(Circulation. 2005;111:321-327.)
© 2005 American Heart Association, Inc.
Interventional Cardiology |
From the Brigham and Womens Hospital (L.M., E.J.O., R.E.K.), Harvard Medical School (A.J.O., L.M., E.J.O., R.E.K., D.E.C.), and Harvard Clinical Research Institute (L.M., R.E.K., D.E.C., C.S.), Boston, Mass; Lenox Hill Heart and Vascular Institute (J.W.M., M.B.L.), New York, NY; Mayo Clinic (D.R.H.), Rochester, Minn; Scripps Clinic (P.S.T.), La Jolla, Calif; Center for Cardiology and Vascular Intervention (J.S.), Hamburg, Germany; Department of Cardiology and Angiology (G.B.), University Hospital, Münster, Germany; Beth Israel Deaconess Medical Center (D.E.C.), Boston, Mass; Ohio Heart Health Center (D.J.K.), Cincinnati, Ohio; and Cordis Corporation (B.G.F., D.J.D.), Warren, NJ.
Correspondence to Richard E. Kuntz, MD, MSc, BC-3-012N, Brigham and Womens Hospital, 75 Francis St, Boston, MA 02116. E-mail rkuntz{at}partners.org
Received August 2, 2004; revision received September 26, 2004; accepted October 28, 2004.
Background Observed rates of restenosis after drug-eluting stenting are low (<10%). Identification of a reliable and powerful angiographic end point will be useful in future trials.
Methods and Results Late loss (postprocedural minimum lumen diameter minus 8-month minimum lumen diameter) was measured in the angiographic cohorts of the SIRIUS (n=703) and E-SIRIUS (n=308) trials. Two techniques, the standard normal approximation and an optimized power transformation, were used to predict binary angiographic restenosis rates and compare them with observed restenosis rates. The mean in-stent late loss observed in the SIRIUS trial was 0.17±0.45 mm (sirolimus) versus 1.00±0.70 mm (control). If a normal distribution was assumed, late loss accurately estimated in-stent binary angiographic restenosis for the control arm (predicted 35.4% versus observed 35.4%) but underestimated it in the sirolimus arm (predicted 0.6% versus observed 3.2%). Power transformation improved the reliability of the estimate in the sirolimus arm (predicted 3.2% [CI 1.0% to 6.7%]) with similar improvements in the E-SIRIUS trial (predicted 4.0% [CI 1.2% to 7.0%] versus observed 3.9%). In the sirolimus-eluting stent arm, in-stent late loss correlated better with target-lesion revascularization than in-segment late loss (c-statistic=0.915 versus 0.665).
Conclusions Because distributions of late loss with a low mean are right-skewed, the use of a transformation improves the accuracy of predicting low binary restenosis rates. Late loss is monotonically correlated with the probability of restenosis and yields a more efficient estimate of the restenosis process in the era of lower binary restenosis rates.
Key Words: angioplasty stents restenosis
This article has been cited by other articles:
![]() |
S. J. Pocock, A. J. Lansky, R. Mehran, J. J. Popma, M. P. Fahy, Y. Na, G. Dangas, J. W. Moses, T. Pucelikova, D. E. Kandzari, et al. Angiographic surrogate end points in drug-eluting stent trials: a systematic evaluation based on individual patient data from 11 randomized, controlled trials. J. Am. Coll. Cardiol., January 1, 2008; 51(1): 23 - 32. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Tomai, B. Reimers, L. De Luca, A. R. Galassi, A. Gaspardone, A. S. Ghini, V. Ferrero, L. Favero, G. Gioffre, F. Prati, et al. Head-to-Head Comparison of Sirolimus- and Paclitaxel-Eluting Stent in the Same Diabetic Patient With Multiple Coronary Artery Lesions: A prospective, randomized, multicenter study Diabetes Care, January 1, 2008; 31(1): 15 - 19. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Baumgart, V. Klauss, F. Baer, F. Hartmann, H. Drexler, W. Motz, H. Klues, S. Hofmann, W. Volker, T. Pfannebecker, et al. One-Year Results of the SCORPIUS Study: A German Multicenter Investigation on the Effectiveness of Sirolimus-Eluting Stents in Diabetic Patients J. Am. Coll. Cardiol., October 23, 2007; 50(17): 1627 - 1634. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. J. Kereiakes, L. Mauri, and R. E. Kuntz Letter by Kereiakes et al Regarding Article, "Stent Thrombosis Late After Implantation of First-Generation Drug-Eluting Stents: A Cause for Concern" Circulation, October 16, 2007; 116(16): e389 - e389. [Full Text] [PDF] |
||||
![]() |
J. Cosgrave, G. Melzi, S. Corbett, G. G.L. Biondi-Zoccai, P. Agostoni, R. Babic, F. Airoldi, A. Chieffo, G. M. Sangiorgi, M. Montorfano, et al. Comparable Clinical Outcomes With Paclitaxel- and Sirolimus-Eluting Stents in Unrestricted Contemporary Practice J. Am. Coll. Cardiol., June 19, 2007; 49(24): 2320 - 2328. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. H. Zou, A. J. O'Malley, and L. Mauri Receiver-Operating Characteristic Analysis for Evaluating Diagnostic Tests and Predictive Models Circulation, February 6, 2007; 115(5): 654 - 657. [Full Text] [PDF] |
||||
![]() |
F. Alfonso, H. Jneid, A. O. Maree, I. F. Palacios, M. Valgimigli, G. Percoco, L. Bolognese, C. Spaulding, P. Henry, E. Teiger, et al. Drug-eluting stents in primary PCI. N. Engl. J. Med., December 7, 2006; 355(23): 2483 - 2484. [Full Text] [PDF] |
||||
![]() |
B. H. Strauss All Drug-Eluting Stents Are Equal, But Some Drug-Eluting Stents Are More Equal Than Others J. Am. Coll. Cardiol., October 3, 2006; 48(7): 1310 - 1311. [Full Text] [PDF] |
||||
![]() |
F. Alfonso, M.-J. Perez-Vizcayno, R. Hernandez, A. Bethencourt, V. Marti, J. R. Lopez-Minguez, J. Angel, R. Mantilla, C. Moris, A. Cequier, et al. A Randomized Comparison of Sirolimus-Eluting Stent With Balloon Angioplasty in Patients With In-Stent Restenosis: Results of the Restenosis Intrastent: Balloon Angioplasty Versus Elective Sirolimus-Eluting Stenting (RIBS-II) Trial J. Am. Coll. Cardiol., June 6, 2006; 47(11): 2152 - 2160. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Seabra-Gomes Percutaneous coronary interventions with drug eluting stents for diabetic patients. Heart, March 1, 2006; 92(3): 410 - 419. [Full Text] [PDF] |
||||
![]() |
M.-C. Morice, A. Colombo, B. Meier, P. Serruys, C. Tamburino, G. Guagliumi, E. Sousa, H.-P. Stoll, and for the REALITY Trial Investigators Sirolimus- vs Paclitaxel-Eluting Stents in De Novo Coronary Artery Lesions: The REALITY Trial: A Randomized Controlled Trial JAMA, February 22, 2006; 295(8): 895 - 904. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. J. Brener Wish List and REALITY: Choice of Stents and End Points for Treatment of De Novo Coronary Artery Lesions JAMA, February 22, 2006; 295(8): 937 - 938. [Full Text] [PDF] |
||||
![]() |
R. Wessely, A. Schomig, and A. Kastrati Sirolimus and Paclitaxel on Polymer-Based Drug-Eluting Stents: Similar But Different J. Am. Coll. Cardiol., February 21, 2006; 47(4): 708 - 714. [Abstract] [Full Text] [PDF] |
||||
![]() |
M Thomas Are drug eluting stents really worth the money? Heart, January 1, 2006; 92(1): 5 - 7. [Full Text] [PDF] |
||||
![]() |
L. Mauri, E. J. Orav, S. C. Candia, D. E. Cutlip, and R. E. Kuntz Robustness of Late Lumen Loss in Discriminating Drug-Eluting Stents Across Variable Observational and Randomized Trials Circulation, November 1, 2005; 112(18): 2833 - 2839. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Dibra, A. Kastrati, J. Mehilli, J. Pache, H. Schuhlen, N. von Beckerath, K. Ulm, R. Wessely, J. Dirschinger, A. Schomig, et al. Paclitaxel-Eluting or Sirolimus-Eluting Stents to Prevent Restenosis in Diabetic Patients N. Engl. J. Med., August 18, 2005; 353(7): 663 - 670. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Mauri, E. J. Orav, and R. E. Kuntz Late Loss in Lumen Diameter and Binary Restenosis for Drug-Eluting Stent Comparison Circulation, June 28, 2005; 111(25): 3435 - 3442. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Costa and D. I. Simon Molecular Basis of Restenosis and Drug-Eluting Stents Circulation, May 3, 2005; 111(17): 2257 - 2273. [Full Text] [PDF] |
||||
![]() |
D. J. Kereiakes, R. E. Kuntz, L. Mauri, and M. W. Krucoff Surrogates, substudies, and real clinical end points in trials of drug-eluting stents J. Am. Coll. Cardiol., April 19, 2005; 45(8): 1206 - 1212. [Full Text] [PDF] |
||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2005 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |