Background Although overdrive pacing for treating atrial flutter is well established, the efficacy of device-based atrial pacing for treating spontaneous atrial tachyarrhythmias in patients with implantable cardioverter defibrillators (ICD) is unknown. This study evaluated the efficacy of novel pacing therapies for treating atrial tachyarrhythmias in patients receiving a dual-chamber ICD to treat ventricular tachyarrhythmias.
Methods and Results A Jewel AF ICD was implanted in 537 patients with ventricular arrhythmia who were followed for 11.4±8.2 months (74% had a documented history of atrial tachyarrhythmias). The device discriminated atrial tachycardia (AT) from atrial fibrillation (AF) on the basis of cycle length and regularity, and it used 3 different methods of overdrive atrial pacing (Ramp, Burst+, and 50-Hz burst) to treat AT episodes and one method (50-Hz burst) to treat AF episodes. Pacing successfully terminated 59% of 1500 spontaneous AT episodes in 127 patients and 30% of 880 AF episodes in 101 patients (P<0.001). With AT and AF episodes combined, pacing efficacy was 48%. Pacing efficacy was significantly reduced at AT cycle lengths
220 ms and AF cycle lengths
160 ms (P<0.01) The median time from pacing to AT/AF termination was 1.1 minute for the pacing successes and 96 minutes for the failures (P<0.001).
Conclusions Atrial pacing terminated 48% of AT/AF episodes in patients with a history of ventricular arrhythmias. Pacing efficacy was greater for device-classified AT than AF. Pacing efficacy was significantly influenced by tachycardia cycle length. Successful pacing significantly reduces the time required for AT/AF termination.
Key Words: atrial flutter pacing tachyarrhythmias fibrillation
This article has been cited by other articles:
![]() |
R. P. Ricci, A. Quesada, J. Almendral, F. Arribas, C. Wolpert, P. Adragao, M. Zoni-Berisso, X. Navarro, T. DeSanto, A. Grammatico, et al. Dual-chamber implantable cardioverter defibrillators reduce clinical adverse events related to atrial fibrillation when compared with single-chamber defibrillators: a subanalysis of the DATAS trial Europace, May 1, 2009; 11(5): 587 - 593. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Almendral, F. Arribas, C. Wolpert, R. Ricci, P. Adragao, E. Cobo, X. Navarro, A. Quesada, and the DATAS Steering Committee and Writing Committee Dual-chamber defibrillators reduce clinically significant adverse events compared with single-chamber devices: results from the DATAS (Dual chamber and Atrial Tachyarrhythmias Adverse events Study) trial Europace, May 1, 2008; 10(5): 528 - 535. [Abstract] [Full Text] [PDF] |
||||
![]() |
Developed in Collaboration With the European Heart, D. P. Zipes, A. J. Camm, M. Borggrefe, A. E. Buxton, B. Chaitman, M. Fromer, G. Gregoratos, G. Klein, A. J. Moss, et al. ACC/AHA/ESC 2006 Guidelines for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death: A Report of the American College of Cardiology/American Heart Association Task Force and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Develop Guidelines for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death) J. Am. Coll. Cardiol., September 5, 2006; 48(5): e247 - e346. [Full Text] [PDF] |
||||
![]() |
Writing Committee Members, D. P. Zipes, A. J. Camm, M. Borggrefe, A. E. Buxton, B. Chaitman, M. Fromer, G. Gregoratos, G. Klein, A. J. Moss, et al. ACC/AHA/ESC 2006 guidelines for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: A report of the American College of Cardiology/American Heart Association Task Force and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Develop Guidelines for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death) Developed in collaboration with the European Heart Rhythm Association and the Heart Rhythm Society Europace, September 1, 2006; 8(9): 746 - 837. [Full Text] [PDF] |
||||
![]() |
