnostic accuracy reported by early investigators varied widely, with sensitivities for detecting renal artery stenosis ranging from 53% to 100% (Tables 2Down and 3Down). The variable accuracy reported by different investigators is largely due to differences in technique, the presence of imaging artifacts, and limited spatial resolution. In addition, it is generally understood that the noncontrast methods were associated with a relatively high technical failure rate. Many of the studies reporting the diagnostic accuracy of these nongadolinium methods did not include technically inadequate studies in the determinations of diagnostic accuracy. The early techniques were least successful in those patients with significant renal insufficiency, which is often associated with reduced renal blood flow. Finally, the limited spatial resolution of early nongadolinium methods resulted in a limited ability to detect accessory renal arteries, thereby further compromising the applicability of the technique.


View this table:
[in this window]
[in a new window]
 
Table 2. MRA for Renal Artery Stenosis: TOF Techniques


View this table:
[in this window]
[in a new window]
 
Table 3. MRA for Renal Artery Stenosis: Phase-Contrast Techniques

Recent studies using CEMRA methods have demonstrated a higher technical success rate, as well as substantially improved diagnostic accuracy relative to noncontrast MRA methods. The improvement in image quality associated with CEMRA methods is largely due to improved spatial resolution, as well as an overall reduction in the number of artifacts. CEMRA methods have emerged as a robust, quick, clinically feasible technique to examine the renal arteries. CEMRA methods are also associated with fewer artifacts and can be performed in patients with poor renal blood flow or with renal insufficiency.

CEMRA methods combine a fast gradient-echo image with an intravenous infusion of a gadolinium-DTPA (diethylenetriamine penta-acetic acid) contrast agent. The rapid (typically <30 seconds) intravenous infusion of contrast results in an {approx}30-second window during which the gadolinium-DTPA is concentrated in the aorta and renal arteries. The signal enhancement created by the infusion of gadolinium-DTPA allows the use of a 3D gradient-echo acquisition that can be acquired far more quickly than with noncontrast MRA techniques. An entire set of 3D images is acquired during a single 30-second breath-hold.13 70 79

Several methods have been proposed to account for the differences in the arrival time of the gadolinium contrast at the region of interest. The contrast arrival time is dependent on the patient’s cardiac output, the severity of valvular insufficiency, and the presence of aortic pathology proximal to the renal arteries. Significant variations in the contrast arrival time after an intravenous bolus of gadolinium-DTPA have been demonstrated,13 80 and these differences may be associated with imaging artifacts.8 Therefore, several investigators have addressed techniques to precisely time the image acquisition commensurate with the arrival of contrast in the aorta.13 14 15 In addition, time-resolved CEMRA imaging techniques have been proposed to image the aorta and renal arteries, as well as to assess delays in renal enhancement due to renal artery stenosis.16 Although these new CEMRA methods have improved image quality, results at academic medical centers may not be uniformly applicable to clinical practice. For example, most of these studies have been performed on 1.5-T MR scanners equipped with faster gradient hardware and software. Therefore, whether these results will be equally successful with older MRI hardware or with low-field scanners remains to be determined.

Several investigators have evaluated the diagnostic accuracies of TOF MRA methods for depicting renal artery stenosis using CA as the reference standard78 81 82 83 84 85 86 87 88 89 (Table 2Up). 2D28 and 3D90 TOF approaches have been evaluated. In general, these methods require higher renal blood flow rates to ensure a technically adequate study. This requirement may partly explain the variable technical success and diagnostic accuracy reported with the TOF methods. For example, studies in younger patients with systemic hypertension and normal renal function84 tend to yield better results than studies in patients with hypertension and renal insufficiency.83 In addition, the variable sensitivity reported with these methods is also due in part to variations in the localization of lesions within the renal arteries. For example, studies with a majority of proximal lesions1 tended to show better accuracy than studies with more lesions in the mid and distal renal arteries.14 In general, TOF methods are only useful for evaluating the proximal renal arteries.

The reports of diagnostic accuracy for phase-contrast techniques also vary widely, and the data are summarized in Table 3Up. 2D and 3D phase-contrast methods have been evaluated.89 91 92 93 94 95 96 97 98 99 100 Phase-contrast methods generally show better visualization of the distal renal arteries because the technique is essentially a subtraction method. However, phase-contrast methods are more sensitive to the acquisition parameters selected by the operator, including the selection of the velocity-encoding value for imaging the renal arteries. Because the velocity of renal artery blood flow varies widely between patients, those approaches that incorporated a rapid-velocity prescan96 tended to perform better than approaches that used an "educated guess" at the correct velocity-encoding value.101 Phase-contrast techniques tend to take longer to acquire MRA images than TOF methods. In addition, phase-contrast methods require a longer TE than TOF techniques. The result is a greater degree of signal loss due to turbulent flow at or distal to a stenosis. The presence of this signal loss, however, correlates with the functional significance of the stenosis.102

Noncontrast TOF and phase-contrast methods are limited by the accuracy with which they detect the small accessory renal arteries owing to their inferior spatial resolution. Most reports detected fewer than 50% of accessory renal arteries. The inability to detect these vessels is primarily due to the small field of view used for the TOF and phase-contrast methods, because accessory renal arteries often lie outside the imaging volume when an axial slab is used for imaging.83 89

The development of breath-hold CEMRA has been instrumental for the successful application of this technique for renal MRA. Several investigators have reported techniques for breath-hold CEMRA, with substantially improved spatial resolution and image quality relative to non–breath-hold methods.13 70 79 The technical success rate of breath-hold renal CEMRA was addressed by Wilman et al15 in a study of 25 consecutive patients. The authors found that visualization of the renal arteries was adequate in 24 patients (96%).

Because the CEMRA methods have been developed more recently, fewer studies are available for analysis; these are summarized in Table 4Down.66 104 105 106 107 108 109 In a prospective study comparing the accuracy of breath-hold renal CEMRA relative to digital subtraction x-ray angiography (DSA), Hany et al109 demonstrated a high accuracy for identifying renal lesions of >50% diameter stenosis (sensitivity 93%, specificity 98%). The authors performed image analysis on 78 main and 11 accessory renal arteries. The CEMRA methods allowed better visualization of the accessory renal arteries relative to nonenhanced techniques, primarily because of improved spatial resolution associated with the breath-hold acquisition, as well as a larger imaging field of view. The authors found that the sensitivity and specificity for detecting these renal artery stenoses were 93% and 98%, respectively. Snidow et al106 examined 32 patients using breath-hold CEMRA and compared the results with CA. The CEMRA protocol was intended primarily to evaluate the aorta and iliac inflow, with the parameters selected to include a screening study of the renal arteries. These authors found a sensitivity and specificity of 100% and 89%, respectively, for obstructive lesions of the main renal arteries. Examination of the accuracy for depiction of accessory renal artery occlusive disease demonstrated a sensitivity of 100% and a specificity of 62%. The relatively lower specificity in that study may be related to the selection of acquisition parameters, which were primarily chosen to ensure evaluation of the aorta and iliac inflow vessels, with inadequate spatial resolution to depict small accessory renal arteries.


View this table:
[in this window]
[in a new window]
 
Table 4. MRA for Renal Artery Stenosis: Contrast-Enhanced Techniques

It is widely recognized that CA has several limitations, although it is the accepted reference standard for identifying renal artery stenosis. CA is a projectional method that must rely on oblique imaging planes to profile a stenosis that appears "en face" in the renal artery. This limitation makes it difficult to precisely profile ostial lesions or to delineate lesions associated with an asymmetrical luminal narrowing. In addition, the CA technique provides information predominately about the vessel lumen, poststenotic dilatation, and relative renal perfusion. DSA does not provide an assessment of alterations in the renal artery blood flow or pressure gradients across a renal artery stenosis. Data regarding the pressure gradient and renal vein renin production may be obtained by use of additional catheter placements under fluoroscopic guidance.

With these limitations of CA in mind, Wasser et al100 used a measurement of the pressure gradient across the renal artery stenosis as a reference standard to compare the diagnostic accuracy of ECG-gated 3D phase-contrast MRA and conventional DSA. The authors used a pressure gradient of 15 mm as the definition of a significant stenosis. DSA detected 10 of 13 stenoses (sensitivity 77%, specificity 92%). MRA demonstrated the loss of distal flow signal intensity in 12 of these stenoses, for a sensitivity of 92% and a specificity of 75%. There was no significant difference between the 2 modalities in grading hemodynamic significance of the stenosis, although the ability to detect differences between the 2 methods was limited by the small subject population.

Schoenberg et al,99 in a study involving ECG-gated phase-contrast MR evaluation of alterations in renal artery blood-flow waveforms, demonstrated changes in the flow waveforms in patients with significant (>50% diameter) renal artery stenoses on CA. The authors, using techniques analogous to Doppler ultrasound measurement of flow waveforms, demonstrated a sensitivity of 100% and a specificity of 93% for identifying stenoses. The authors relied on a qualitative assessment of the renal artery flow waveform, as well as quantitative assessment of changes in the time to maximal flow after the ECG trigger. These data suggest that physiological waveform alterations may indeed be detected by MR phase-contrast flow measurements.

In an effort to step beyond diagnostic correlation studies comparing MRA with CA or pressure measurements, Prince et al102 sought to demonstrate that findings on MRA studies could be used to predict the likelihood of success after revascularization, as measured by a reduction in the need for antihypertensive medications or an improvement in renal function after revascularization. The investigators performed a retrospective study that combined CEMRA for detection of renal artery stenosis with phase-contrast MRA for prediction of patient outcome after revascularization. They found that signal loss distal to a stenosis on the phase-contrast MRA examinations was significantly more prominent in those patients with hemodynamically significant stenoses. Poststenotic arterial dilatation, reduced renal length, and reduction in parenchymal renal thickness were good predictors of patient outcome after revascularization.

In summary, recent developments in renal MRA have led to a significant improvement in the technical success rate and diagnostic accuracy of the methods. Specifically, 3D breath-hold CEMRA methods have been shown to provide a more reliable depiction of renal artery morphology. The new methods are easier to perform, are capable of providing higher spatial resolution, and are associated with fewer artifacts. Although the older noncontrast (phase-contrast and TOF) methods are less reliable for demonstrating renal artery anatomy, the phase-contrast techniques have several important attributes that may make it feasible to demonstrate the hemodynamic significance of a renal artery stenosis. Recent work in MRA has demonstrated that MRA methods may be used to demonstrate the hemodynamic significance of renal artery stenosis and predict patient outcome in response to revascularization therapy.

Peripheral MRA

Peripheral arterial occlusive disease is a local manifestation of the systemic atherosclerotic process. Patients present with a wide range of symptoms depending on the severity of disease and the degree of peripheral ischemia. Treatment options depend on the underlying arterial anatomy and include percutaneous balloon angioplasty, bypass surgery, and amputation. Patients with focal disease will most likely have percutaneous balloon angioplasty, whereas those with long occluded segments or diffuse disease usually undergo surgery. Regardless of the type of revascularization, a preoperative road map is required before intervention, and for the past 3 decades, CA has served this purpose. The recent development and refinement of peripheral vascular MRA has challenged catheter-based approaches and at some centers has replaced it in the evaluation and surgical planning of patients with symptomatic peripheral vascular disease. Its use in individual institutions depends on the expertise of the MR radiologist performing the examination and the ability of vascular surgeons and interventional radiologists to plan interventions based on the MRA images.

Most of the initial work on peripheral MRA was performed on infrainguinal vessels with a local surface coil used as an RF receiver. These vessel segments provide an ideal environment for MRA because they are not tortuous, they are unidirectional, and they have steady, nonturbulent flow. The first reports of imaging peripheral vessels with MRA began to appear in the late 1980s and early 1990s.110 111 112 113 114 115 116 117 118 Although most of these were feasibility studies, peripheral MRA was found to be remarkably accurate at depicting peripheral vascular disease compared with CA. 2D TOF was the pulse sequence most commonly used, sometimes combined with phase-contrast MRA.

