Circulation, Vol 83, 213-223, Copyright © 1991 by American Heart Association
ER Chafizadeh and WA Zoghbi
To test whether the continuity equation can be applied to the noninvasive
assessment of prosthetic aortic valve function, Doppler echocardiography
was performed in 67 patients (mean age, 58 +/- 14 years) within 10 +/- 6
days after valve replacement with St. Jude Medical valves. All patients
were clinically stable and without evidence of valve dysfunction. Valve
size ranged from 19 to 31 mm, and ejection fraction ranged from 30% to 75%.
With the parasternal long- axis view, the left ventricular outflow diameter
measured just proximal to the prosthetic valve correlated well with valve
size (r = 0.92). Doppler-derived maximal gradients ranged from 9 to 71 mm
Hg. Effective prosthetic aortic valve area by the continuity equation
ranged between 0.73 cm2 for a 19-mm valve and 4.23 cm2 for a 31-mm valve.
With analysis of variance, effective orifice area differentiated various
valve sizes (p less than 10(-14)) better than did gradients alone (p =
0.003) and correlated better with actual valve orifice area (r = 0.83
versus - 0.40). A Doppler velocity index, the ratio of peak velocity in the
left ventricular outflow to that of the aortic jet, averaged 0.41 +/- 0.09
and was less dependent on valve size (r = 0.43). Thus, the continuity
equation can be applied to the assessment of prosthetic St. Jude valves in
the aortic position. By accounting for flow through the valve, it provides
an improved assessment over the sole use of gradients in the evaluation of
prosthetic valve function.
ARTICLES
Doppler echocardiographic assessment of the St. Jude Medical prosthetic valve in the aortic position using the continuity equation
Department of Medicine, Baylor College of Medicine, Houston, Tex.
This article has been cited by other articles:
![]() |
K. Yoshikawa, S. Fukunaga, K. Arinaga, H. Hori, E. Nakamura, T. Ueda, E. Tayama, and S. Aoyagi Long-Term Results of Aortic Valve Replacement With a Small St. Jude Medical Valve in Japanese Patients Ann. Thorac. Surg., April 1, 2008; 85(4): 1303 - 1308. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. R. Bridges, S. M. O'Brien, J. C. Cleveland, E. B. Savage, J. S. Gammie, F. H. Edwards, E. D. Peterson, and F. L. Grover Association between indices of prosthesis internal orifice size and operative mortality after isolated aortic valve replacement J. Thorac. Cardiovasc. Surg., April 1, 2007; 133(4): 1012 - 1021. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Haaverstad, N. Vitale, A. Karevold, G. Cappabianca, A. Tromsdal, P. S. Olsen, L. Kober, H. Ihlen, K. A. Rein, and J. L Svennevig Clinical and echocardiographic assessment of the Medtronic Advantage aortic valve prosthesis: the Scandinavian multicentre, prospective study Heart, April 1, 2007; 93(4): 500 - 505. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. B. Vanky, E. Hakanson, E. Tamas, and R. Svedjeholm Risk Factors for Postoperative Heart Failure in Patients Operated on for Aortic Stenosis Ann. Thorac. Surg., April 1, 2006; 81(4): 1297 - 1304. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. R. Moon, M. K. Pasque, N. A. Munfakh, S. J. Melby, J. S. Lawton, N. Moazami, J. E. Codd, T. D. Crabtree, H. B. Barner, and R. J. Damiano Jr Prosthesis-Patient Mismatch After Aortic Valve Replacement: Impact of Age and Body Size on Late Survival Ann. Thorac. Surg., February 1, 2006; 81(2): 481 - 489. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Mohty-Echahidi, J. F. Malouf, S. E. Girard, H. V. Schaff, D. E. Grill, M. E. Enriquez-Sarano, and F. A. Miller Jr Impact of Prosthesis-Patient Mismatch on Long-Term Survival in Patients With Small St Jude Medical Mechanical Prostheses in the Aortic Position Circulation, January 24, 2006; 113(3): 420 - 426. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. A. Schroeder and J. B. Mark Is the Valve OK or Not? Immediate Evaluation of a Replaced Aortic Valve Anesth. Analg., November 1, 2005; 101(5): 1288 - 1291. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Penta de Peppo, J. Zeitani, P. Nardi, G. Iaci, P. Polisca, R. De Paulis, and L. Chiariello Small "Functional" Size after Mechanical Aortic Valve Replacement: No Risk in Young to Middle-Age Patients Ann. Thorac. Surg., June 1, 2005; 79(6): 1915 - 1920. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. W.H. Sutherland, T. E. Perry, Y. Yu, M. C. Sherwood, E. Rabkin, Y. Masuda, G. A. Garcia, D. L. McLellan, G. C. Engelmayr Jr, M. S. Sacks, et al. From Stem Cells to Viable Autologous Semilunar Heart Valve Circulation, May 31, 2005; 111(21): 2783 - 2791. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. G. Davila-Roman, A. D. Waggoner, E. D. Kennard, R. Holubkov, W. R. E. Jamieson, L. Englberger, T. P. Carrel, and H. V. Schaff Prevalence and severity of paravalvular regurgitation in the Artificial Valve Endocarditis Reduction Trial (AVERT) echocardiography study J. Am. Coll. Cardiol., October 6, 2004; 44(7): 1467 - 1472. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. B. Chambers, L. Oo, A. Narracott, P. M. Lawford, and C. I. Blauth Nominal size in six bileaflet mechanical aortic valves: A comparison of orifice size and biologic equivalence J. Thorac. Cardiovasc. Surg., June 1, 2003; 125(6): 1388 - 1393. [Abstract] [Full Text] [PDF] |
||||
![]() |
X. Y. Jin and J. R. Pepper Do stentless valves make a difference? Eur. J. Cardiothorac. Surg., July 1, 2002; 22(1): 95 - 100. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. D. Milano, M. De Carlo, G. Mecozzi, A. D'Alfonso, G. Scioti, C. Nardi, and U. Bortolotti Clinical outcome in patients with 19-mm and 21-mm St. Jude aortic prostheses: comparison at long-term follow-up Ann. Thorac. Surg., January 1, 2002; 73(1): 37 - 43. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Bech-Hanssen, K. Caidahl, I. Wallentin, P. Ask, and B. Wranne Assessment of effective orifice area of prosthetic aortic valves with Doppler echocardiography: An in vivo and in vitro study J. Thorac. Cardiovasc. Surg., August 1, 2001; 122(2): 287 - 295. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Kadir, I. Y.P. Wan, C. Walsh, P. Wilde, A. J. Bryan, and G. D. Angelini Hemodynamic performance of the 21-mm Sorin Bicarbon mechanical aortic prostheses using dobutamine Doppler echocardiography Ann. Thorac. Surg., July 1, 2001; 72(1): 49 - 53. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. E. Girard, F. A. Miller Jr., T. A. Orszulak, C. J. Mullany, S. Montgomery, W. D. Edwards, H. D. Tazelaar, J. F. Malouf, and A. J. Tajik Reoperation for prosthetic aortic valve obstruction in the era of echocardiography: trends in diagnostic testing and comparison with surgical findings J. Am. Coll. Cardiol., February 1, 2001; 37(2): 579 - 584. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Takakura, T. Sasaki, K. Hashimoto, T. Hachiya, K. Onoguchi, M. Oshiumi, and S. Takeuchi Hemodynamic evaluation of 19-mm Carpentier-Edwards pericardial bioprosthesis in aortic position Ann. Thorac. Surg., February 1, 2001; 71(2): 609 - 613. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Pibarot and J. G. Dumesnil Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention J. Am. Coll. Cardiol., October 1, 2000; 36(4): 1131 - 1141. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. D. Maslow, J. M. Haering, S. Heindel, J. Mashikian, R. Levine, and P. Douglas An Evaluation of Prosthetic Aortic Valves Using Transesophageal Echocardiography: The Double-Envelope Technique Anesth. Analg., September 1, 2000; 91(3): 509 - 516. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Aoyagi, M. Nishimi, H. Kawano, E. Tayama, S. Fukunaga, N. Hayashida, H. Akashi, and T. Kawara Obstruction of st jude medical valves in the aortic positionSignificance of a combination of cineradiography and echocardiography J. Thorac. Cardiovasc. Surg., July 1, 2000; 120(1): 142 - 147. [Abstract] [Full Text] [PDF] |
||||
![]() |
J R Gonzalez-Juanatey, M V Fernandez, F G Sampedro, J M Garcia-Acuna, J B Garcia-Bengoechea, A A Cendon, and M G de la Pena Haemodynamic performance of aortic pericardial bioprostheses and bileaflet prostheses at rest and during exercise: implications for the surgical management of patients with small aortic roots Heart, August 1, 1999; 82(2): 149 - 155. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Barbetseas, S. F. Nagueh, C. Pitsavos, P. K. Toutouzas, M. A. Quinones, and W. A. Zoghbi Differentiating thrombus from pannus formation in obstructed mechanical prosthetic valves: an evaluation of clinical, transthoracic and transesophageal echocardiographic parameters J. Am. Coll. Cardiol., November 1, 1998; 32(5): 1410 - 1417. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Y. Leung, J. Wong, L. Rodriguez, M. Pu, P. M. Vandervoort, and J. D. Thomas Application of Color Doppler Flow Mapping to Calculate Orifice Area of St Jude Mitral Valve Circulation, September 22, 1998; 98(12): 1205 - 1211. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Cam, H. Gercekoglu, S. Celik, M. Gursurer, G. Tayyareci, H. Karabulut, A. Narin, T. Tezel, and B. Yigiter Dobutamine Stress Test to Evaluate Different Sizes of Prosthetic Aortic Valves Asian Cardiovasc Thorac Ann, September 1, 1998; 6(3): 166 - 173. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Demirkilic, E. Kuralay, E. Ozal, H. Bingol, C. Sag, A. T Yilmaz, and H. Tatar Medium-Term Outcome With Small Size ATS Medical Valves in Aortic Position Asian Cardiovasc Thorac Ann, September 1, 1998; 6(3): 183 - 187. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Chakraborty, S. Quek, Ding Zee Pin, Chee Tek Siong, Tan Ley Kheng, and B. Chakraborty Evaluation of Normal Hemodynamic Profile of Carbomedics Prosthetic Valves by Doppler Echocardiography Angiology, December 1, 1997; 48(12): 1055 - 1061. [Abstract] [PDF] |
||||
![]() |
I. Kadir, M. B. Izzat, P. Wilde, B. Reeves, A. J. Bryan, and G. D. Angelini Dynamic Evaluation of the 21-mm Medtronic Intact Aortic Bioprosthesis by Dobutamine Echocardiography Ann. Thorac. Surg., April 1, 1997; 63(4): 1128 - 1132. [Abstract] [Full Text] |
||||
![]() |
M. L. McDonald, R. C. Daly, H. V. Schaff, C. J. Mullany, F. A. Miller, J. J. Morris, and T. A. Orszulak Hemodynamic Performance of Small Aortic Valve Bioprostheses: Is There a Difference? Ann. Thorac. Surg., February 1, 1997; 63(2): 362 - 366. [Abstract] [Full Text] |
||||
![]() |
B. Chakraborty, S. Quek, Ding Zee Pin, Chee Tek Siong, and Tan Lay Kheng Doppler Echocardiographic Assessment of Normally Functioning Starr-Edwards, Carbomedics and Carpentier-Edwards Valves in Aortic Position Angiology, May 1, 1996; 47(5): 481 - 489. [Abstract] [PDF] |
||||
![]() |
T. Carrel, U. Zingg, R. Jenni, B. Aeschbacher, and M. I. Turina Early In Vivo Experience With the Hemodynamic Plus St. Jude Medical Heart Valves in Patients With Narrowed Aortic Annulus Ann. Thorac. Surg., May 1, 1996; 61(5): 1418 - 1422. [Abstract] [Full Text] |
||||
![]() |
M. B. Izzat, I. Birdi, P. Wilde, A. J. Bryan, and G. D. Angelini COMPARISON OF HEMODYNAMIC PERFORMANCES OF ST. JUDE MEDICAL AND CARBOMEDICS 21 MM AORTIC PROSTHESES BY MEANS OF DOBUTAMINE STRESS ECHOCARDIOGRAPHY J. Thorac. Cardiovasc. Surg., February 1, 1996; 111(2): 408 - 415. [Abstract] [Full Text] |
||||
![]() |
T. Ota, K. Iwahashi, H. Matsuda, T. Tsukube, K. Ataka, and M. Okada Reduction of Left Ventricular Hypertrophy with St. Jude Medical 19 mm Valve Prosthesis Angiology, November 1, 1995; 46(11): 981 - 987. [Abstract] [PDF] |
||||
![]() |
M. B. Izzat, I. Birdi, P. Wilde, A. J. Bryan, and G. D. Angelini Evaluation of the Hemodynamic Performance of Small CarboMedics Aortic Prostheses Using Dobutamine-Stress Doppler Echocardiography Ann. Thorac. Surg., October 1, 1995; 60(4): 1048 - 1052. [Abstract] [Full Text] |
||||
![]() |
R. De Paulis, L. Sommariva, F. Russo, F. Tomai, A. Tondo, C. Pagliaricci, C. Bassano, and L. Chiariello Doppler echocardiography evaluation of the CarboMedics valve in patients with small aortic anulus and valve prosthesis-body surface area mismatch J. Thorac. Cardiovasc. Surg., July 1, 1994; 108(1): 57 - 62. [Abstract] [Full Text] |
||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1991 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |