Donate Help Contact The AHA Sign In Home
American Heart Association
Circulation
Search: search_blue_button Advanced Search
Published Online
on August 24, 2009

Circulation. 2009
Published online before print August 24, 2009, doi: 10.1161/CIRCULATIONAHA.109.859280
A more recent version of this article appeared on September 8, 2009
This Article
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
120/10/867    most recent
CIRCULATIONAHA.109.859280v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by Dewey, M.
Right arrow Articles by Hamm, B.
PubMed
Right arrow PubMed Citation
Right arrow Articles by Dewey, M.
Right arrow Articles by Hamm, B.
Related Collections
Right arrow Cardiovascular imaging agents/Techniques
Right arrow CT and MRI
Right arrowRelated Article

Submitted on February 18, 2009
Accepted on June 22, 2009

Noninvasive Coronary Angiography by 320-Row Computed Tomography With Lower Radiation Exposure and Maintained Diagnostic Accuracy. Comparison of Results With Cardiac Catheterization in a Head-to-Head Pilot Investigation

Marc Dewey MD, PhD*, Elke Zimmermann MD, Florian Deissenrieder MS, Michael Laule MD, PhD, Hans-Peter Dübel MD, PhD, Peter Schlattmann PhD, Fabian Knebel MD, Wolfgang Rutsch MD, PhD, and Bernd Hamm MD, PhD

From the Charité, Medical School, Departments of Radiology (M.D., E.Z., F.D., B.H.), Cardiology (M.L., H.D., F.K., W.R.), and Medical Statistics (P.S.), Humboldt Universität zu Berlin, Freie Universität Berlin, Berlin, Germany.

* To whom correspondence should be addressed. E-mail: dewey{at}charite.de.

Background—Noninvasive coronary angiography with the use of multislice computed tomography (CT) scanners is feasible with high sensitivity and negative predictive value; however, the radiation exposure associated with this technique is rather high. We evaluated coronary angiography using whole-heart 320-row CT, which avoids exposure-intensive overscanning and overranging.

Methods and Results—A total of 30 consecutive patients with suspected coronary artery disease referred for clinically indicated conventional coronary angiography (CCA) were included in this prospective intention-to-diagnose study. CT was performed with the use of up to 320 simultaneous detector rows before same-day CCA, which, together with quantitative analysis, served as the reference standard. The per-patient sensitivity and specificity for CT compared with CCA were 100% (95% confidence interval [CI], 72 to 100) and 94% (95% CI, 73 to 100), respectively. Per-vessel versus per-segment sensitivity and specificity were 89% (95% CI, 62 to 98) and 96% (95% CI, 90 to 99) versus 78% (95% CI, 56 to 91) and 98% (95% CI, 96 to 99), respectively. Interobserver agreement between the 2 readers was significantly better for CCA (97% of 121 coronary arteries) than for CT (90%; P=0.04). Percent diameter stenosis determined with the use of CT showed good correlation with CCA (P<0.001, R=0.81) without significant underestimation or overestimation (-3.1±24.4%; P=0.08). Intraindividual comparison of CT with CCA revealed a significantly smaller effective radiation dose (median, 4.2 versus 8.5 mSv; P<0.05) and amount of contrast agent required (median, 80 versus 111 mL; P<0.001) for 320-row CT. The majority of patients (87%) indicated that they would prefer CT over CCA for future diagnostic imaging (P<0.001).

Conclusions—CT with the use of emerging technology has the potential to significantly reduce the radiation dose and amount of contrast agent required compared with CCA while maintaining high diagnostic accuracy.


Key words: angiography • computed tomography • coronary disease • coronary vessels • imaging


Related Article:

Clinical Summaries
Circulation 2009 120: 823-824. [Extract] [Full Text]