Donate Help Contact The AHA Sign In Home
American Heart Association
Circulation
Search: search_blue_button Advanced Search
Circulation. 1999;100:II-236-II-243

This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Artrip, J. H.
Right arrow Articles by Wang, J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Artrip, J. H.
Right arrow Articles by Wang, J.
Related Collections
Right arrow Other heart failure
Right arrow Animal models of human disease
Right arrow Other Treatment
Right arrow Coronary circulation

(Circulation. 1999;100:II-236.)
© 1999 American Heart Association, Inc.


Thoracic Transplantation and Ventricular Assist Devices

Physiological and Hemodynamic Evaluation of Nonuniform Direct Cardiac Compression

Presented in part at the 71th Scientific Sessions of the American Heart Association, Dallas, Tex, November 8–11, 1998.

John H. Artrip, MD; Geng-Hua Yi, MD; Howard R. Levin, MD; Daniel Burkhoff, MD, PhD; Jie Wang, MD, PhD

From the Department of Surgery, Division of Cardiothoracic Surgery (J.H.A.), and Department of Medicine, Division of Circulatory Physiology (G.-H.Y., D.B., J.W.), College of Physicians and Surgeons, Columbia University, New York, NY; and Cardio Technologies Inc (H.R.L.), Pine Brook, NJ.

Correspondence to Jie Wang, MD, PhD, Department of Medicine, Division of Circulatory Physiology, College of Physicians and Surgeons, Columbia University, MHB 5-435, 177 Fort Washington Ave, New York, NY 10032. E-mail jw147{at}columbia.edu

Background—Biventricular direct cardiac compression (DCC) has the potential to support the failing heart without the complications associated with a blood/device interface encountered with the use of current ventricular assist devices. A clinically designed DCC device that provides compression pressure around the base of the heart in synchrony with native ventricular contractions was evaluated with the use of an ex vivo and in vivo canine model of heart failure.

Methods and Results—The device was tested over a series of ventricular preloads with the use of an ex vivo canine heart preparation and computerized afterload system that mimicked the conditions of heart failure. The end-systolic pressure-volume relation of the left and right ventricles was shifted upward in parallel by DCC, with the magnitude of the shift averaging 40% of the device compression pressure. The device was tested in vivo with the use of a canine model of acute ischemic heart failure in which graded reductions in ventricular function were created through serial coronary artery embolizations. Under the most severe condition of heart failure, DCC improved cardiac output (CO) by 104% (0.80±0.33 to 1.63±0.40 L/min) and mean arterial pressure by 95% (45.6±11 to 89.0±18.2 mm Hg). The CO was typically restored to {approx}60% of the normal baseline value, despite attempts to further increase CO by increasing the amount or duration of compression pressure.

Conclusions—Nonuniform DCC significantly improves the left and right ventricular pressure-generating capability and, in the setting of acute heart failure, can increase CO and mean arterial pressure. Such DCC devices can potentially avoid the complications associated with currently available ventricular support devices that involve a blood/device interface.


Key Words: heart failure • hemodynamics • ischemia • physiology • cardiac assist devices