Donate Help Contact The AHA Sign In Home
American Heart Association
Circulation
Search: search_blue_button Advanced Search
Published Online
on May 5, 2008

Circulation. 2008
Published online before print May 5, 2008, doi: 10.1161/CIRCULATIONAHA.107.743120
This Article
Right arrow Full Text (PDF)
Right arrow Data Supplement
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Google Scholar
Right arrow Articles by Chung, E. S.
Right arrow Articles by Murillo, J.
PubMed
Right arrow PubMed Citation
Right arrow Articles by Chung, E. S.
Right arrow Articles by Murillo, J.
Related Collections
Right arrow Congestive
Right arrow Pacemaker
Right arrow Echocardiography

Submitted on October 24, 2007
Accepted on February 27, 2008

Results of the Predictors of Response to CRT (PROSPECT) Trial

Eugene S. Chung MD*, Angel R. Leon MD, Luigi Tavazzi MD, Jing-Ping Sun MD, Petros Nihoyannopoulos MD, John Merlino MD, William T. Abraham MD, Stefano Ghio MD, Christophe Leclercq MD, Jeroen J. Bax MD, Cheuk-Man Yu MD, FRCP, John Gorcsan III MD, Martin St. John Sutton FRCP, Johan De Sutter MD, PhD, and Jaime Murillo MD

From the Lindner Clinical Trial Center, The Ohio Heart and Vascular Center, and The Christ Hospital, Cincinnati, Ohio (E.S.C.); Emory University–Crawford Long Hospital, Atlanta, Ga (A.R.L., J.-P.S., J.M.); Ospedale Policlinico San Matteo, Pavia, Italy (L.T., S.G.); Imperial College London and Hammersmith Hospital, London, UK (P.N.); Ohio State University Heart Center, Columbus (W.T.A.); University Hospital, Rennes, France (C.L.); Leiden University Medical Center, Leiden, the Netherlands (J.J.B.); Prince of Wales Hospital, Hong Kong, China (C.-M.Y.); University of Pittsburgh, Pittsburgh, Pa (J.G.); University of Pennsylvania Medical Center, Philadelphia (M.S.J.S.); University Hospital Gent, Gent, Belgium (J.D.S.); and Sentara Norfolk General Hospital, Norfolk, Va (J.M.).

* To whom correspondence should be addressed. E-mail: chunge{at}ohioheart.org.

Background—Data from single-center studies suggest that echocardiographic parameters of mechanical dyssynchrony may improve patient selection for cardiac resynchronization therapy (CRT). In a prospective, multicenter setting, the Predictors of Response to CRT (PROSPECT) study tested the performance of these parameters to predict CRT response.

Methods and Results—Fifty-three centers in Europe, Hong Kong, and the United States enrolled 498 patients with standard CRT indications (New York Heart Association class III or IV heart failure, left ventricular ejection fraction ≤35%, QRS ≥130 ms, stable medical regimen). Twelve echocardiographic parameters of dyssynchrony, based on both conventional and tissue Doppler–based methods, were evaluated after site training in acquisition methods and blinded core laboratory analysis. Indicators of positive CRT response were improved clinical composite score and ≥15% reduction in left ventricular end-systolic volume at 6 months. Clinical composite score was improved in 69% of 426 patients, whereas left ventricular end-systolic volume decreased ≥15% in 56% of 286 patients with paired data. The ability of the 12 echocardiographic parameters to predict clinical composite score response varied widely, with sensitivity ranging from 6% to 74% and specificity ranging from 35% to 91%; for predicting left ventricular end-systolic volume response, sensitivity ranged from 9% to 77% and specificity from 31% to 93%. For all the parameters, the area under the receiver-operating characteristics curve for positive clinical or volume response to CRT was ≤0.62. There was large variability in the analysis of the dyssynchrony parameters.

Conclusion—Given the modest sensitivity and specificity in this multicenter setting despite training and central analysis, no single echocardiographic measure of dyssynchrony may be recommended to improve patient selection for CRT beyond current guidelines. Efforts aimed at reducing variability arising from technical and interpretative factors may improve the predictive power of these echocardiographic parameters in a broad clinical setting.


Key words: dyssynchrony • echocardiography • echocardiography, Doppler • heart failure • pacemakers