In 1991, Mulligan et al119 compared 2D TOF peripheral MRA with CA and ultrasound in 12 patients. There was 71% agreement between MRA and CA. Other investigators used 2D TOF MRA and had similar results.120 121 122 123 In 1992, Owen et al124 studied tibial vessels with MRA and CA in patients with severe peripheral vascular disease. Surprisingly, more arterial segments were seen on MRA than on contrast angiography, with these "angiographically occult" runoff vessels confirmed at surgery. Surgical plans were also changed on the basis of the additional information obtained from MRA. The results of the study by Owen et al124 provoked much discussion113 125 126 127 and led to several follow-up investigations. In a study of 51 patients, Carpenter et al128 found that in 48% of vessel segments, MRA demonstrated information that was not present on CA. The additional information led to a change in the surgical plan for 22% of the patients. Similar investigations by McCauley et al129 in 1994, McDermott et al130 in 1995, and Huber et al131 in 1997 also demonstrated more vascular segments, as did a multicenter trial performed by the American College of Radiology.132 In the latter study, 155 patients from 6 hospitals were studied with similar imaging hardware and software. Peripheral MRA and CA were compared with the gold standard of intraoperative x-ray angiography. There was 81% agreement between CA and MRA, whereas MRA was found to be more sensitive in the identification of diseased tibial runoff vessels. This was especially true in the most distal segments of the ankle and foot. The superiority of MRA over CA in imaging pedal vessels is a result of the fundamental differences between the 2 techniques. CA requires opacification of vessels with contrast in order for them to be visualized radiographically, whereas 2D TOF MRA only needs flowing blood. By the appropriate selection of imaging parameters, blood flows <10 cm/s can be readily detected by MRA. It has been shown that because of the high sensitivity of MRA for visualizing tibial vessels, the addition of this test can be cost-effective.131 133

Because of the initial success of 2D TOF MRA in visualizing the infrainguinal arteries, many investigators used the same technique to image the abdominal and pelvic vessels. Unfortunately, the results in these investigations varied widely among institutions. Snidow and colleagues134 compared CA and 2D TOF MRA in 50 patients and only found 52% agreement between techniques. The same group in another comparison study concluded, "For evaluation of symptomatic lower extremity ischemia, 2D TOF MRA cannot be considered a reliable substitute for x-ray angiography."135

Other centers have demonstrated better success using similar techniques. Carpenter et al135A performed a 2D TOF MRA and CA study in 47 patients and formulated separate interventional plans for each modality. The results of the 2 imaging studies were identical in 41 (87%) of 47 patients. MRA accurately detected patent and occluded arterial segments (sensitivity 99.6%; specificity 100%; positive predictive value 100%; negative predictive value 98.6%). Therapeutic plans based on either MRA or CA were identical for every patient. MRA imaging times, however, were >2 hours and only included the pelvis and a single extremity. In addition, almost 25% of patients needed a second MRA because of flow-related artifacts. This showed that although it was possible to plan revascularization with 2D TOF MRA, it was not clinically practical.

To overcome the flow-related artifacts and prolonged imaging times that have plagued 2D TOF MRA, several variations in imaging techniques have been examined.136 137 138 139 140 141 142 143 144 145 146 Although helpful, these approaches failed to accomplish the goal of a robust technique for an aortogram and runoff examination that could be performed in <2 hours.

Although peripheral MRA has proved accurate in the depiction of focal peripheral vasculature, the examination can be extremely lengthy. To visualize small tibial vessels, surface coils are needed that can image the entire lower extremity during a single series. At present, the examination requires several overlapping series that, when pieced together, produce a vascular road map. The surface coil must be repositioned after each series is performed. Extended-length surface coils have been developed in the hope of speeding examination times without any loss of spatial resolution.144 147 148

As with renal, aortic, and carotid angiography, CEMRA has proved helpful for peripheral MRA. In 1992, Lossef et al112 evaluated the efficacy of gadopentetate dimeglumine (Gd-DTPA) in the lower extremities by use of a flow phantom, 7 healthy volunteers, and 7 patients with peripheral vascular disease. It was determined that the intravenous administration of a bolus of Gd-DTPA improved MRA quality in the phantom, volunteers, and patients. However, in some cases, venous overlap, imaging artifacts, and suboptimal visualization of subtle lesions limited interpretation.112 In a feasibility study, Prince and colleagues149 used 3D CEMRA to evaluate the abdominal aorta and the renal, visceral, and iliac arteries in 16 patients. They found that by imaging dynamically during the arterial phase of a 5-minute injection, preferential arterial enhancement could be achieved with minimal enhancement of venous structures and no flow-related artifacts. Additional studies have confirmed the speed and accuracy of peripheral CEMRA.102 150 Poon et al151 compared the diagnostic accuracy of CEMRA for the measurement of percent stenosis in the iliac arteries rather than using an arbitrary cutoff for significant disease. A recent refinement of CEMRA for peripheral vascular disease has used a stepping table with digital subtraction, comparable to the technique used to perform a peripheral runoff examination with x-ray angiography, although at lower resolution.152 153 Those studies showed that whereas stenosis measurements for 2D TOF MRA were significantly different from those obtained on CA, the values obtained with CEMRA were statistically indistinguishable from CA. The use of CEMRA for the inflow (aortoiliac) vessels has become commonplace as part of a complete peripheral MRA examination at many centers.

The timing of arrival of the contrast bolus with imaging is critical. After the arterial first pass of a bolus of MR contrast, there is rapid equilibration followed by renal excretion. If imaging is delayed and the arterial phase is missed, venous and soft tissue enhancement occurs. In addition, sometimes there is a need for several areas to be imaged sequentially, which would require multiple administrations of contrast. For these reasons, subtraction techniques were developed that allow for multiple injections of contrast.150 154 155 These techniques, alone or in combination with moving-table technology, may permit further inroads of CEMRA into TOF MRA of the extremity. Initial results have also been reported for peripheral MRA with blood-pool contrast agents.21 Dynamic images comparable to conventional agents were obtained at substantially lower doses. Steady-state images demonstrated enhancement of the blood pool for up to 50 minutes, and vessels as small as 1 mm in diameter were seen. Overlap of arteries and veins on MIP images remains an issue for these agents. Possible approaches include the use of multiplanar reformatting and development of automated segmentation techniques.21

After initial feasibility and comparison studies were performed, investigators studied the utility of MRA to plan bypass surgery.122 156 157 158 This requires imaging from the infrarenal aorta through the plantar arch. In the largest series reported to date, 80 consecutive patients successfully underwent peripheral bypass surgery based solely on a preoperative MRA.159 Most authors suggest that this should be attempted only at centers with expertise in peripheral MRA and where these techniques have been validated against the gold standard of CA.

MRA has proved useful in a variety of disease states in addition to peripheral vascular disease, including popliteal entrapment and aneurysms.160 161 162 163 164 165 It has also been found useful in identifying patient tibial vessels before fibular graft harvesting.166

There has been and continues to be much progress since peripheral MRA was introduced almost a decade ago, and at selected centers with high expertise, MRA has been able to replace CA for patients with symptomatic peripheral occlusive disease. If properly performed and interpreted, MRA can replace diagnostic CA by combining the speed of CEMRA of the iliac and femoral arteries with the high sensitivity of 2D TOF of the tibial arteries. The eventual role of CEMRA with multiple injections, moving-table technology, and/or blood-pool contrast agents remains to be determined.

MRA of the Coronary and Pulmonary Arteries

Of all vascular regions, coronary MRA has been considered the most technically demanding, largely because of the small size of the coronaries (3 to 4 mm in diameter),167 their tortuous and complex pathway, the abundant signal from surrounding epicardial fat, and the significant motion associated with both respiration168 and cardiac motion.169 Tremendous progress has been made since the first clinical reports of human coronary MRA were published in 1991,24 yet current clinical applications are limited to evaluation of anomalous coronary arteries and assessment of coronary artery bypass graft patency.

Virtually all current applications of native coronary MRA use ECG gating to minimize motion artifacts related to cardiac motion. Early attempts at coronary MRA with conventional spin-echo MR were met with limited success. In 1984, Lieberman et al 170 were able to visualize portions of the coronaries in only 7 of 23 patients, whereas Paulin and colleagues23 identified the origin of the left main coronary artery in all 6 subjects examined and the RCA ostium in 4 of the 6 individuals. No coronary stenoses were identified in either report.

Recent advances in both MR hardware and software have led to the development of imaging strategies that enable visualization of the proximal portion of the major epicardial coronary arteries in the vast majority of subjects. Although not yet standardized, current methods for coronary MRA combine ECG gating with fast imaging sequences and techniques for suppressing signal from surrounding epicardial fat and for myocardial suppression of artifacts related to respiratory motion.

The greatest body of clinical investigation has been performed with segmented gradient-echo sequences, first described in an animal model by Burstein171 and subsequently reported in humans by Edelman et al.24 This method reduces the number of heartbeats required to complete an MR image and uses an acquisition window of 100 to 150 ms during mid diastole, a period of relative bulk cardiac diastasis but with relatively high coronary blood flow. Typically, 8 interleaved phase-encoding steps are acquired during each cardiac cycle, necessitating 16 successive heartbeats to complete a 128x256 matrix. This allows for the acquisition of a single 2D image during a single 15- to 20-second breath-hold, thereby minimizing respiratory motion artifacts. A slice thickness of 3 to 4 mm and a 240-mm field of view (in-plane spatial resolution of {approx}1.9x0.9 mm) are often used. Forty or more breath-holds may be needed to complete a coronary MRA study. As with other gradient-echo approaches, rapidly moving laminar blood flow appears as bright signal because of the inflow of unsaturated protons. Focal areas of turbulence, such as those caused by coronary stenoses, appear dark owing to signal loss from dephasing.

Several other general approaches to coronary MRA have been described, including 3D TOF172 methods that offer enhanced signal along with reconstruction advantages for visualization of the tortuous coronary vasculature. However, 3D imaging is prone to blurring,173 because scan times usually far exceed that of a single breath-hold. MR subtraction methods that involve selective tagging of blood in the aortic root and suppression of background tissue have also been described.174 This approach has great appeal for defining the proximal coronary artery but may be more limiting for more distal regions. Other imaging techniques include spiral acquisitions175 and echo-planar imaging.176 177 178

The identification of the origin and proximal/mid portions of the native coronary arteries is currently possible in the vast majority of individuals. A summary of recent coronary MRA studies reporting on visualization of coronary arteries in healthy volunteers and patients with coronary artery disease is presented in Table 5Down. Both 2D24 179 180 181 182 and 3D172 178 179 183 184 185 186 approaches have been used at multiple centers with similar results. The origin and proximal coronary arteries are visualized in >95% of subjects. In general, more extensive portions of the LAD and RCA are seen.180 181 182 In angiographically normal vessels, proximal coronary MRA diameter corresponds well to CA.180 187


View this table:
[in this window]
[in a new window]
 
Table 5. Current Experience With Coronary MRA in Humans

The ability of coronary MRA to reliably identify the native vessels, coupled with the inherent 3D representation of MR data, renders this technique ideally suited for the noninvasive identification and characterization of anomalous coronary arteries. Although they are rare, occurring in only 0.6% to 1.2% of adults referred for CA,188 anomalies associated with passage of the anomalous vessel between the aorta and pulmonary artery can result in impaired myocardial perfusion and have been associated with sudden death.189 Although the presence of anomalous vessels is well defined by CA, their passage anterior or posterior to the aorta can sometimes be difficult to appreciate. With the use of breath-hold 2D TOF approaches, accurate identification and the course of anomalous coronary vessels have been well described.190 191 192 In many studies, coronary MRA has been found to be superior to CA because it has identified prior misclassification (by CA) of anomalous vessels.190 192 These data strongly support a clinical role for coronary MRA as a noninvasive tool for defining the anatomic course in patients with known or suspected anomalous coronary vessels.

Although coronary MRA is sufficiently advanced to allow for definition of anomalous coronary vessels, the data do not support a general role for identification of native vessel stenoses. The spatial resolution of the most widely tested breath-hold coronary MRA approach precludes quantitative coronary MRA. Most series have used the visual identification of a prominent signal "void" as being indicative of a diameter stenosis of >50%. Although the technique has been shown to accurately identify proximal and mid-coronary stenoses in some centers,193 194 195 with good correlation with location of stenoses and with length of the coronary lesion slightly overestimated by MR,194 data from other centers have been more variable (Table 6Down).196 197 198 199


View this table:
[in this window]
[in a new window]
 
Table 6. Summary of Current Coronary MRA Studies for the Detection of Significant Coronary Artery Disease (>50% Diameter Stenosis)

Much of the variability in coronary MRA success is likely related to the lack of sequence standardization (data acquisition interval and TE) and the dependence on the breath-hold for respiratory artifact suppression. Frequently, >=40 sustained breath-holds are required to complete a scan, a requirement that often leads to both patient and operator fatigue. Improvements in data acquisition with the use of either far fewer breath-holds or free-breathing strategies promise to simplify the approach, improve spatial resolution, and likely improve accuracy. However, a growing obstacle to coronary MRA is the use of intracoronary stents. Although not a contraindication for MRA200 (up to field strengths of 1.5 T), local signal voids induced by the stent preclude assessment of the stented portion of the vessel.201

The relatively stationary position of saphenous vein and internal mammary artery bypass grafts offsets many of the technical challenges of native vessel coronary MRA. Furthermore, their relatively straight and predictable course has allowed imaging of bypass grafts with conventional ECG-gated spin-echo202 203 204 205 and gradient-echo202 206 207 MR techniques. For most of these investigations, contiguous transverse images are obtained at levels corresponding to that expected for the bypass graft. The graft is then characterized as "patent" if the normal signal void (spin echo) or bright signal (gradient echo) of laminar blood flow is seen in at least 2 contiguous anatomic levels in the expected region of the bypass graft. If a signal void is seen at only 1 level, a graft is considered "indeterminate," and if no signal of laminar blood flow is seen, the graft is considered occluded. Both spin-echo and gradient-echo approaches have a similar accuracy (Table 7Down). The addition of CEMRA208 209 210 may further improve accuracy for bypass grafts. In an unusual case,211 coronary MRA was able to correctly identify the bypass graft and confirm patency of the native LAD distal to the graft anastomosis when CA failed to do so.


View this table:
[in this window]
[in a new window]
 
Table 7. Summary of Studies Evaluating Coronary MRA for the Assessment of Graft Patency After Aortocoronary Bypass Grafting

As with intravascular stents, a major obstacle for bypass graft imaging is the local signal loss or artifact associated with implanted metallic objects such as hemostatic clips and ostial graft rings. Artifacts from graft markers or clips may be indistinguishable from signal "voids" related to blood flow through a stenosis in the bypass graft. In addition, grafts with very tight stenoses, in which blood flow would be expected to be low, often result in insufficient contrast to characterize a graft as patent.204 A determination of the ultimate role of MRA in bypass graft evaluation in clinical practice will require further study.

Although current coronary MRA strategies are acceptable for evaluation of anomalous coronary arteries and bypass graft patency, the major impact of coronary MRA will likely be felt when methods are sufficiently robust to allow for evaluation of the native coronary arteries. Current advances in MR hardware and software have focused on 2 different areas: suppression of respiratory motion artifacts and CEMRA.

Although breath-holding has been widely used to minimize respiratory motion artifacts, this approach requires significant patient training and cooperation, as well as operator expertise. Breath-hold variability introduces slice registration errors, and the breath-hold duration limits the use of potential enhancements, such as 3D acquisitions, signal averaging, and higher-resolution coronary MRA. Imaging during free (unrestricted) breathing offers obvious advantages over breath-holding. A novel technique for minimizing respiratory-related motion is with the use of real-time navigator echoes25 27 212 213 214 215 216 217 that interrogate an interface such as the diaphragm-lung or heart-lung.25 Positional changes of this interface along the axis of the navigator can then be followed through the normal respiratory cycle. Image acquisition then can be gated to respiratory motion: data are acquired (prospective gating) or accepted (retrospective gating) only when the interface is within a preselected window near the end-expiratory position. A more elaborate variation is to prospectively correct the slice position on the basis of navigator data, thereby allowing for the use of wider gating windows with improved time efficiency and image quality.27 218 An alternative method for minimization of respiratory motion artifacts would be to decrease the acquisition time so that an entire coronary MRA data set could be obtained in several brief breath-holds.27 178 212 218 Enhanced gradient functions and use of MR contrast agents (see below) will potentially allow for such rapid imaging.

Intravenous contrast media have had an important impact on aortic, renal, and peripheral MRA11 and promise to provide similar enhancements for coronary MRA. Recently, the use of contrast media has been described for coronary MRA in experimental animals for assessment of bypass graft patency208 209 210 219 and for visualization of the native coronary arteries in humans.183

Coronary MRA is a noninvasive diagnostic method currently in a state of rapid evolution. Advances in technology will undoubtedly lead to enhanced resolution, improved accuracy, and shorter scan times. It is almost certain that coronary MRA will be a prominent diagnostic clinical tool in the future. In anticipation of this, the development and use of more MR "friendly" intracoronary stents, graft markers, and vascular clips are strongly encouraged.

Pulmonary MRA has also been slow to develop into a clinically useful application. Initial approaches used 2D or 3D TOF techniques, the former with and the latter without breath-holding.22 220 221 222 223 Some encouraging results were achieved, but the techniques were not reliable enough to allow widespread clinical use. Recently, the use of CEMRA has been extended to the pulmonary arteries.224 225 226 227 228 229 It is possible to cover the entire pulmonary tree in a single breath-hold during the injection of contrast. Meaney et al224 reported on 30 patients who underwent both breath-hold pulmonary CEMRA and CA for suspected pulmonary embolism. A spatial resolution of 1.25x2.5x3.0 mm was achieved in that study. For the 3 readers, sensitivity ranged from 75% to 100%, with specificities of 95% to 100%. As scan speeds improve, resolution promises to further increase. Steiner et al225 reported spatial resolution of 1.7x1.7x2.0 mm in a 23-second breath-hold. Recently, the blood pool contrast agent NC100150 has been used to achieve free-breathing pulmonary MRA with navigator echoes in normal subjects.226 With continuing improvements in technology and validation in larger groups of diseased patients, pulmonary MRA may become competitive with the other minimally invasive imaging modalities for pulmonary embolism, including radionuclide ventilation-perfusion scanning and CT angiography.

Conclusions
MRA continues to show rapid evolution. The hallmark of the last few years has been the widespread interest in CEMRA, which has essentially displaced many of the flow-dependent MR strategies used previously. Current CEMRA uses extracellular-fluid gadolinium contrast agents in first-pass mode. Progress in several areas will determine the evolution of MRA in the near future. Several intravascular contrast agents are in clinical trials, and the role of these agents appears promising but remains to be determined. Hardware and software advances continue, with development of ever-faster scanning, more powerful image processing, and dedicated devices such as cardiac and peripheral coils and movable tables for peripheral MRA.

The current status of MRA can be summarized as follows (Table 8Down): carotid MRA has established itself as a complement to DUS for preoperative evaluation, markedly lessening the need for CA. MRA has gained widespread acceptance for imaging the aorta. The use of aortography can now be limited to special situations. In the renal arteries, MRA has an established role for main renal artery screening. Additional advances will be required to replace CA for evaluation of intrarenal vessels and nuclear medicine for renal functional evaluation. In peripheral vascular disease, MRA can replace CA now in patients in whom catheter introduction or iodinated contrast poses special risks (eg, because of contrast allergy, renal failure, or severe inflow disease). With the introduction of CEMRA and the development of moving-table technology and blood-pool contrast agents, MRA appears poised to replace CA for routine imaging in the near future. Coronary and pulmonary MRA are areas of very active investigation, and a routine clinical role for these evolving technologies remains to be established. In particular, they will need to prove themselves not only against CA, but also with other "noninvasive" technology such as electron beam coronary CT and helical CT pulmonary angiography, respectively.


View this table:
[in this window]
[in a new window]
 
Table 8. Summary of Recommendations for MRA

References

1. Edelman RR, Mattle HP, Wallner B, Bajakian R, Kleefield J, Kent C, Skillman JJ, Mendel JB, Atkinson DJ. Extracranial carotid arteries: evaluation with "black blood" MR angiography. Radiology. 1990;177:45–50.[Abstract/Free Full Text]

2. Edelman RR, Chien D, Kim D. Fast selective black blood MR imaging. Radiology. 1991;181:655–660.[Abstract/Free Full Text]

3. Keller P. Time of flight magnetic resonance angiography. Neuroimaging Clin N Am. 1992;4:639–656.

4. Dumoulin CL. Phase contrast MR angiography techniques. Magn Reson Imaging Clin N Am. 1995;3:399–411.[Medline] [Order article via Infotrieve]

5. Atkinson DJ, Edelman RR. Cineangiography of the heart in a single breath-hold with a segmented turboFLASH sequence. Radiology. 1991;178:357–360.[Abstract/Free Full Text]

6. Walker MF, Souza SP, Dumoulin CL. Quantitative flow measurement in phase contrast MR angiography. J Comput Assist Tomogr. 1988;12:304–313.[Medline] [Order article via Infotrieve]

7. Saloner D. MRA: principles and display. In: Higgins CB, Hricak H, Helms CA, eds. Magnetic Resonance Imaging of the Body. 3rd ed. Philadelphia, Pa: Lippincott-Raven; 1997:1345–1368.

8. Laub G. Displays for MR angiography. Magn Reson Med. 1990;14:222–229.[Medline] [Order article via Infotrieve]

9. Anderson CM, Saloner D, Tsuruda JS, Shapeero LG, Lee R. Artifacts in maximum-intensity-projection display of MR angiograms. Am J Roentgenol. 1990;154:623–629.[Free Full Text]

10. Blatter DD, Parker DL, Robison RO. Cerebral MR angiography with multiple overlapping thin slab acquisition, I: quantitative analysis of vessel visibility. Radiology. 1991;179:805–811.[Abstract/Free Full Text]

11. Prince MR. Gadolinium-enhanced MR aortography. Radiology. 1994;191:155–164.[Abstract/Free Full Text]

12. Prince MR. Body MR angiography with gadolinium contrast agents. Magn Reson Imaging Clin N Am. 1996;4:11–24.[Medline] [Order article via Infotrieve]

13. Earls JP, Rofsky NM, DeCorato DR, Krinsky GA, Weinreb JC. Breath-hold single-dose gadolinium-enhanced three-dimensional MR aortography: usefulness of a timing examination and MR power injector. Radiology. 1996;201:705–710.[Abstract/Free Full Text]

14. Foo TK, Saranathan M, Prince MR, Chenevert TL. Automated detection of bolus arrival and initiation of data acquisition in fast, three-dimensional, gadolinium-enhanced MR angiography. Radiology. 1997;203:275–280.[Abstract/Free Full Text]

15. Wilman AH, Riederer SJ, King BF, Debbins JP, Rossman PJ, Ehman RL. Fluoroscopically triggered contrast-enhanced three-dimensional MR angiography with elliptical centric view order: application to the renal arteries. Radiology. 1997;205:137–146.[Abstract/Free Full Text]

16. Korosec FR, Frayne R, Grist TM, Mistretta CA. Time-resolved contrast-enhanced 3D MR angiography. Magn Reson Med. 1996;36:345–351.[Medline] [Order article via Infotrieve]

17. Mayo-Smith WW, Saini S, Slater G, Kaufman JA, Sharma P, Hahn PF. MR contrast material for vascular enhancement: value of superparamagnetic iron oxide. AJR Am J Roentgenol. 1996;166:73–77.[Abstract/Free Full Text]

18. Anzai Y, Prince MR, Chenevert TL, Maki JH, Londy F, London M, McLachlan SJ. MR angiography with an ultrasmall superparamagnetic iron oxide blood pool agent. J Magn Reson Imaging. 1997;7:209–214.[Medline] [Order article via Infotrieve]

19. Wilkstrom LJ, Johansson LO, Ericsson BA, Borseth A, Akeson PA, Ahlstrom KH. Abdominal vessel enhancement with an ultrasmall, superparamagnetic iron oxide blood pool agent: evaluation of dose and echo time dependence at different field strengths. Acad Radiol. 1999;6:292–298.[Medline] [Order article via Infotrieve]

20. Lauffer RB, Parmalee DJ, Dunham SU, Ouellet HS, Dolan RP, Witte S, McMurry TJ, Walovitch RC. MS-325: an albumin-targeted contrast agent for MR angiography. Radiology. 1998;207:529–538.[Abstract/Free Full Text]

21. Grist TM, Korosec FR, Peters DC, Witte S, Walovitch RC, Dolan RP, Bridson WE, Yucel EK, Mistretta CA. Steady-state and dynamic MR angiography with MS-325: initial experience in humans. Radiology. 1998;207:539–544.[Abstract/Free Full Text]

22. Grist TM, Sostman HD, MacFall JR, Foo TK, Spritzer CE, Witty L, Newman GE, Debatin JF, Tapson V, Saltzman HA. Pulmonary angiography with MR imaging: preliminary clinical experience. Radiology. 1993;189:523–530.[Abstract/Free Full Text]

23. Paulin S, von Schulthess GK, Fossel E, Krayenbuehl HP. MR imaging of the aortic root and proximal coronary arteries. AJR Am J Roentgenol. 1987;148:665–670.[Abstract/Free Full Text]

24. Edelman RR, Manning WJ, Burstein D, Paulin S. Coronary arteries: breath-hold MR angiography. Radiology. 1991;181:641–643.[Abstract/Free Full Text]

25. McConnell MV, Khasgiwala VC, Savord BJ, Chen MH, Chuang ML, Edelman RR, Manning WJ. Comparison of respiratory suppression methods and navigator locations for MR coronary angiography. AJR Am J Roentgenol. 1997;168:1369–1375.[Abstract/Free Full Text]

26. Grimm RC, Rossman PJ, Reiderer SJ, Ehman RL. Real-time adaptive motion correction using navigator echoes. Proceedings of the 3rd Annual Meeting of the Society of Magnetic Resonance, Nice, France, 1995. Berkeley, Calif: Society of Magnetic Resonance; 1995:741.

27. Danias PG, McConnell MV, Khasgiwala VC, Chuang ML, Edelman RR, Manning WJ. Prospective navigator correction of image position for coronary MR angiography. Radiology. 1997;203:733–736.[Abstract/Free Full Text]

28. Keller PJ, Drayer BP, Fram EK, Williams KD, Dumoulin CL, Souza SP. MR angiography with two-dimensional acquisition and three-dimensional display: work in progress. Radiology. 1989;173:527–532.[Abstract/Free Full Text]

29. Masaryk TJ, Modic MT, Ruggieri PM, Ross JS, Laub G, Lenz GW, Tkach JA, Haacke EM, Selman WR, Harik SI. Three-dimensional (volume) gradient echo imaging of the carotid bifurcation: preliminary clinical experience. Radiology. 1989;171:801–806.[Abstract/Free Full Text]

30. Cloft HJ, Murphy KJ, Prince MR, Brunberg JA. 3D gadolinium-enhanced MR angiography of the carotid arteries. Magn Reson Imaging. 1996;14:593–600.[Medline] [Order article via Infotrieve]

31. Parker DL, Yuan C, Blatter DD. MR angiography by multiple thin slab 3D acquisition. Magn Reson Med. 1991;17:434–451.[Medline] [Order article via Infotrieve]

32. Krinsky G, Rofsky N, Flyer M, Giangola G, Maya M, DeCoroto D, Earls J, Weinreb J. Gadolinium-enhanced three-dimensional MR angiography of acquired arch vessel disease. AJR Am J Roentgenol. 1996;167:981–987.[Free Full Text]

33. Carpenter JP, Holland GA, Golden MA, Barker CF, Lexa FJ, Gilfeather M, Schnall MD. Magnetic resonance angiography of the aortic arch. J Vasc Surg. 1997;25:145–151.[Medline] [Order article via Infotrieve]

34. Urchuk SN, Plewes DB. Mechanisms of flow-induced signal loss in MR angiography. J Magn Reson Imaging. 1992;2:453–462.[Medline] [Order article via Infotrieve]

35. Rossnick S, Laub G, Braeckle R. Three dimensional display of blood vessels in MRI. Comput Cardiol. 1986:193–195.

36. De Marco JK, Nesbit GM, Wesbey GE, Richardson D. Prospective evaluation of extracranial carotid stenosis: MR angiography with maximum-intensity projections and multiplanar reformation compared with conventional angiography. AJR Am J Roentgenol. 1994;163:1205–1212.[Abstract/Free Full Text]

37. Anderson CM, Lee RE, Levin DL, de la Torre Alonso S, Saloner D. Measurement of internal carotid artery stenosis from source MR angiograms. Radiology. 1994;193:219–226.[Abstract/Free Full Text]

38. Levi CR, Mitchell A, Fitt G, Donnan GA. The accuracy of magnetic resonance angiography in the assessment of extracranial carotid artery occlusive disease: a comparison with digital subtraction angiography using NASCET criteria for stenosis measurement. Cerebrovasc Dis. 1996;6:231–236.

39. Korogi Y, Takahashi M, Mabuchi N, Miki H, Shiga H, Watabe T, O’Uchi T, Nakagawa T, Horikawa Y, Fujiwara S, Furuse M. Intracranial vascular stenosis and occlusion: diagnostic accuracy of three-dimensional, Fourier transform, time-of-flight MR angiography. Radiology. 1994;193:187–193.[Abstract/Free Full Text]

40. Liberopoulos K, Kaponis A, Kokkinis K, Pagratis N, Nicolakopoulou Z, Douskou M, Stringaris K, Clonaris C, Balas P. Comparative study of magnetic resonance angiography, digital subtraction angiography, duplex ultrasound examination with surgical and histological findings of atherosclerotic carotid bifurcation disease. Int Angiol. 1996;15:131–137.[Medline] [Order article via Infotrieve]

41. Link J, Brinkmann G, Steffens JC, Graessner J, Muller-Hulsbeck S, Heller M. MR angiography of the carotid arteries in 3D TOF technique with sagittal double-volume acquisition using new head-neck coil [in German]. Rofo Fortschr Geb Rontgenstr Neuen Bildgeb Verfahr. 1996;165:544–550.[Medline] [Order article via Infotrieve]

42. Vogl TJ, Heinzinger K, Juergens M, Kutter R, Hepp W, Balzer JO, Haupt G, Banzer D, Felix R. Multiple slab MR angiography of the A. carotis interna: a preoperative comparative study [in German]. Rofo Fortschr Geb Rontgenstr Neuen Bildgeb Verfahr. 1995;162:404–411.[Medline] [Order article via Infotrieve]

43. Patel MR, Kuntz KM, Klufas RA, Kim D, Kramer J, Polak JF, Skillman JJ, Whittemore AD, Edelman RR, Kent KC. Preoperative assessment of the carotid bifurcation: can magnetic resonance angiography and duplex ultrasonography replace contrast arteriography? Stroke. 1995;26:1753–1758.[Abstract/Free Full Text]

44. Nicholas GG, Osborne MA, Jaffe JW, Reed JF III. Carotid artery stenosis: preoperative noninvasive evaluation in a community hospital. J Vasc Surg. 1995;22:9–16.[Medline] [Order article via Infotrieve]

45. Kent KC, Kuntz KM, Patel MR, Kim D, Klufas RA, Whittemore AD, Polak JF, Skillman JJ, Edelman RR. Perioperative imaging strategies for carotid endarterectomy: an analysis of morbidity and cost-effectiveness in symptomatic patients. JAMA. 1995;274:888–893.[Abstract/Free Full Text]

46. Mittl RL Jr, Broderick M, Carpenter JP, Goldberg HI, Listerud J, Mishkin MM, Berkowitz HD, Atlas SW. Blinded-reader comparison of magnetic resonance angiography and duplex ultrasonography for carotid artery bifurcation stenosis. Stroke. 1994;25:4–10.[Abstract]

47. Young GR, Humphrey PR, Shaw MD, Nixon TE, Smith ET. Comparison of magnetic resonance angiography, duplex ultrasound, and digital subtraction angiography in assessment of extracranial internal carotid artery stenosis. J Neurol Neurosurg Psychiatry. 1994;57:1466–1478.[Abstract/Free Full Text]

48. Young GR, Sandercock PA, Slattery J, Humphrey PR, Smith ET, Brock L. Observer variation in the interpretation of intra-arterial angiograms and the risk of inappropriate decisions about carotid endarterectomy. J Neurol Neurosurg Psychiatry. 1996;60:152–157.[Abstract/Free Full Text]

49. Vanninen R, Manninen H, Soimakallio S. Imaging of carotid artery stenosis: clinical efficacy and cost-effectiveness. AJNR Am J Neuroradiol. 1995;16:1875–1883.[Abstract]

50. Gagne PJ, Matchett J, MacFarland D, Hauer-Jensen M, Barone GW, Eidt JF, Barnes RW. Can the NASCET technique for measuring carotid stenosis be reliably applied outside the trial? J Vasc Surg. 1996;24:449–455; discussion 455–456.[Medline] [Order article via Infotrieve]

51. Eliasziw M, Fox AJ, Sharpe BL, Barnett HJ. Carotid artery stenosis: external validity of the North American Symptomatic Carotid Endarterectomy Trial measurement method. Radiology. 1997;204:229–233.[Abstract/Free Full Text]

52. Pan XM, Saloner D, Reilly LM, Bowersox JC, Murray SP, Anderson CM, Gooding GA, Rapp JH. Assessment of carotid artery stenosis by ultrasonography, conventional angiography, and magnetic resonance angiography: correlation with ex vivo measurement of plaque stenosis. J Vasc Surg. 1995;21:82–88.[Medline] [Order article via Infotrieve]

53. Prestigiacomo CJ, Connolly ES Jr, Quest DO. Use of carotid ultrasound as a preoperative assessment of extracranial carotid artery blood flow and vascular anatomy. Neurosurg Clin N Am. 1996;7:577–587.[Medline] [Order article via Infotrieve]

54. Obuchowski NA, Modic MT, Magdinec M, Masaryk TJ. Assessment of the efficacy of noninvasive screening for patients with asymptomatic neck bruits. Stroke. 1997;28:1330–1339.[Abstract/Free Full Text]

55. Carriero A, Ucchino S, Magarelli N, Legnini M, Macri MA, Napolitano AM, Bonomo L. Carotid bifurcation stenosis: a comparative study between MR angiography and duplex scanning with respect to digital subtraction angiography. J Neuroradiol. 1995;22:103–111.[Medline] [Order article via Infotrieve]

56. Kuntz KM, Skillman JJ, Whittemore AD, Kent KC. Carotid endarterectomy in asymptomatic patients: is contrast angiography necessary? A morbidity analysis. J Vasc Surg. 1995;22:706–714.[Medline] [Order article via Infotrieve]

57. Turnipseed WD, Kennell TW, Turski PA, Acher CW, Hoch JR. Combined use of duplex imaging and magnetic resonance angiography for evaluation of patients with symptomatic ipsilateral high-grade carotid stenosis. J Vasc Surg. 1993;17:832–839; discussion 839–840.[Medline] [Order article via Infotrieve]

58. Polak JF, Kalina P, Donaldson MC, O’Leary DH, Whittemore AD, Mannick JA. Carotid endarterectomy: preoperative evaluation of candidates with combined Doppler sonography and MR angiography. Radiology. 1993;186:333–338.[Abstract/Free Full Text]

59. Culebras A, Kase CS, Masdeu JC, Fox AJ, Bryan RN, Grossman CB, Lee DH, Adams HP, Thies W. Practice guidelines for the use of imaging in transient ischemic attacks and acute stroke: a report of the Stroke Council, American Heart Association. Stroke. 1997;28:1480–1497.[Free Full Text]

60. Kotval PS. Doppler waveform parvus and tardus: a sign of proximal flow obstruction. J Ultrasound Med. 1989;8:435–440.[Abstract]

61. Kaufman JA, Yucel EK, Waltman AC, Geller SC, Prince MR, Cambria RP, Brewster DC, Abbott WM. MR angiography in the preoperative evaluation of abdominal aortic aneurysms: a preliminary study. J Vasc Interv Radiol. 1994;5:489–496.[Medline] [Order article via Infotrieve]

62. Durham JR, Hackworth CA, Tober JC, Bova JG, Bennett WF, Schmalbrock P, Van Aman ME, Horowitz JD, Wright JG, Smead WL. Magnetic resonance angiography in the preoperative evaluation of abdominal aortic aneurysms. Am J Surg. 1993;166:173–177; discussion 177–178.[Medline] [Order article via Infotrieve]

63. Ecklund K, Hartnell GG, Hughes LA, Stokes KR, Finn JP. MR angiography as the sole method in evaluating abdominal aortic aneurysms: correlation with conventional techniques and surgery. Radiology. 1994;192:345–350.[Abstract/Free Full Text]

64. Prince MR, Narasimham DL, Stanley JC, Wakefield TW, Messina LM, Zelenock GB, Jacoby WT, Marx MV, Williams DM, Cho KJ. Gadolinium-enhanced magnetic resonance angiography of abdominal aortic aneurysms. J Vasc Surg. 1995;21:656–669.[Medline] [Order article via Infotrieve]

65. Laissy JP, Soyer P, Tebboune D, Tiah D, Hvass U, Menu Y. Abdominal aortic aneurysms: assessment with gadolinium-enhanced time-of-flight coronal MR angiography (MRA). Eur J Radiol. 1995;20:1–8.[Medline] [Order article via Infotrieve]

66. Kaufman JA, Geller SC, Petersen MJ, Cambria RP, Prince MR, Waltman AC. MR imaging (including MR angiography) of abdominal aortic aneurysms: comparison with conventional angiography. AJR Am J Roentgenol. 1994;163:203–210.[Abstract/Free Full Text]

67. Petersen MJ, Cambria RP, Kaufman JA, LaMuraglia GM, Gertler JP, Brewster DC, Geller SC, Waltman AC, L’Italien GJ, Abbott WM. Magnetic resonance angiography in the preoperative evaluation of abdominal aortic aneurysms. J Vasc Surg. 1995;21:891–898; discussion 899.[Medline] [Order article via Infotrieve]

68. Shetty AN, Shirkhoda A, Bis KG, Alcantara A. Contrast-enhanced three-dimensional MR angiography in a single breath-hold: a novel technique. AJR Am J Roentgenol. 1995;165:1290–1292.[Free Full Text]

69. Prince MR, Narasimham DL, Stanley JC, Chenevert TL, Williams DM, Marx MV, Cho KJ. Breath-hold gadolinium-enhanced MR angiography of the abdominal aorta and its major branches. Radiology. 1995;197:785–792.[Abstract/Free Full Text]

70. Leung DA, McKinnon GC, Davis CP, Pfammatter T, Krestin GP, Debatin JF. Breath-hold, contrast-enhanced, three-dimensional MR angiography. Radiology. 1996;200:569–571.[Abstract/Free Full Text]

71. Nienaber CA, von Kokolitsch Y, Nicolas V, Siglow V, Piepho A, Brockhoff C, Koschyk DH, Spielmann RP. The diagnosis of thoracic aortic dissection by noninvasive imaging procedures. N Engl J Med. 1993;328:1–9.[Abstract/Free Full Text]

72. Yucel EK, Steinberg FL, Egglin TK, Geller SC, Waltman AC, Athanasoulis CA. Penetrating aortic ulcers: diagnosis with MR imaging. Radiology. 1990;177:779–781.[Abstract/Free Full Text]

73. Laissy JP, Blanc F, Soyer P, Assayag P, Sibert A, Tebboune D, Arrive L, Brochet E, Hvass U, Langlois J, Menu Y. Thoracic aortic dissection: diagnosis with transesophageal echocardiography versus MR imaging. Radiology. 1995;194:331–336.[Abstract/Free Full Text]

74. Di Cesare E, Costanzi A, Fedele F, Di Renzi P, D’Eusanio G, Lupattelli L, Passariello R. MRI postoperative monitoring in patients surgically treated for aortic dissection. Magn Reson Imaging. 1996;14:1149–1156.[Medline] [Order article via Infotrieve]

75. Murray JG, Manisali M, Flamm SD, VanDyke CW, Lieber ML, Lytle BW, White RD. Intramural hematoma of the thoracic aorta: MR image findings and their prognostic implications. Radiology. 1997;204:349–355.[Abstract/Free Full Text]

76. Hartnell GG, Finn JP, Zenni M, Cohen MC, Dupuy DE, Wheeler HG, Longmaid HE III. MR imaging of the thoracic aorta: comparison of spin-echo, angiographic, and breath-hold techniques. Radiology. 1994;191:697–704.[Abstract/Free Full Text]

77. Prince MR, Narasimham DL, Jacoby WT, Williams DM, Cho KJ, Marx MV, Deeb GM. Three-dimensional gadolinium-enhanced MR angiography of the thoracic aorta. AJR Am J Roentgenol. 1996;166:1387–1397.[Abstract/Free Full Text]

78. Kim D, Edelman R, Kent K, Porter D, Stillman J. Abdominal aorta and renal artery stenosis: evaluation with MR angiography. Radiology. 1990;174:727–731.[Abstract/Free Full Text]

79. Siegelman ES, Gilfeather M, Holland GA, Carpenter JP, Golden MA, Townsend RR, Schnall MD. Breath-hold ultrafast three-dimensional gadolinium-enhanced MR angiography of the renovascular system. AJR Am J Roentgenol. 1997;168:1035–1040.[Free Full Text]

80. Hany TF, McKinnon GC, Leung DA, Pfammatter T, Debatin JF. Optimization of contrast timing for breath-hold three-dimensional MR angiography. J Magn Reson Imaging. 1997;7:551–556.[Medline] [Order article via Infotrieve]

81. Yucel EK, Kaufman JA, Prince M, Bazari H, Fang LS, Waltman AC. Time of flight renal MR angiography: utility in patients with renal insufficiency. Magn Reson Imaging. 1993;11:925–930.[Medline] [Order article via Infotrieve]

82. Kent KC, Edelman RR, Kim D, Steinman TI, Porter DH, Skillman JJ. Magnetic resonance imaging: a reliable test for the evaluation of proximal atherosclerotic renal arterial stenosis. J Vasc Surg. 1991;13:311–318.[Medline] [Order article via Infotrieve]

83. Debatin JF, Spritzer CE, Grist TM, Beam C, Svetkey LP, Newman GE, Sostman HD. Imaging of the renal arteries: value of MR angiography. AJR Am J Roentgenol. 1991;157:981–990.[Abstract/Free Full Text]

84. Strotzer M, Fellner CM, Geissler A, Gmeinwieser J, Kohler SM, Kramer BK, Kromer EP. Noninvasive assessment of renal artery stenosis: a comparison of MR angiography, color Doppler sonography, and intraarterial angiography. Acta Radiol. 1995;36:243–247.[Medline] [Order article via Infotrieve]

85. Servois V, Laissy JP, Feger C, Sibert A, Delahousse M, Baleynaud S, Mery JP, Menu Y. Two-dimensional time-of-flight magnetic resonance angiography of renal arteries without maximum intensity projection: a prospective comparison with angiography in 21 patients screened for renovascular hypertension. Cardiovasc Intervent Radiol. 1994;17:138–142.[Medline] [Order article via Infotrieve]

86. Hertz SM, Holland GA, Baum RA, Haskal ZJ, Carpenter JP. Evaluation of renal artery stenosis by magnetic resonance angiography. Am J Surg. 1994;168:140–143.[Medline] [Order article via Infotrieve]

87. Fellner C, Strotzer M, Geissler A, Kohler SM, Kramer BK, Spies V, Held P, Gmeinwieser J. Renal arteries: evaluation with optimized 2D and 3D time-of flight MR angiography. Radiology. 1995;196:681–687.[Abstract/Free Full Text]

88. Laissy JP, Shouman-Claeys E. Time-of-flight magnetic resonance angiography of aorta and renal arteries: state of the art. Eur J Radiol. 1996;21:201–211.[Medline] [Order article via Infotrieve]

89. Duda SH, Schick F, Teufl F, Muller-Schimpfle M, Erley C, Schneider W, Miller S, Claussen CD. Phase-contrast MR angiography for detection of arteriosclerotic renal artery stenosis. Acta Radiol. 1997;38:287–291.[Medline] [Order article via Infotrieve]

90. Lewin JS, Laub G, Hausmann R. Three-dimensional time-of-flight MR angiography: applications in the abdomen and thorax. Radiology. 1991;179:261–264.[Abstract/Free Full Text]

91. Vock P, Terrier F, Wegmuller H, Mahler F, Gertsch P, Souza SP, Dumoulin CL. Magnetic resonance angiography of abdominal vessels: early experience using the three-dimensional phase-contrast technique. Br J Radiol. 1991;64:10–16.[Abstract/Free Full Text]

92. Grist TM. Magnetic resonance angiography of renal artery stenosis. Am J Kidney Dis. 1994;24:700–712.[Medline] [Order article via Infotrieve]

93. Gedroyc WM, Negus R, al-Kutoubi A, Palmer A, Taube D, Hulme B. Magnetic resonance angiography of renal transplants. Lancet. 1992;339:789–791.[Medline] [Order article via Infotrieve]

94. Loubeyre P, Cahen R, Grozel F, Trolliet P, Pouteil-Noble C, Labeeuw M, Tran Minh VA. Transplant renal artery stenosis: evaluation of diagnosis with magnetic resonance angiography compared with color duplex sonography and arteriography. Transplantation. 1996;62:446–450.[Medline] [Order article via Infotrieve]

95. Loubeyre P, Trolliet P, Cahen R, Grozel F, Labeeuw M, Minh VA. MR angiography of renal artery stenosis: value of the combination of three-dimensional time-of-flight and three-dimensional phase contrast MR angiography sequences. AJR Am J Roentgenol. 1996;167:489–494.[Abstract/Free Full Text]

96. De Cobelli F, Mellone R, Salvioni M, Vanzulli A, Sironi S, Manunta P, Lanzani C, Bianchi G, Del Maschio A. Renal artery stenosis: value of screening with three-dimensional phase-contrast MR angiography with a phased array multicoil. Radiology. 1996;201:697–703.[Abstract/Free Full Text]

97. de Haan MW, Kouwenhoven M, Thelissen RP, Koster D, Kessels AG, de Leeuw PW, van Engelshoven JM. Renovascular disease in patients with hypertension: detection with systolic and diastolic gating in three-dimensional, phase-contrast MR angiography. Radiology. 1996;198:449–456.[Abstract/Free Full Text]

98. Silverman JM, Freidman ML, Van Allan RJ. Detection of main renal artery stenosis using phase-contrast cine MR angiography. AJR Am J Roentgenol. 1996;166:1131–1137.[Abstract/Free Full Text]

99. Schoenberg SO, Knopp MV, Bock M, Kallinowski F, Just A, Essig M, Hawighorst H, Schad L, van Kaick G. Renal artery stenosis: grading of hemodynamic changes with cine phase-contrast MR blood flow measurements. Radiology. 1997;203:45–53.[Abstract/Free Full Text]

100. Wasser MN, Westenberg J, van der Hulst VP, van Baalen J, van Bockel JH, van Erkel AR, Pattynama PM. Hemodynamic significance of renal artery stenosis: digital subtraction angiography versus systolically gated three-dimensional phase-contrast MR angiography. Radiology. 1997;202:333–338.[Abstract/Free Full Text]

101. Gedroyc WM, Neerhut P, Negus R, Palmer A, al Kutoubi A, Taube D, Hulme B. Magnetic resonance angiography of renal artery stenosis. Clin Radiol. 1995;50:436–439.[Medline] [Order article via Infotrieve]

102. Prince MR, Schoenberg SO, Ward JS, Londy FJ, Wakefield TW, Stanley JC. Hemodynamically significant atherosclerotic renal artery stenosis: MR angiographic features. Radiology. 1997;205:128–136.[Abstract/Free Full Text]

103. Deleted in proof.

104. Holland GA, Dougherty L, Carpenter JP, Golden MA, Gilfeather M, Slossman F, Schnall MD, Axel L. Breath-hold ultrafast three-dimensional gadolinium-enhanced MR angiography of the aorta and the renal and other visceral abdominal arteries. AJR Am J Roentgenol. 1996;166:971–981.[Abstract/Free Full Text]

105. Holland GA. Author’s correction. AJR Am J Roentgenol. 1996;166:541.[Free Full Text]

106. Snidow JJ, Johnson MS, Harris VJ, Margosian PM, Aisen AM, Lalka SG, Cikrit DF, Trerotola SO. Three-dimensional gadolinium-enhanced MR angiography for aortoiliac inflow assessment plus renal artery screening in a single breath hold. Radiology. 1996;198:725–732.[Abstract/Free Full Text]

107. Steffens JC, Link J, Grassner J, Mueller-Huelsbeck S, Brinkmann G, Reuter M, Heller M. Contrast-enhanced, K-space-centered, breath-hold MR angiography of the renal arteries and the abdominal aorta. J Magn Reson Imaging. 1997;7:617–622.[Medline] [Order article via Infotrieve]

108. Rieumont MJ, Kaufman JA, Geller SC, Yucel EK, Cambria RP, Fang LS, Bazari H, Waltman AC. Evaluation of renal artery stenosis with dynamic gadolinium-enhanced MR angiography. AJR Am J Roentgenol. 1997;169:39–44.[Abstract/Free Full Text]

109. Hany TF, Debatin JF, Leung DA, Pfammater T. Evaluation of the aortoiliac and renal arteries: comparison of breath hold, contrast enhanced three dimensional MR angiography with conventional catheter angiography. Radiology. 1997;204:357–362.[Abstract/Free Full Text]

110. Baumgartner I, Koch M, Maier S, Franzeck UK, von Schulthess GK, Bollinger A. Duplex sonography, magnetic resonance arteriography and conventional arteriography in the evaluation of peripheral arterial occlusive disease [in German]. Vasa Suppl. 1991;33:303.[Medline] [Order article via Infotrieve]

111. Yucel EK. Magnetic resonance angiography of the lower extremity and renal arteries. Semin Ultrasound CT MR. 1992;13:291–302.[Medline] [Order article via Infotrieve]

112. Lossef SV, Rajan SS, Patt RH, Carvlin M, Calcagno D, Gomes MN, Barth KH. Gadolinium-enhanced magnitude contrast MR angiography of popliteal and tibial arteries. Radiology. 1992;184:349–355.[Abstract/Free Full Text]

113. Higgins CB. The potential role of magnetic resonance imaging in ischemic vascular disease. N Engl J Med. 1992;326:1624–1626.[Medline] [Order article via Infotrieve]

114. Harms SE, Flamig DP. Magnetic resonance angiography: application to the peripheral circulation. Invest Radiol. 1992;27:S80–S83.

115. Carpenter JP. Magnetic resonance angiography in peripheral artery disease. Hosp Pract (Off Ed). 1992;27:79–82, 85–76, 92 passim.

116. Owen RS, Baum RA, Carpenter JP, Holland GA, Cope C. Symptomatic peripheral vascular disease: selection of imaging parameters and clinical evaluation with MR angiography. Radiology. 1993;187:627–635.[Abstract/Free Full Text]

117. Schiebler ML, Listerud J, Holland G, Owen R, Baum R, Kressel HY. Magnetic resonance angiography of the pelvis and lower extremities: works in progress. Invest Radiol. 1992;27:S90–S96.

118. Yucel EK. Magnetic resonance angiography of the peripheral arteries. Magn Reson Imaging Clin N Am. 1993;1:229–238.[Medline] [Order article via Infotrieve]

119. Mulligan SA, Matsuda T, Lanzer P, Gross GM, Routh WD, Keller FS, Koslin DB, Berland LL, Fields MD, Doyle M, Cranney GB, Lee JY, Pohost GM. Peripheral arterial occlusive disease: prospective comparison of MR angiography and color duplex US with conventional angiography. Radiology. 1991;178:695–700.[Abstract/Free Full Text]

120. Yucel EK, Dumoulin CL, Waltman AC. MR angiography of lower extremity arterial disease: preliminary experience. J Magn Reson Imaging. 1992;2:303–309.[Medline] [Order article via Infotrieve]

121. Hertz SM, Baum RA, Owen RS, Holland GA, Logan DR, Carpenter JP. Comparison of magnetic resonance angiography and contrast arteriography in peripheral arterial stenosis. Am J Surg. 1993;166:112–116; discussion 116.[Medline] [Order article via Infotrieve]

122. Cambria RP, Yucel EK, Brewster DC, L’Italien G, Gertler JP, LaMuraglia GM, Kaufman JA, Waltman AC, Abbott WM. The potential for lower extremity revascularization without contrast arteriography: experience with magnetic resonance angiography. J Vasc Surg. 1993;17:1050–1056; discussion 1056–1057.[Medline] [Order article via Infotrieve]

123. Baumgartner I, Maier SE, Koch M, Schneider E, von Schulthess GK, Bollinger A. Magnetic resonance arteriography, duplex sonography and conventional arteriography for the evaluation of peripheral arterial occlusive disease [in German]. Rofo Fortschr Geb Rontgenstr Neuen Bildgeb Verfahr. 1993;159:167–173.[Medline] [Order article via Infotrieve]

124. Owen RS, Carpenter JP, Baum RA, Perloff LJ, Cope C. Magnetic resonance imaging of angiographically occult runoff vessels in peripheral arterial occlusive disease. N Engl J Med. 1992;326:1577–1581.[Abstract]

125. Druy EM. Magnetic resonance versus conventional angiography in peripheral arterial occlusive disease. N Engl J Med. 1992;327:1319; discussion 1320.[Medline] [Order article via Infotrieve]

126. Picus D, Hicks ME, Darcy MD, Vesely TM. Magnetic resonance imaging of angiographically occult run off vessels in peripheral arterial occlusive disease. Invest Radiol. 1993;28:656–658.[Medline] [Order article via Infotrieve]

127. Kaufman SL. Magnetic resonance versus conventional angiography in peripheral arterial occlusive disease. N Engl J Med. 1992;327:1319–1320.

128. Carpenter JP, Owen RS, Baum RA, Cope C, Barker CF, Berkowitz HD, Golden MA, Perloff LJ. Magnetic resonance angiography of peripheral runoff vessels. J Vasc Surg. 1992;16:807–813; discussion 813–815.[Medline] [Order article via Infotrieve]

129. McCauley TR, Monib A, Dickey KW, Clemett J, Meier GH, Egglin TK, Gusberg RJ, Rosenblatt M, Pollak JS. Peripheral vascular occlusive disease: accuracy and reliability of time-of-flight MR angiography. Radiology. 1994;192:351–357.[Abstract/Free Full Text]

130. McDermott VG, Meakem TJ, Carpenter JP, Baum RA, Stolpen AH, Holland GA, Schnall MD. Magnetic resonance angiography of the distal lower extremity. Clin Radiol. 1995;50:741–746.[Medline] [Order article via Infotrieve]

131. Huber TS, Back MR, Ballinger RJ, Culp WC, Flynn TC, Kubilis PS, Seeger JM. Utility of magnetic resonance arteriography for distal lower extremity revascularization. J Vasc Surg. 1997;26:415–423.[Medline] [Order article via Infotrieve]

132. Baum RA, Rutter CM, Sunshine JH, Blebea JS, Blebea J, Carpenter JP, Dickey KW, Quinn SF, Gomes AS, Grist TM, McNeil BJ. Multicenter trial to evaluate vascular magnetic resonance angiography of the lower extremity. American College of Radiology Rapid Technology Assessment Group. JAMA. 1995;274:875–880.[Abstract/Free Full Text]

133. Yin D, Baum RA, Carpenter JP, Langlotz CP, Pentecost MJ. Cost-effectiveness of MR angiography in cases of limb-threatening peripheral vascular disease. Radiology. 1995;194:757–764.[Abstract/Free Full Text]

134. Snidow JJ, Harris VJ, Johnson MS, Cikrit DF, Lalka SG, Sawchuk AP, Trerotola SO. Iliac artery evaluation with two-dimensional time-of-flight MR angiography: update. J Vasc Interv Radiol. 1996;7:213–220.[Medline] [Order article via Infotrieve]

135. Snidow JJ, Harris VJ, Trerotola SO, Cikrit DF, Lalka SG, Buckwalter KA, Johnson MS. Interpretations and treatment decisions based on MR angiography versus conventional arteriography in symptomatic lower extremity ischemia. J Vasc Interv Radiol. 1995;6:595–603.[Medline] [Order article via Infotrieve]

135. Carpenter JP, Owen RS, Holland GA, Baum RA, Barker CF, Perloff LJ, Golden MA, Cope C. Magnetic resonance angiography of the aorta, iliac, and femoral arteries. Surgery.. 1994;116:17–23.[Medline] [Order article via Infotrieve]

136. Mulligan SA, Doyle M, Matsuda T, Koslin DB, Kenney PJ, Barton RE, Pohost GM. Aortoiliac disease: two-dimensional inflow MR angiography with lipid suppression. J Magn Reson Imaging. 1993;6:829–834.

137. Sivananthan UM, Ridgway JP, Bann K, Verma SP, Cullingworth J, Ward J, Rees MR. Fast magnetic resonance angiography using turbo-Flash sequences in advanced aortoiliac disease. Br J Radiol. 1993;66:1103–1110.[Abstract/Free Full Text]

138. Swan JS, Kennel TW, Wojtowycz MM, Grist TM. Increased presaturation pulse gaps in two-dimensional time-of-flight MR angiography: a pitfall in diseased lower extremities. J Vasc Interv Radiol. 1993;4:569–571.[Medline] [Order article via Infotrieve]

139. Swan JS, Korosec FR, Heisey DM, Grist TM. Evaluation of presaturation pulse thickness and gap for lower extremity peripheral two-dimensional time-of-flight MR angiography with a signa 1.5-T unit. J Vasc Interv Radiol. 1995;6:965–970.[Medline] [Order article via Infotrieve]

140. Ekelund L, Sjoqvist L, Thuomas KA, Asberg B. MR angiography of abdominal and peripheral arteries: techniques and clinical applications. Acta Radiol. 1996;37:3–13.[Medline] [Order article via Infotrieve]

141. Parker DL, Buswell HR, Goodrich KC, Alexander AL, Keck N, Tsuruda JS. The application of magnetization transfer to MR angiography with reduced total power. Magn Reson Med. 1995;34:283–286.[Medline] [Order article via Infotrieve]

142. Debatin JF, Leung DA, Wildermuth S, Holtz D, McKinnon GC. Advances in vascular echoplanar imaging. Cardiovasc Intervent Radiol. 1995;18:277–287.[Medline] [Order article via Infotrieve]

143. Glickerman DJ, Obregon RG, Schmiedl UP, Harrison SD, Macaulay SE, Simon HE, Kohler TR. Cardiac-gated MR angiography of the entire lower extremity: a prospective comparison with conventional angiography. AJR Am J Roentgenol. 1996;167:445–451.[Abstract/Free Full Text]

144. Adamis MK, Li W, Wielopolski PA, Kim D, Sax EJ, Kent KC, Edelman RR. Dynamic contrast-enhanced subtraction MR angiography of the lower extremities: initial evaluation with a multisection two-dimensional time-of-flight sequence. Radiology. 1995;196:689–695.[Abstract/Free Full Text]

145. Leyendecker JR, Johnson SP, Diffin DC, Elsass K, Bifano SL. Time-of-flight MR arteriography of below-knee arteries with maximum-intensity-projection reconstruction: is interpretation of the axial source images helpful? AJR Am J Roentgenol. 1997;169:1145–1149.[Abstract/Free Full Text]

146. Nielsen HT, Olcott EW, Nishimura DG. Improved 2D time-of-flight angiography using a radial-line k-space acquisition. Magn Resonance Med. 1997;37:285–291.

147. Alley MT, Grist TM, Swan JS. Development of a phased-array coil for the lower extremities. Magn Reson Med. 1995;34:260–267.[Medline] [Order article via Infotrieve]

148. Kojima KY, Szumowski J, Sheley RC, Quinn SF. Lower extremities: MR angiography with a unilateral telescopic phased-array coil. Radiology. 1995;196:871–875.[Abstract/Free Full Text]

149. Prince MR, Yucel EK, Kaufman JA, Harrison DC, Geller SC. Dynamic gadolinium-enhanced three-dimensional abdominal MR arteriography. J Magn Reson Imaging. 1993;3:877–881.[Medline] [Order article via Infotrieve]

150. Rofsky NM, Johnson G, Adelman MA, Rosen RJ, Krinsky GA, Weinreb JC. Peripheral vascular disease evaluated with reduced-dose gadolinium-enhanced MR angiography. Radiology. 1997;205:163–169.[Abstract/Free Full Text]

151. Poon E, Yucel EK, Pagan-Marin H, Kayne H. Iliac artery stenosis measurements: comparison of two-dimensional time-of-flight and three-dimensional dynamic gadolinium-enhanced MR angiography. AJR Am J Roentgenol.. 1997;169:1139–1144.[Abstract/Free Full Text]

152. Wang Y, Lee HM, Khilnani NM, Trost DW, Jagust MB, Winchester PA, Bush HL, Sos TA, Sostman HD. Bolus-chase MR digital subtraction angiography in the lower extremity. Radiology. 1998;207:263–269.[Abstract/Free Full Text]

153. Meaney JF, Ridgway JP, Chakraverty S, Robertson I, Kessel D, Radjenovic A, Kouwenhoven M, Kassner A, Smith MA. Stepping-table gadolinium-enhanced digital subtraction MR angiography of the aorta and lower extremity arteries: preliminary experience. Radiology. 1999;211:59–67.[Abstract/Free Full Text]

154. Wang Y, Johnston DL, Breen JF. Huston J III, Jack CR, Julsrud PR, Kiely MJ, King BF, Riederer SL, Ehman RL. Dynamic MR digital subtraction angiography using contrast enhancement, fast data acquisition, and complex subtraction. Magn Reson Med. 1996;36:551–556.[Medline] [Order article via Infotrieve]

155. Watanabe Y, Dohke M, Okumura A, Amoh Y, Ishimori T, Oda K, Dodo Y. Dynamic subtraction MR angiography: first-pass imaging of the main arteries of the lower body. AJR Am J Roentgenol.. 1998;170:357–360.[Free Full Text]

156. Quinn SF, Demlow TA, Hallin RW, Eidemiller LR, Szumowski J. Femoral MR angiography versus conventional angiography: preliminary results. Radiology. 1993;189:181–184.[Abstract/Free Full Text]

157. Hoch JR, Tullis MJ, Kennell TW, McDermott J, Acher CW, Turnipseed WD. Use of magnetic resonance angiography for the preoperative evaluation of patients with infrainguinal arterial occlusive disease. J Vasc Surg. 1996;23:792–800; discussion 801.[Medline] [Order article via Infotrieve]

158. Cambria RP, Kaufman JA, L’Italien GJ, Gertler JP, LaMuraglia GM, Brewster DC, Geller S, Atamian S, Waltman AC, Abbott WM. Magnetic resonance angiography in the management of lower extremity arterial occlusive disease: a prospective study. J Vasc Surg. 1997;25:380–389.[Medline] [Order article via Infotrieve]

159. Carpenter JP, Baum RA, Holland GA, Barker CF. Peripheral vascular surgery with magnetic resonance angiography as the sole preoperative imaging modality. J Vasc Surg. 1994;20:861–869; discussion 869–871.[Medline] [Order article via Infotrieve]

160. Di Cesare E, Marsili L, Marino G, Masciocchi C, Morettini G, Spartera C, Lupattelli L, Passariello R. Stress MR imaging for evaluation of popliteal artery entrapment. J Magn Reson Imaging. 1994;4:617–622.[Medline] [Order article via Infotrieve]

161. Fujiwara H, Sugano T, Fujii N. Popliteal artery entrapment syndrome: accurate morphological diagnosis utilizing MRI. J Cardiovasc Surg (Torino). 1992;33:160–162.[Medline] [Order article via Infotrieve]

162. Porcellini M, Selvetella L, Bernardo B, Del Viscovo L, Parisi B, Baldassarre M. Popliteal artery entrapment syndrome: diagnosis and management. G Chir. 1997;18:182–186.[Medline] [Order article via Infotrieve]

163. Huch Boni RA, Brunner U, Bollinger A, Debatin JF, Hauser M, Krestin GP. Management of congenital angiodysplasia of the lower limb: magnetic resonance imaging and angiography versus conventional angiography. Br J Radiol. 1995;68:1308–1315.[Abstract/Free Full Text]

164. Robbins KV, Meadors FA, Koonce TW, Dedman WD, King MT. Radiological case of the month: cystic adventitial disease of the popliteal artery. J Ark Med Soc. 1997;94:129–130.[Medline] [Order article via Infotrieve]

165. Rak KM, Yakes WF, Ray RL, Dreisbach JN, Parker SH, Luethke JM, Stavros AT, Slater DD, Burke BJ. MR imaging of symptomatic peripheral vascular malformations. AJR Am J Roentgenol. 1992;159:107–112.[Abstract/Free Full Text]

166. Futran ND, Stack BC Jr, Payne LP. Use of color Doppler flow imaging for preoperative assessment in fibular osteoseptocutaneous free tissue transfer. Otolaryngol Head Neck Surg. 1997;117:660–663.[Medline] [Order article via Infotrieve]

167. Dodge JT Jr, Brown BG, Bolson EL, Dodge HT. Lumen diameter of normal human coronary arteries: influence of age, sex, anatomic variation, and left ventricular hypertrophy or dilation. Circulation. 1992;86:232–246.[Abstract/Free Full Text]

168. Wang Y, Riederer SJ, Ehman RL. Respiratory motion of the heart: kinematics and the implications for the spatial resolution in coronary imaging. Magn Reson Med. 1995;33:713–719.[Medline] [Order article via Infotrieve]

169. McDonald IG. The shape and movements of the human left ventricle during systole: a study by cineangiography and by cineradiography of epicardial markers. Am J Cardiol. 1970;26:221–230.[Medline] [Order article via Infotrieve]

170. Lieberman JM, Botti RE, Nelson AD. Magnetic resonance imaging of the heart. Radiol Clin North Am. 1984;22:847–858.[Medline] [Order article via Infotrieve]

171. Burstein D. MR imaging of coronary artery flow in isolated and in vivo hearts. J Magn Reson Imaging. 1991;1:337–346.[Medline] [Order article via Infotrieve]

172. Li D, Paschal CB, Haacke EM, Adler LP. Coronary arteries: three dimensional MR imaging with fat saturation and magnetization transfer contrast. Radiology. 1993;187:401–406.[Abstract/Free Full Text]

173. Wang Y, Grist TM, Korosec FR, Christy PS, Alley MT, Polzin JA, Mistretta CA. Respiratory blur in 3D coronary MR imaging. Magn Reson Med. 1995;33:541–548.[Medline] [Order article via Infotrieve]

174. Wang SJ, Hu BS, Macovski A, Nishimura DG. Coronary angiography using fast selective inversion recovery. Magn Reson Med. 1991;18:417–423.[Medline] [Order article via Infotrieve]

175. Meyer CH, Hu BS, Nishimura DG, Macovski A. Fast spiral coronary artery imaging. Magn Reson Med. 1992;28:202–213.[Medline] [Order article via Infotrieve]

176. Bornert P, Jensen D. Coronary artery imaging at 0.5T using echo planar imaging. Magn Reson Med. 1995;34:779–785.[Medline] [Order article via Infotrieve]

177. Wielopolski PA, Manning WJ, Edelman RR. Single breath-hold volumetric imaging of the heart using magnetization prepared 3-dimensional segmented echo planar imaging. J Magn Reson Imaging. 1995;5:403–409.[Medline] [Order article via Infotrieve]

178. Wielopolski P, de Feyter P. Single breath-hold three-dimensional MR coronary angiography. Proc Int Soc Magn Reson Med. 1996;1:451. Abstract.

179. Hofman MB, Paschal CB, Li D, Haacke EM, van Rossum AC, Sprenger M. MRI of coronary arteries: 2D breath hold vs 3D respiratory-gated acquisition. J Comput Assist Tomogr. 1995;19:56–62.[Medline] [Order article via Infotrieve]

180. Manning WJ, Li W, Boyle NG, Edelman RR. Fat-suppressed breath-hold magnetic resonance coronary angiography. Circulation. 1993;87:94–104.[Abstract/Free Full Text]

181. Oshinski JN, Hofland L, Mukundan S Jr, Dixon WT, Parks WJ, Pettigrew RI. Two-dimensional coronary MR angiography without breath holding. Radiology. 1996;201:737–743.[Abstract/Free Full Text]

182. Pennell DJ, Keegan J, Firmin DN, Gatehouse PD, Underwood SR, Longmore DB. Magnetic resonance imaging of coronary arteries: technique and preliminary results. Br Heart J. 1993;70:315–326.[Abstract/Free Full Text]

183. Goldfarb JW, Li W, Griswold MA. Contrast enhanced breath-hold 3D coronary magnetic resonance angiography. Proc Inst Soc Magn Reson Med. 1997;1:441. Abstract.

184. Li D, Kaushikkar S, Haacke EM, Woodard PK, Dhawale PJ, Kroeker RM, Laub G, Kuginuki Y, Gutierrez FR. Coronary arteries: three dimensional MR imaging with retrospective respiratory gating. Radiology. 1996;201:857–863.[Abstract/Free Full Text]

185. Paschal CB, Haacke EM, Adler LP. Three-dimensional MR imaging of the coronary arteries: preliminary clinical experience. J Magn Reson Imaging. 1993;3:491–500.[Medline] [Order article via Infotrieve]

186. Post JC, van Rossum AC, Hofman MB, Valk J, Visser CA. Three-dimensional respiratory-gated MR angiography of coronary arteries: comparison with conventional coronary angiography. AJR Am J Roentgenol. 1996;166:1399–1404.[Abstract/Free Full Text]

187. Scheidegger MB, Vassalli G, Hess OM, Boesiger P. Validation of coronary artery MR angiography: comparison of measure vessel diameters with quantitative contrast angiography. Proc Soc Magn Reson. 1994;1:497. Abstract.

188. Kimbiris D, Iskandrian AS, Segal BL, Bemis CE. Anomalous aortic origin of coronary arteries. Circulation. 1978;58:606–615.[Free Full Text]

189. Cheitlin MD, DeCastro CM, McAllister HA. Sudden death as a complication of anomalous left coronary origin from the anterior sinus of Valsalva, a not so minor congenital anomaly. Circulation. 1974;50:780–787.[Abstract/Free Full Text]

190. Post JC, van Rossum AC, Bronzwaer JG, de Cock CC, Hofman MB, Valk J, Visser CA. Magnetic resonance angiography of anomalous coronary arteries: a new gold standard for delineating the proximal course? Circulation. 1995;92:3163–3171.[Abstract/Free Full Text]

191. McConnell MV, Ganz P, Selwyn AP, Li W, Edelman RR, Manning WJ. Identification of anomalous coronary arteries and their anatomic course by magnetic resonance coronary angiography. Circulation. 1995;92:3158–3162.[Abstract/Free Full Text]

192. Vliegen HW, Doornbos J, de Roos A, Jukema JW, Bekedam MA, van der Wall EE. Value of fast gradient echo magnetic resonance angiography as an adjunct to coronary arteriography in detecting and confirming the course of clinically significant coronary artery anomalies. Am J Cardiol. 1997;79:773–776.[Medline] [Order article via Infotrieve]

193. Manning WJ, Li W, Edelman RR. A preliminary report comparing magnetic resonance coronary angiography with conventional angiography. N Engl J Med. 1993;328:828–832.[Abstract/Free Full Text]

194. Pennell DJ, Bogren HG, Keegan J, Firmin DN, Underwood SR. Assessment of coronary artery stenosis by magnetic resonance imaging. Heart. 1996;75:127–133.[Abstract/Free Full Text]

195. Natatori T, Hachiya J, Korenaga T, Hanaoka H, Yoshino A. Clinical application of coronary MR angiography: studies on shortening of time for examination. Proc Soc Magn Reson. 1995:1390. Abstract.

196. Post JC, van Rossum AC, Hofman MB, de Cock CC, Valk J, Visser CA. Clinical utility of two-dimensional magnetic resonance angiography in detecting coronary artery disease. Eur Heart J. 1997;18:426–433.[Abstract/Free Full Text]

197. Muller MF, Fleisch M, Kroeker R, Chatterjee T, Meier B, Vock P. Proximal coronary artery stenosis: three-dimensional MRI with fat saturation and navigator echo. J Magn Reson Imaging. 1997;7:644–651.[Medline] [Order article via Infotrieve]

198. Hu BS, Meyer CH, Macovski A, Nishimura DG. Multi-spiral magnetic resonance coronary angiography. Proc Int Soc Magn Reson Med. 1996;1:176. Abstract.

199. Duerinckx AJ, Urman MK. Two-dimensional coronary MR angiography: analysis of initial clinical results. Radiology. 1994;193:731–738.[Abstract/Free Full Text]

200. Scott NA, Pettigrew RI. Absence of movement of coronary stents after placement in a magnetic resonance imaging field. Am J Cardiol. 1994;73:900–901.[Medline] [Order article via Infotrieve]

201. Duerinckx AJ, Atkinson D, Hurwitz R, Mintorovitch J, Whitney W. Coronary MR angiography after coronary stent placement. AJR Am J Roentgenol. 1995;165:662–664.[Free Full Text]

202. Galjee MA, van Rossum AC, Doesburg T, van Eenige MJ, Visser CA. Value of magnetic resonance imaging in assessing patency and function of coronary artery bypass grafts: an angiographically controlled study. Circulation. 1996;93:660–666.[Abstract/Free Full Text]

203. Jenkins JP, Love HG, Foster CJ, Isherwood I, Rowlands DJ. Detection of coronary artery bypass graft patency as assessed by magnetic resonance imaging. Br J Radiol. 1988;61:2–4.[Abstract/Free Full Text]

204. Rubinstein RI, Askenase AD, Thickman D, Feldman MS, Agarwal JB, Helfant RH. Magnetic resonance imaging to evaluate patency of aortocoronary bypass grafts. Circulation. 1987;76:786–791.[Abstract/Free Full Text]

205. White RD, Caputo GR, Mark AS, Modin GW, Higgins CB. Coronary artery bypass graft patency: noninvasive evaluation with MR imaging. Radiology. 1987;164:681–686.[Abstract/Free Full Text]

206. White RD, Pflugfelder PW, Lipton MJ, Higgins CB. Coronary artery bypass grafts: evaluation of patency with cine MR imaging. AJR Am J Roentgenol. 1988;150:1271–1274.[Abstract/Free Full Text]

207. Aurigemma GP, Reichek N, Axel L, Schiebler M, Harris C, Kressel HY. Noninvasive determination of coronary artery bypass graft patency by cine magnetic resonance imaging. Circulation. 1989;80:1595–1602.[Abstract/Free Full Text]

208. Wintersperger BJ, von Smekal A, Engelmann MG. Determination of coronary artery bypass graft (CABG) patency using gadolinium enhanced 3D magnetic resonance angiography: correlation with coronary angiography. Proc Int Soc Magn Reson Med. 1997;1:444. Abstract.

209. Vrachliotis TG, Bis KG, Aliabadi D, Shetty AN, Safian R, Simonetti O. Contrast-enhanced breath-hold MR angiography for evaluating patency of coronary artery bypass grafts. AJR Am J Roentgenol. 1997;168:1073–1080.[Abstract/Free Full Text]

210. Davis CP, Hany TF, Kaufmann P, von Shulthess GK. Coronary artery bypass grafts: 3D MR angiography. In: 5th Annual Proceedings of the International Society of Magnetic Resonance Medicine. 1997;2:834. Abstract.

211. Warner OJ, Ohri SK, Pennell DJ, Smith PL. Magnetic resonance coronary artery imaging for redo coronary operations. Ann Thorac Surg. 1996;62:1513–1516.[Abstract/Free Full Text]

212. McConnell MV, Khasgiwala VC, Savord BJ, Chen MH, Chuang ML, Edelman RR, Manning WJ. Prospective adaptive navigator correction for breath-hold MR coronary angiography. Magn Reson Med. 1997;37:148–152.[Medline] [Order article via Infotrieve]

213. Ehman RL, Felmlee JP. Adaptive technique for high-definition MR imaging of moving structures. Radiology. 1989;173:255–263.[Abstract/Free Full Text]

214. Fu ZW, Wang Y, Grimm RC, Rossman PJ, Felmlee JP, Riederer SJ, Ehman RL. Orbital navigator echoes for motion measurements in magnetic resonance imaging. Magn Reson Med. 1995;34:746–753.[Medline] [Order article via Infotrieve]

215. Sachs TS, Meyer CH, Hu BS, Kohli J, Nishimura DG, Macovski A. Real-time motion detection in spiral MRI using navigators. Magn Reson Med. 1994;32:639–645.[Medline] [Order article via Infotrieve]

216. Taylor AM, Jhooti P, Weissmann F, Keegan J, Firmin DN, Pennell DJ. MR navigator-echo monitoring of temporal changes in diaphragm position: implications for MR coronary angiography. J Magn Reson Imaging. 1997;7:629–636.[Medline] [Order article via Infotrieve]

217. Wang Y, Rossman PJ, Grimm RC, Riederer SJ, Ehman RL. Navigator-echo-based real-time respiratory gating and triggering for reduction of respiration effects in three-dimensional coronary MR angiography. Radiology. 1996;198:55–60.[Abstract/Free Full Text]

218. Meyer CH, Hu BS, Kerr AB. High resolution multislice spiral coronary angiography with real-time interactive localization. Proc Int Soc Magn Reson Med. 1997;1:439. Abstract.

219. Hofman MB, Henson RE, Kovacs SJ, Fischer SE, Lauffer RB, Adzamli K, De Becker J, Wickline SA, Lorenz CH. Blood pool agent strongly improves 3D magnetic resonance coronary angiography using an inversion pre-pulse. Proc Int Soc Magn Reson Med. 1997;1:442.

220. Moody AR, Liddicoat A, Krarup K. Magnetic resonance pulmonary angiography and direct imaging of embolus for the detection of pulmonary emboli. Invest Radiol. 1997;32:431–440.[Medline] [Order article via Infotrieve]

221. Sostman HD, Layish DT, Tapson VF, Spritzer CE, DeLong DM, Trotter P, MacFall JR, Patz EF Jr, Goodman PC, Woodard PK, Foo TK, Farber JL. Prospective comparison of helical CT and MR imaging in clinically suspected acute pulmonary embolism. J Magn Reson Imaging. 1996;6:275–281.[Medline] [Order article via Infotrieve]

222. Schiebler ML, Holland GA, Hatabu H, Listerud J, Foo T, Palevsky H, Edmunds H, Gefter WB. Suspected pulmonary embolism: prospective evaluation with pulmonary MR angiography. Radiology. 1993;189:125–131.[Abstract/Free Full Text]

223. Wielopolski PA, Haacke EM, Adler LP. Three-dimensional MR imaging of the pulmonary vasculature: preliminary experience. Radiology. 1992;183:465–472.[Abstract/Free Full Text]

224. Meaney JF, Weg JG, Chenevert TL, Stafford-Johnson D, Hamilton BH, Prince MR. Diagnosis of pulmonary embolism with magnetic resonance angiography. N Engl J Med. 1997;336:1422–1427.[Abstract/Free Full Text]

225. Steiner P, McKinnon GC, Romanowski B, Goehde SC, Hany T, Debatin JF. Contrast-enhanced, ultrafast 3D pulmonary MR angiography in a single breath-hold: initial assessment of imaging performance. J Magn Reson Imaging. 1997;7:177–182.[Medline] [Order article via Infotrieve]

226. Ahlstrom KH, Johansson LO, Rodenburg JB, Ragnarsson AS, Akeson P, Borseth A. Pulmonary MR angiography with ultrasmall superparamagnetic iron oxide particles as a blood pool agent and a navigator echo for respiratory gating: pilot study. Radiology. 1999;211:865–869.[Abstract/Free Full Text]

227. Hatabu H, Gaa J, Kim D, Li W, Prasad PV, Edelman RR. Pulmonary perfusion and angiography: evaluation with breath-hold enhanced three-dimensional fast imaging steady-state precession MR imaging with short TR and TE. AJR Am J Roentgenol.. 1996;167:1653–1655.

228. Loubeyre P, Revel D, Douek P, Delignette A, Baldy C, Genin G, Amiel M. Dynamic contrast-enhanced MR angiography of pulmonary embolism: comparison with pulmonary angiography. AJR Am J Roentgenol.. 1994;162:1035–1039.[Abstract/Free Full Text]

229. Isoda H, Ushimi T, Masui T, Mochizuki T, Goto S, Suzuki K, Shirakawa T, Ohta A, Takahashi M, Kaneko M. Clinical evaluation of pulmonary 3D time-of-flight MRA with breath holding using contrast media. J Comput Assist Tomogr. 1995;19:911–919.[Medline] [Order article via Infotrieve]




This article has been cited by other articles:


Home page
StrokeHome page
R. E. Latchaw, M. J. Alberts, M. H. Lev, J. J. Connors, R. E. Harbaugh, R. T. Higashida, R. Hobson, C. S. Kidwell, W. J. Koroshetz, V. Mathews, et al.
Recommendations for Imaging of Acute Ischemic Stroke: A Scientific Statement From the American Heart Association
Stroke, November 1, 2009; 40(11): 3646 - 3678.
[Full Text] [PDF]


Home page
CirculationHome page
M. A. Creager, C. J. White, W. R. Hiatt, M. H. Criqui, S. C. Josephs, M. J. Alberts, W. H. Pearce, B. H. Gray, and K. J. Rocha-Singh
Atherosclerotic Peripheral Vascular Disease Symposium II: Executive Summary
Circulation, December 16, 2008; 118(25): 2811 - 2825.
[Full Text] [PDF]


Home page
Am. J. Neuroradiol.Home page
F. Runck, R.P. Steiner, W.A. Bautz, and M.M. Lell
MR Imaging: Influence of Imaging Technique and Postprocessing on Measurement of Internal Carotid Artery Stenosis
AJNR Am. J. Neuroradiol., October 1, 2008; 29(9): 1736 - 1742.
[Abstract] [Full Text] [PDF]


Home page
RadiologyHome page
A. M. Gharib, V. B. Ho, D. R. Rosing, D. A. Herzka, M. Stuber, A. E. Arai, and R. I. Pettigrew
Coronary Artery Anomalies and Variants: Technical Feasibility of Assessment with Coronary MR Angiography at 3 T
Radiology, April 1, 2008; 247(1): 220 - 227.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Roentgenol.Home page
L. Saba and G. Mallarini
MDCTA of Carotid Plaque Degree of Stenosis: Evaluation of Interobserver Agreement
Am. J. Roentgenol., January 1, 2008; 190(1): W41 - W46.
[Abstract] [Full Text] [PDF]


Home page
J. Clin. Endocrinol. Metab.Home page
C. A. Bondy, P. L. Van, V. K. Bakalov, and V. B. Ho
Growth Hormone Treatment and Aortic Dimensions in Turner Syndrome
J. Clin. Endocrinol. Metab., May 1, 2006; 91(5): 1785 - 1788.
[Abstract] [Full Text] [PDF]


Home page
RadiologyHome page
N. H. Bunce, C. H. Lorenz, J. Keegan, J. Lesser, E. M. Reyes, D. N. Firmin, and D. J. Pennell
Coronary Artery Anomalies: Assessment with Free-breathing Three-dimensional Coronary MR Angiography
Radiology, April 1, 2003; 227(1): 201 - 208.
[Abstract] [Full Text] [PDF]


Home page
RadiologyHome page
R. Wyttenbach, S. Gianella, M. Alerci, A. Braghetti, L. Cozzi, and A. Gallino
Prospective Blinded Evaluation of Gd-DOTA- versus Gd-BOPTA-enhanced Peripheral MR Angiography, as Compared with Digital Subtraction Angiography
Radiology, April 1, 2003; 227(1): 261 - 269.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
T. Geva, G. F. Greil, A. C. Marshall, M. Landzberg, and A. J. Powell
Gadolinium-Enhanced 3-Dimensional Magnetic Resonance Angiography of Pulmonary Blood Supply in Patients With Complex Pulmonary Stenosis or Atresia: Comparison With X-Ray Angiography
Circulation, July 23, 2002; 106(4): 473 - 478.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
G. F. Greil, A. J. Powell, H. P. Gildein, and T. Geva
Gadolinium-enhanced three-dimensional magnetic resonance angiography of pulmonary and systemic venous anomalies
J. Am. Coll. Cardiol., January 16, 2002; 39(2): 335 - 341.
[Abstract] [Full Text] [PDF]


Home page
Circ. Res.Home page
Z. A. Fayad and V. Fuster
Clinical Imaging of the High-Risk or Vulnerable Atherosclerotic Plaque
Circ. Res., August 17, 2001; 89(4): 305 - 316.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Roentgenol.Home page
M. J. Lipton, L. M. Boxt, and Z. M. Hijazi
Role of the Radiologist in Cardiac Diagnostic Imaging
Am. J. Roentgenol., December 1, 2000; 175(6): 1495 - 1506.
[Full Text] [PDF]


This Article
Free upon publication Free Article
Right arrow Extract
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Yucel, E. K.
Right arrow Articles by Pearce, W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Yucel, E. K.
Right arrow Articles by Pearce, W.
Related Collections
Right arrow AHA Statements and Guidelines
Right arrow Coronary imaging: angiography/ultrasound/Doppler/CC
Right arrow CT and MRI