(Circulation. 2006;113:11-19.)
© 2006 American Heart Association, Inc.
Arrhythmia/Electrophysiology |
From the Cardiac Arrhythmia Research Institute, Departments of Medicine (K.Y., H.N., R.L., W.J.M.) and Pathology (J.V.P.), University of Oklahoma Health Sciences Center and Veterans Administration Medical Center, Oklahoma City, Okla; and Heart Lung Center, Utrecht, the Netherlands (F.H.M.W.).
Correspondence to Hiroshi Nakagawa, MD, PhD, Cardiac Arrhythmia Research Institute, University of Oklahoma Health Sciences Center, 1200 Everett Dr (TUH-6E-103), Oklahoma City, OK 73104. E-mail hiroshi-nakagawa{at}ouhsc.edu
Received December 30, 2004; de novo received February 22, 2005; revision received October 10, 2005; accepted October 13, 2005.
Background Electrode cooling by circulating fluid within the electrode (closed loop) or open irrigation facilitates radiofrequency (RF) ablation. This study compared lesion parameters between closed loop and open irrigation with the use of a canine model.
Methods and Results In 8 anesthetized dogs, the skin over the thigh muscle was incised and raised, forming a cradle superfused with heparinized blood (activated clotting time >350 seconds) at 37°C. A 7F 4-mm closed loop electrode (irrigation 36 mL/min) and 7.5F 3.5-mm open irrigation electrode (irrigation 17 mL/min) were positioned perpendicular to the thigh muscle at 10 g contact weight. RF was applied (n=121) at 20 or 30 W for 60 seconds in low (0.1 m/s) or high (0.5 m/s) pulsatile blood flow. Temperatures were measured in the electrode, electrode-tissue interface, and within the tissue at 3- and 7-mm depths. After each RF, the cradle was emptied to examine the electrode and interface for thrombus. There was no difference between closed loop and open irrigation in impedance, lesion depth, or tissue temperature at 20 or 30 W. Interface temperature and electrode temperature were greater in the closed loop application. Thrombus occurred in 32 of 63 closed loop versus 0 of 58 open irrigation RF applications (P<0.05) with interface temperature
80°C in all 32 (electrode temperature <40°C in 1, 40°C to 50°C in 26, and >50°C in 5). With closed loop, interface temperature and thrombus incidence were greater at 30 W and low blood flow. With open irrigation, interface temperature remained low (
71°C) with no difference between 20 and 30 W or between low and high blood flow. Steam pop occurred at 20 W in 4 of 35 closed loop and 0 of 30 open irrigation and at 30 W in 15 of 28 closed loop and 4 of 28 open irrigation applications (P<0.05).
Conclusions Lower interface temperature, thrombus, and steam pop, especially in low blood flow, indicate greater interface cooling with open irrigation.
This article has been cited by other articles:
![]() |
E. M. Aliot, W. G. Stevenson, J. M. Almendral-Garrote, F. Bogun, C. H. Calkins, E. Delacretaz, P. D. Bella, G. Hindricks, P. Jais, M. E. Josephson, et al. EHRA/HRS Expert Consensus on Catheter Ablation of Ventricular Arrhythmias: Developed in a partnership with the European Heart Rhythm Association (EHRA), a Registered Branch of the European Society of Cardiology (ESC), and the Heart Rhythm Society (HRS); in collaboration with the American College of Cardiology (ACC) and the American Heart Association (AHA) Europace, June 1, 2009; 11(6): 771 - 817. [Full Text] [PDF] |
||||
![]() |