M. Gulizia, S. Mangiameli, S. Orazi, G. Chiaranda, G. Boriani, G. Piccione, N. DiGiovanni, A. Colletti, C. Puntrello, G. Butera, et al. Randomized comparison between Ramp and Burst+ atrial antitachycardia pacing therapies in patients suffering from sinus node disease and atrial fibrillation and implanted with a DDDRP device Europace, July 1, 2006; 8(7): 465 - 473. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Schwartzman, M. N. Harvey, R. H. Hoyt, J. L. Koehler, M. R. Ujhelyi, and D. E. Euler Utility of adjunctive single oral bolus propafenone therapy in patients with atrial defibrillators. Europace, March 1, 2006; 8(3): 211 - 215. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Vollmann, E. Kollet, L. Lüthje, S. Busse, and C. Unterberg Comparison of immediate and delayed automatic antitachycardia pacing for the termination of atrial tachyarrhythmias Europace, January 1, 2005; 7(3): 248 - 254. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Quesada, J.ús Almendral, F. Arribas, R. Ricci, C. Wolpert, P. Adragao, E. Cobo, X. Navarro, and on behalf DATAS investigators The DATAS rationale and design: a controlled, randomized trial to assess the clinical benefit of dual chamber (DDED) defibrillator Europace, January 1, 2004; 6(2): 142 - 150. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.C. Geller, S. Reek, C. Timmermans, T. Kayser, H.-F. Tse, C. Wolpert, W. Jung, A.J. Camm, C.-P. Lau, H. J.J. Wellens, et al. Treatment of atrial fibrillation with an implantable atrial defibrillator -- long term results Eur. Heart J., December 1, 2003; 24(23): 2083 - 2089. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.-N. Pak, Y.-B. Liu, H. Hayashi, Y. Okuyama, P.-S. Chen, and S.-F. Lin Synchronization of ventricular fibrillation with real-time feedback pacing: implication to low-energy defibrillation Am J Physiol Heart Circ Physiol, December 1, 2003; 285(6): H2704 - H2711. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Hügl, C. W. Israel, C. Unterberg, T. Lawo, J. C. Geller, I. M. Kennis, D. E. Euler, J. L. Koehler, D. A. Hettrick, and on behalf of the AT500 Verification Study Investig Incremental programming of atrial anti-tachycardia pacing therapies in bradycardia-indicated patients: effects on therapy efficacy and atrial tachyarrhythmia burden Europace, January 1, 2003; 5(4): 403 - 409. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Santini and R. Ricci Atrial fibrillation and heart failure Europace, January 1, 2003; 5(s1): S55 - S59. [Full Text] [PDF] |
||||
![]() |
T. H. Everett IV, J. G. Akar, L.-C. Kok, J. R. Moorman, and D. E. Haines Use of global atrial fibrillation organization to optimize the success of burst pace termination J. Am. Coll. Cardiol., November 20, 2002; 40(10): 1831 - 1840. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Saksena and N. Madan Management of the Patient With an Implantable Cardioverter-Defibrillator in the Third Millennium Circulation, November 19, 2002; 106(21): 2642 - 2646. [Full Text] [PDF] |
||||
![]() |
A. M. Gillis, C. Unterberg-Buchwald, H. Schmidinger, S. Massimo, K. Wolfe, D. J. Kavaney, M. F. Otterness, S. H. Hohnloser, and GEM III AT Worldwide Investigators Safety and efficacy of advanced atrial pacing therapies for atrial tachyarrhythmias in patients with a new implantable dual chamber cardioverter-defibrillator J. Am. Coll. Cardiol., November 6, 2002; 40(9): 1653 - 1659. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Cooper, M. S. Katcher, and M. V. Orlov Implantable Devices for the Treatment of Atrial Fibrillation N. Engl. J. Med., June 27, 2002; 346(26): 2062 - 2068. [Full Text] [PDF] |
||||
![]() |
S.R. Raj and D.G. Wyse Rhythm control for paroxysmal atrial fibrillation after AV junction ablation: is it worth the effort? Eur. Heart J., June 1, 2002; 23(11): 845 - 848. [Full Text] [PDF] |
||||
![]() |
J. C. Daubert and P. Mabo Implantable devices to treat atrial fibrillation: real prospects or just new gimmicks? Europace, January 1, 2002; 4(2): 161 - 164. [PDF] |
||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2001 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |