(Circulation. 2000;101:1970.)
© 2000 American Heart Association, Inc.
Clinical Investigation and Reports |
From the I. Medizinische Klinik und Poliklinik (C.S., C.N., C.A., R.A.H., E.A.), the Nuklearmedizinische Klinik und Poliklinik (I.W.), the Institut für Medizinische Statistik und Epidemiologie (K.U.), Klinikum rechts der Isar, and the Institut für Radiochemie (R.H., X.L.), Technische Universität München, Munich; and the Institut für Strahlenschutz (C.H., W.P., D.F.R.), GSF-Forschungszentrum für Umwelt und Gesundheit, Neuherberg, Germany.
Correspondence to Eckhard Alt, MD, I. Medizinische Klinik und Poliklinik, Klinikum rechts der Isar, Technische Universität München, Ismaningerstraße 22, 81675 München, Germany. E-mail alt{at}med1.med.tu-muenchen.de
BackgroundRecent studies have shown that ionizing radiation reduces neointima formation after balloon angioplasty and stent implantation in experimental models of restenosis and first clinical trials. The objective of this study was to determine the dose distribution of a new ß-particleemitting radioactive gold stent and to evaluate the dose-dependent vascular response in the coronary overstretch pig model.
Methods and ResultsSixteen Göttinger minipigs underwent placement of 11 nonradioactive and 36 ß-particleemitting stents with activity levels of 10.4±0.6, 14.9±2.4, 22.8±1.3, 35.8±2.8, and 55.4±5.3 µCi of 198Au. Three months after implantation, the percent area stenosis, neointimal thickness, neointimal area, and vessel injury were analyzed by quantitative histomorphometry. The lifetime radiation doses at a depth of 1 mm were 3.3±0.2, 4.7±0.5, 7.2±0.4, 11.4±0.9, and 17.6±1.7 Gy for the different activity groups. No dose-response relationship was observed in the radioactive stents with respect to percent area stenosis (P=0.297), mean neointimal thickness (P=0.82), or mean neointimal area (P=0.65). Significantly lower neointima formation and less luminal narrowing was seen in the control group than in the ß-particleemitting stents (P<0.001). Multilinear regression analysis revealed that only radioactivity made a significant independent contribution to the degree of percent area stenosis (P<0.001).
ConclusionsNeointima formation in pigs is markedly increased by ß-particleemitting stents with 198Au as the radioisotope. This study provides evidence that dosages of 3 to 18 Gy of low-dose-rate ß-particle irradiation via endovascular stents cause pronounced luminal narrowing in the animal model at 3 months.
Key Words: coronary disease stenosis stents radioisotopes
This article has been cited by other articles:
![]() |
C. Bartels, A. Erasmi, F. Sayk, R. Eggers, A. Dendorfer, T. Feyerabend, W. Eichler, and Hans.-H. Sievers Prophylactic gamma radiation of unaffected vein grafts failed to prevent vein graft disease in a chronic hypercholesterolemic porcine model Eur. J. Cardiothorac. Surg., July 1, 2003; 24(1): 92 - 97. [Abstract] [Full Text] [PDF] |
||||
|