Donate Help Contact The AHA Sign In Home
American Heart Association
Circulation
Search: search_blue_button Advanced Search
Circulation. 1997;95:1022-1029

This Article
Right arrow Full Text
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Vonesh, M. J.
Right arrow Articles by McPherson, D. D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Vonesh, M. J.
Right arrow Articles by McPherson, D. D.

(Circulation. 1997;95:1022-1029.)
© 1997 American Heart Association, Inc.


Articles

In Vitro Identification of Angioplasty-Induced Injury by Use of Vascular Acoustic Emissions

Michael J. Vonesh, PhD; Lyle F. Mockros, PhD; Charles J. Davidson, MD; Krishnan B. Chandran, DSc; David D. McPherson, MD

the Department of Medicine, Section of Cardiology (M.J.V., C.J.D., D.D.M.), and the Department of Biomedical Engineering (M.J.V., L.F.M.), Feinberg Cardiovascular Research Institute, Northwestern University, Chicago, Ill, and the College of Engineering, Department of Biomedical Engineering (K.B.C.), University of Iowa, Iowa City.

Correspondence to David D. McPherson, MD, Section of Cardiology, Northwestern University, 250 E Superior St, Wesley Pavilion, Suite 582, Chicago, IL 60611.

Background We have developed a novel method of diagnosing stress-induced vascular injury. This approach uses the sound energy released from atherosclerotic arterial tissue during in vitro balloon angioplasty to characterize type and severity of induced trauma.

Methods and Results Thirty-two postmortem human peripheral arterial specimens 1.0 cm long were subjected to in vitro balloon angioplasty with simultaneous acoustic emission monitoring. Specimens were examined before and after angioplasty to ascertain the extent of angioplasty-induced injury. Gross observation was used to identify dissection. A three-dimensional intravascular ultrasound reconstruction technique was used to estimate the luminal surface area of the specimen. Change in luminal surface area (postangioplasty minus preangioplasty) was used to quantify induced injury. The energy content and spectral distribution of the digitally acquired vascular acoustic emission (VAE) signals were computed. Comparisons of angioplasty-induced trauma with VAE signal characteristics were made. Dissection (mural laceration of variable depth) was observed in 15 of 32 specimens. Eleven showed no evidence of induced dissection, and 6 had preexisting intimal disruptions. The energy content of the VAE signals collected from specimens with dissection was greater than that obtained from those in which dissection was absent: 845±89.4 mJ (mean±SEM; n=15) versus 128±40.8 mJ (n=11; P<.001). Comparison of induced trauma and VAE signal energy demonstrated a proportional relationship (r=.87, P<.001, n=32).

Conclusions VAE signals contain information characterizing type and severity of angioplasty-induced arterial injury. Because vascular injury is related to adverse procedural outcome, development of VAE technology as an adjunct to conventional diagnostic modalities may facilitate optimal balloon angioplasty delivery and postprocedural care.


Key Words: angioplasty • atherosclerosis • coronary disease • balloon