M. Linhart, H. Mollnau, A. Bitzen, S. Wurtz, J. W. Schrickel, R. Andrie, F. Stockigt, C. Weiss, G. Nickenig, L. M. Lickfett, et al. In vitro comparison of platinum-iridium and gold tip electrodes: lesion depth in 4 mm, 8 mm, and irrigated-tip radiofrequency ablation catheters Europace, May 1, 2009; 11(5): 565 - 570. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. G. Stevenson, D. J. Wilber, A. Natale, W. M. Jackman, F. E. Marchlinski, T. Talbert, M. D. Gonzalez, S. J. Worley, E. G. Daoud, C. Hwang, et al. Irrigated Radiofrequency Catheter Ablation Guided by Electroanatomic Mapping for Recurrent Ventricular Tachycardia After Myocardial Infarction: The Multicenter Thermocool Ventricular Tachycardia Ablation Trial Circulation, December 16, 2008; 118(25): 2773 - 2782. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. H.M. Wittkampf Image integration in 3D catheter mapping systems: proof of the pudding Eur. Heart J., December 2, 2008; 29(24): 2957 - 2958. [Full Text] [PDF] |
||||
![]() |
K. Yokoyama, H. Nakagawa, D. C. Shah, H. Lambert, G. Leo, N. Aeby, A. Ikeda, J. V. Pitha, T. Sharma, R. Lazzara, et al. Novel Contact Force Sensor Incorporated in Irrigated Radiofrequency Ablation Catheter Predicts Lesion Size and Incidence of Steam Pop and Thrombus Circ Arrhythm Electrophysiol, December 1, 2008; 1(5): 354 - 362. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. J. Blanc, J. Almendral, M. Brignole, M. Fatemi, K. Gjesdal, E. Gonzalez-Torrecilla, P. Kulakowski, G. Y.H. Lip, D. Shah, C. Wolpert, et al. Consensus document on antithrombotic therapy in the setting of electrophysiological procedures Europace, May 1, 2008; 10(5): 513 - 527. [Full Text] [PDF] |
||||
![]() |
T. Miura, K. Yamazaki, S. Kihara, S. Saito, M. Miyagishima, S. Aomi, and H. Kurosawa Transatrial Repair of Submitral Left Ventricular Pseudoaneurysm Ann. Thorac. Surg., February 1, 2008; 85(2): 643 - 645. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Calkins, J. Brugada, D. L. Packer, R. Cappato, S.-A. Chen, H. J.G. Crijns, R. J. Damiano Jr, D. W. Davies, D. E. Haines, M. Haissaguerre, et al. HRS/EHRA/ECAS Expert Consensus Statement on Catheter and Surgical Ablation of Atrial Fibrillation: Recommendations for Personnel, Policy, Procedures and Follow-Up: A report of the Heart Rhythm Society (HRS) Task Force on Catheter and Surgical Ablation of Atrial Fibrillation Developed in partnership with the European Heart Rhythm Association (EHRA) and the European Cardiac Arrhythmia Society (ECAS); in collaboration with the American College of Cardiology (ACC), American Heart Association (AHA), and the Society of Thoracic Surgeons (STS). Endorsed and Approved by the governing bodies of the American College of Cardiology, the American Heart Association, the European Cardiac Arrhythmia Society, the European Heart Rhythm Association, the Society of Thoracic Surgeons, and the Heart Rhythm Society. Europace, June 1, 2007; 9(6): 335 - 379. [Full Text] [PDF] |
||||
![]() |
W. G. Stevenson and K. Soejima Catheter Ablation for Ventricular Tachycardia Circulation, May 29, 2007; 115(21): 2750 - 2760. [Full Text] [PDF] |
||||
![]() |
M. H. Kanj, O. Wazni, T. Fahmy, S. Thal, D. Patel, C. Elay, L. Di Biase, M. Arruda, W. Saliba, R. A. Schweikert, et al. Pulmonary Vein Antral Isolation Using an Open Irrigation Ablation Catheter for the Treatment of Atrial Fibrillation: A Randomized Pilot Study J. Am. Coll. Cardiol., April 17, 2007; 49(15): 1634 - 1641. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. P. Saul and D. Haemmerich Letter by Saul and Haemmerich Regarding Article "Comparison of Electrode Cooling Between Internal and Open Irrigation in Radiofrequency Ablation Lesion Depth and Incidence of Thrombus and Steam Pop" Circulation, June 27, 2006; 113(25): e936 - e936. [Full Text] [PDF] |
||||
![]() |
H. Nakagawa, K. Yokoyama, R. Lazzara, W. M. Jackman, F. H.M. Wittkampf, and J. V. Pitha Response to Letter Regarding Article "Comparison of Electrode Cooling Between Internal and Open Irrigation in Radiofrequency Ablation Lesion Depth and Incidence of Thrombus and Steam Pop" Circulation, June 27, 2006; 113(25): e937 - e937. [Full Text] [PDF] |
||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2006 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |