From the Thoraxcenter Erasmus MC (J.A., P.W.S.), Rotterdam, the Netherlands; Ospedale San Raffaele (A.C.), Milan, Italy; Jagiellonian University (D.D., K.Z.), Krakow, Poland; John Radcliffe Hospital (A.P.B.), Oxford, England; PSK No. 7 Zaklad Kardiologii (J.D.), Katowice, Poland; the Centre Hospitalier Universitarie Jean Minjoz (F.S.), Besancon, France; and Boston Scientific Corp (M.E.R., J.K.), Natick, Mass.
Correspondence to P.W. Serruys, MD, PhD, Thoraxcenter, Ba 583, Erasmus MC, Dr Molewaterplein 40, 3015 GD Rotterdam, The Netherlands. E-mail p.w.j.c.serruys{at}erasmusmc.nl
Received April 26, 2005; revision received October 16, 2005; accepted October 18, 2005.
Background The purpose of this study was to evaluate long-term vascular responses as long as 2 years after implantation of polymer-based, paclitaxel-eluting stents in contrast to uncoated stents.
Methods and Results TAXUS II is a randomized, double-blind trial comparing slow-release (SR) and moderate-release (MR) TAXUS stents with bare-metal control stents (BMSs). One hundred sixty-one event-free patients (SR, 43; MR, 41; and BMS, 77) underwent serial intravascular ultrasound (IVUS) analysis after the procedure and at 6 months and 2 years. At 2 years, neointimal responses continued to be significantly suppressed in the SR and MR groups when compared with the BMS group (BMS, 1.49±1.12 mm2; SR, 0.94±0.76 mm2 [P=0.004]; and MR, 1.06±0.90 mm2 [P=0.02]). Between 6 months and 2 years, the BMS group showed compaction of the neointima (
, 0.22±1.05 mm2 [P=0.08]). In contrast, both the SR and MR groups exhibited an increase (
SR, 0.30±0.76 mm2 (P=0.01); MR, 0.41±0.94 mm2 [P=0.009]). Between 6 months and 2 years, the initial increase in plaque outside the stent regressed in the BMS and SR groups to levels comparable to those after the procedure, whereas expansive remodeling partially regressed in the MR group (
between after the procedure and 2 years BMS, 0.34±1.28 mm2 [P=0.05]; SR, 0.02±1.40 mm2 [P=0.93]; MR, 0.32±1.56 mm2 [P=0.27]).
Conclusions The 2-year follow-up demonstrates that neointimal suppression was dose independent and that this effect was still sustained at 2 years. However, the increase in area outside the stent seen at 6 months regressed to different extents in a dose-dependent manner at 2 years.
This article has been cited by other articles:
![]() |
Y. Onuma, P. Serruys, P. den Heijer, K. S. Joesoef, H. Duckers, E. Regar, N. Kukreja, S. Tanimoto, H. M. Garcia-Garcia, H. van Beusekom, et al. MAHOROBA, first-in-man study: 6-month results of a biodegradable polymer sustained release tacrolimus-eluting stent in de novo coronary stenoses Eur. Heart J., June 2, 2009; 30(12): 1477 - 1485. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. A. Byrne, R. Iijima, J. Mehilli, S. Pinieck, O. Bruskina, A. Schomig, and A. Kastrati Durability of Antirestenotic Efficacy in Drug-Eluting Stents With and Without Permanent Polymer J. Am. Coll. Cardiol. Intv., April 1, 2009; 2(4): 291 - 299. [Abstract] [Full Text] [PDF] |
||||
![]() |
H-J Kang, Y-S Kim, B-K Koo, K W Park, H-Y Lee, D-W Sohn, B-H Oh, Y-B Park, and H-S Kim Effects of stem cell therapy with G-CSF on coronary artery after drug-eluting stent implantation in patients with acute myocardial infarction Heart, May 1, 2008; 94(5): 604 - 609. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Aoki, G. S. Mintz, N. J. Weissman, J. T. Mann, L. Cannon, J. Greenberg, E. Grube, A.R. Z. Masud, J. Koglin, L. Mandinov, et al. Chronic Arterial Responses to Overlapping Paclitaxel-Eluting Stents: Insights From Serial Intravascular Ultrasound Analyses in the TAXUS-V and -VI Trials J. Am. Coll. Cardiol. Intv., April 1, 2008; 1(2): 161 - 167. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Daemen and P. W. Serruys Drug-Eluting Stent Update 2007: Part II: Unsettled Issues Circulation, August 21, 2007; 116(8): 961 - 968. [Full Text] [PDF] |
||||
![]() |
M. Valgimigli, G. Campo, C. Arcozzi, P. Malagutti, R. Carletti, F. Ferrari, D. Barbieri, G. Parrinello, G. Percoco, and R. Ferrari Two-Year Clinical Follow-Up After Sirolimus-Eluting Versus Bare-Metal Stent Implantation Assisted by Systematic Glycoprotein IIb/IIIa Inhibitor Infusion in Patients With Myocardial Infarction: Results From the STRATEGY Study J. Am. Coll. Cardiol., July 10, 2007; 50(2): 138 - 145. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. J. Weissman, S. G. Ellis, E. Grube, K. D. Dawkins, J. D. Greenberg, T. Mann, L. A. Cannon, P. A. Cambier, S. Fernandez, G. S. Mintz, et al. Effect of the polymer-based, paclitaxel-eluting TAXUS Express stent on vascular tissue responses: a volumetric intravascular ultrasound integrated analysis from the TAXUS IV, V, and VI trials Eur. Heart J., July 1, 2007; 28(13): 1574 - 1582. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. S. Mintz What to Do About Late Incomplete Stent Apposition? Circulation, May 8, 2007; 115(18): 2379 - 2381. [Full Text] [PDF] |
||||
![]() |
E. Camenzind, P. G. Steg, and W. Wijns A Cause for Concern Circulation, March 20, 2007; 115(11): 1440 - 1455. [Full Text] [PDF] |
||||
![]() |
P. W. Serruys and J. Daemen Late Stent Thrombosis: A Nuisance in Both Bare Metal and Drug-Eluting Stents Circulation, March 20, 2007; 115(11): 1433 - 1439. [Full Text] [PDF] |
||||
![]() |
B. Lagerqvist, S. K. James, U. Stenestrand, J. Lindback, T. Nilsson, L. Wallentin, and the SCAAR Study Group Long-Term Outcomes with Drug-Eluting Stents versus Bare-Metal Stents in Sweden N. Engl. J. Med., March 8, 2007; 356(10): 1009 - 1019. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. J. Murphy, T. W. Johnson, M. H. Chamberlain, S. I. Rizvi, M. Wyatt, S. J. George, G. D. Angelini, K. R. Karsch, M. Oberhoff, and A. C. Newby Short- and long-term effects of cytochalasin D, paclitaxel and rapamycin on wall thickening in experimental porcine vein grafts Cardiovasc Res, February 1, 2007; 73(3): 607 - 617. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. Yang, J. Ge, H. Liu, K. Zhao, X. Liu, X. Qu, W. Li, Y. Huang, A. Sun, and Y. Zou Arsenic trioxide eluting stent reduces neointima formation in a rabbit iliac artery injury model Cardiovasc Res, December 1, 2006; 72(3): 483 - 493. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Camenzind Treatment of In-Stent Restenosis -- Back to the Future? N. Engl. J. Med., November 16, 2006; 355(20): 2149 - 2151. [Full Text] [PDF] |
||||
![]() |
R. Gupta, F. Al-Ali, A. J. Thomas, M. B. Horowitz, T. Barrow, N. A. Vora, K. Uchino, M. D. Hammer, L. R. Wechsler, and T. G. Jovin Safety, Feasibility, and Short-Term Follow-Up of Drug-Eluting Stent Placement in the Intracranial and Extracranial Circulation Stroke, October 1, 2006; 37(10): 2562 - 2566. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. S. Mintz and N. J. Weissman Intravascular Ultrasound in the Drug-Eluting Stent Era J. Am. Coll. Cardiol., August 1, 2006; 48(3): 421 - 429. [Abstract] [Full Text] [PDF] |
||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2005 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |