Donate Help Contact The AHA Sign In Home
American Heart Association
Circulation
Search: search_blue_button Advanced Search
Circulation. 2005;112:954-960
Published online before print August 8, 2005, doi: 10.1161/CIRCULATIONAHA.104.492439
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
112/7/954    most recent
CIRCULATIONAHA.104.492439v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Bruce, G. K.
Right arrow Articles by Packer, D. L.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Bruce, G. K.
Right arrow Articles by Packer, D. L.
Related Collections
Right arrow Ablation/ICD/surgery
Right arrow Arrhythmias, clinical electrophysiology, drugs

(Circulation. 2005;112:954-960.)
© 2005 American Heart Association, Inc.


Arrhythmia/Electrophysiology

Discrepancies Between Catheter Tip and Tissue Temperature in Cooled-Tip Ablation

Relevance to Guiding Left Atrial Ablation

G. Keith Bruce, MD; T. Jared Bunch, MD; Mark A. Milton, MD; Alvaro Sarabanda, MD, PhD; Susan B. Johnson, BS; Douglas L. Packer, MD

From the Division of Cardiovascular Disease, Department of Internal Medicine, Mayo Clinic, Rochester, Minn.

Correspondence to Douglas L. Packer, MD, Heart Rhythm Services, Mayo Clinic, St. Mary’s Hospital, 1216 2nd St SW, AL 2-416, Rochester, MN 55902. E-mail packer.douglas{at}mayo.edu

Received September 14, 2004; revision received April 25, 2005; accepted May 3, 2005.

Background— It is not known whether catheter tip temperatures with a cooled-tip ablation can be reliably extrapolated to estimate actual tissue temperatures. The relationship between catheter tip temperatures, tissue temperatures, power, and microbubble formation is not known.

Methods and Results— Nine dogs underwent 111 radiofrequency energy deliveries at the pulmonary vein ostia with a cooled-tip catheter. Catheter tip and tissue temperatures were markedly discrepant. Catheter tip temperature plateaus at 36°C to 39°C with increasing power, whereas tissue temperature increases to a mean of 75±3°C at 45 W (maximum temperature >100°C). Seventy-two energy deliveries were performed, titrating power to microbubble formation guided by intracardiac echocardiography. Type I and II microbubble formation occurred in 45 (63%) and 19 (26%) ablations, respectively. Type I microbubble emergence occurred at lower powers (21±8 versus 26±4 W; P=0.05), catheter tip temperatures (38±5°C versus 48±10°C; P=0.02), and tissue temperatures (65±19°C versus 81±9°C; P<0.001) than type II microbubble formation. Maximum impedance decreases during ablation before microbubble formation were less with type I microbubble (20±9 versus 37±11 {Omega}; P<0.001) compared with type II microbubbles. One quarter of type I microbubbles abruptly transitioned to type II microbubbles with significant changes in power or catheter tip temperature. No microbubbles were seen in 19 ablations (26%) despite powers up to 26±9 W and tissue temperatures up to 81±17°C.

Conclusions— Catheter tip and tissue temperatures are markedly discrepant during cooled-tip ablation. Type I and II microbubble formation occurs at overlapping power and catheter tip and tissue temperature ranges. Neither the absence of microbubbles nor the presence of type I microbubble formation ensures against excessive tissue heating. The appearance of microbubbles may indicate possible tissue overheating and signal a need to decrease energy.


Key Words: ablation • arrhythmia • atrium • catheter ablation • echocardiography




This article has been cited by other articles:


Home page
EuropaceHome page
M. Linhart, H. Mollnau, A. Bitzen, S. Wurtz, J. W. Schrickel, R. Andrie, F. Stockigt, C. Weiss, G. Nickenig, L. M. Lickfett, et al.
In vitro comparison of platinum-iridium and gold tip electrodes: lesion depth in 4 mm, 8 mm, and irrigated-tip radiofrequency ablation catheters
Europace, May 1, 2009; 11(5): 565 - 570.
[Abstract] [Full Text] [PDF]


Home page
Circ Arrhythmia ElectrophysiolHome page
Y. Okumura, B. D. Henz, S. B. Johnson, T. J. Bunch, C. J. O'Brien, D. O. Hodge, A. Altman, A. Govari, and D. L. Packer
Three-Dimensional Ultrasound for Image-Guided Mapping and Intervention: Methods, Quantitative Validation, and Clinical Feasibility of a Novel Multimodality Image Mapping System
Circ Arrhythmia Electrophysiol, June 1, 2008; 1(2): 110 - 119.
[Abstract] [Full Text] [PDF]


Home page
EuropaceHome page
H. Calkins, J. Brugada, D. L. Packer, R. Cappato, S.-A. Chen, H. J.G. Crijns, R. J. Damiano Jr, D. W. Davies, D. E. Haines, M. Haissaguerre, et al.
HRS/EHRA/ECAS Expert Consensus Statement on Catheter and Surgical Ablation of Atrial Fibrillation: Recommendations for Personnel, Policy, Procedures and Follow-Up: A report of the Heart Rhythm Society (HRS) Task Force on Catheter and Surgical Ablation of Atrial Fibrillation Developed in partnership with the European Heart Rhythm Association (EHRA) and the European Cardiac Arrhythmia Society (ECAS); in collaboration with the American College of Cardiology (ACC), American Heart Association (AHA), and the Society of Thoracic Surgeons (STS). Endorsed and Approved by the governing bodies of the American College of Cardiology, the American Heart Association, the European Cardiac Arrhythmia Society, the European Heart Rhythm Association, the Society of Thoracic Surgeons, and the Heart Rhythm Society.
Europace, June 1, 2007; 9(6): 335 - 379.
[Full Text] [PDF]


Home page
EuropaceHome page
H. Oral and F. Morady
Radiofrequency energy delivery for pulmonary vein isolation: is less more?
Europace, November 1, 2006; 8(11): 966 - 967.
[Full Text] [PDF]