(Circulation. 2008;117:2727-2733.)
© 2008 American Heart Association, Inc.
Arrhythmia/Electrophysiology |
From the Division of Cardiology (J.E., B.A.S., S.O., C.S.) and Basel Institute for Clinical Epidemiology (M.T.K.), University Hospital, Basel, Switzerland; Charité-University Hospital Benjamin Franklin, Berlin, Germany (M.Z.); Division of Cardiology, University of Göttingen, Göttingen, Germany (M.Z.); and Herz-Zentrum Bad Krozingen, Bad Krozingen, Germany (D.K.).
Correspondence to Christian Sticherling, MD, FESC, University Hospital Basel, Division of Cardiology, Petersgraben 4, 4031 Basel, Switzerland. E-mail csticherling{at}uhbs.ch
Received September 17, 2007; accepted March 4, 2008.
Background— Defibrillator lead malfunction is a potential long-term complication in patients with an implantable cardioverter-defibrillator (ICD). The aim of this study was to determine the incidence and causes of lead malfunction necessitating surgical revision and to evaluate 2 approaches to treat lead malfunction.
Methods and Results— We included 1317 consecutive patients with an ICD implanted at 3 European centers between 1993 and 2004. The types and causes of lead malfunction were recorded. If the integrity of the high-voltage part of the lead could be ascertained, an additional pace/sense lead was implanted. Otherwise, the patients received a new ICD lead. Of the 1317 patients, 38 experienced lead malfunction requiring surgical revision and 315 died during a median follow-up of 6.4 years. At 5 years, the cumulative incidence was 2.5% (95% confidence interval, 1.5 to 3.6). Lead malfunction resulted in inappropriate ICD therapies in 76% of the cases. Implantation of a pace/sense lead was feasible in 63%. Both lead revision strategies were similar with regard to lead malfunction recurrence (P=0.8). However, the cumulative incidence of recurrence was high (20% at 5 years; 95% confidence interval, 1.7 to 37.7).
Conclusions— ICD lead malfunction necessitating surgical revision becomes a clinically relevant problem in 2.5% of ICD recipients within 5 years. In selected cases, simple implantation of an additional pace/sense lead is feasible. Regardless of the chosen approach, the incidence of recurrent ICD lead-related problems after lead revision is 8-fold higher in this population.
Related Article:
This article has been cited by other articles:
![]() |
G. Z. Duray, J. Schmitt, S. Cicek-Hartvig, S. H. Hohnloser, and C. W. Israel Complications leading to surgical revision in implantable cardioverter defibrillator patients: comparison of patients with single-chamber, dual-chamber, and biventricular devices Europace, March 1, 2009; 11(3): 297 - 302. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Corrado, G. Gasparini, and A. Raviele Lead malfunctions in implantable cardioverter defibrillators: where are we and where should we go? Europace, March 1, 2009; 11(3): 276 - 277. [Full Text] [PDF] |
||||
![]() |
T. Kleemann and K. Seidl Letter by Kleemann and Seidl Regarding Article, "Necessity for Surgical Revision of Defibrillator Leads Implanted Long-Term: Causes and Management" Circulation, January 6, 2009; 119(1): e8 - e8. [Full Text] [PDF] |
||||
![]() |
C. Sticherling, J. Eckstein, B. A. Schaer, S. Osswald, M. T. Koller, D. Kalusche, and M. Zabel Response to Letter Regarding Article, "Necessity for Surgical Revision of Chronically Implanted Defibrillator Leads: Causes and Management" Circulation, January 6, 2009; 119(1): e9 - e9. [Full Text] [PDF] |
||||
![]() |
G. Boriani, M. Biffi, A. Marziali, I. Diemberger, and C. Martignani A changing scenario in the clinical use of implantable defibrillators: the need for long-term data on lead performance Europace, January 1, 2009; 11(1): 1 - 3. [Full Text] [PDF] |
||||
![]() |
A. Goette, F. Cantu, L. van Erven, P. Geelen, F. Halimi, J. L. Merino, J. M. Morgan, and on behalf of the Scientific Initiative Committee o Performance and survival of transvenous defibrillation leads: need for a European data registry Europace, January 1, 2009; 11(1): 31 - 34. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. G. Hauser A Better Method for Preventing Adverse Clinical Events Caused by Implantable Cardioverter-Defibrillator Lead Fractures? Circulation, November 18, 2008; 118(21): 2117 - 2119. [Full Text] [PDF] |
||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2008 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |