(Circulation. 1999;100:2344.)
© 1999 American Heart Association, Inc.
Clinical Investigation and Reports |
From the University of Alabama, Birmingham (G.N.K., R.C.R.), Johns Hopkins University (J.A.B.), Baltimore, Md; University of Southern California (D.T.K.), Los Angeles; Ohio State University (C.J.L.), Columbus; Mayo Clinic Foundation (M.A.L.), Rochester, Minn; Clinique Alleray-Labrousse (G.P.), Paris, France; Accufix Research Institute (J.F., M.K.O., L.G.E.), Denver, Colo; and Medical Data Research Center (G.L.G.), Providence Health System, Portland, Ore.
Correspondence to G. Neal Kay, MD, 321J Tinsley Harrison Tower, University of Alabama at Birmingham, Birmingham, AL 35294.
BackgroundThe Telectronics Accufix pacing leads were recalled in November 1994 after 2 deaths and 2 nonfatal injuries were reported. This multicenter clinical study (MCS) of patients with Accufix leads was designed to determine the rate of spontaneous injury related to the J retention wire and results of lead extraction.
Methods and ResultsThe MCS included 2589 patients with Accufix
atrial pacing leads that were implanted at or who were followed up at
12 medical centers. Patients underwent cinefluoroscopic imaging of
their lead every 6 months. The risk of J retention wire fracture was
5.6%/y at 5 years and 4.7%/y at 10 years after implantation. The
annual risk of protrusion was 1.5%. A total of 40 spontaneous injuries
were reported to a worldwide registry (WWR) that included data from
34 672 patients (34 892 Accufix leads), including pericardial
tamponade (n=19), pericardial effusion (n=5), atrial perforation (n=3),
J retention wire embolization (n=4), and death (n=6). The risk of
injury was 0.02%/y (95% CI, 0.0025 to 0.072) in the MCS and 0.048%/y
(95% CI, 0.035 to 0.067) in the WWR. A total of 5299 leads (13%) have
been extracted worldwide. After recall in the WWR, fatal extraction
complications occurred in 0.4% of intravascular procedures (16 of
4023), with life-threatening complications in 0.5% (n=21). Extraction
complications increased with implant duration, female sex, and J
retention wire protrusion.
ConclusionsAccufix pacing leads pose a low, ongoing risk of injury. Extraction is associated with substantially higher risks, and a conservative management approach is indicated for most patients.
Key Words: pacemakers pacing complications
This article has been cited by other articles:
![]() |
M. S. Amin, D. B. Matchar, M. A. Wood, and K. A. Ellenbogen Management of recalled pacemakers and implantable cardioverter-defibrillators: a decision analysis model. JAMA, July 26, 2006; 296(4): 412 - 420. [Abstract] [Full Text] [PDF] |
||||
![]() |
Conservative Approach for Extraction of Accufix Leads Journal Watch Cardiology, January 21, 2000; 2000(121): 10 - 10. [Full Text] |
||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1999 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |