Donate Help Contact The AHA Sign In Home
American Heart Association
Circulation
Search: search_blue_button Advanced Search
Circulation. 1999;100:203-208

This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Strickberger, S. A.
Right arrow Articles by Cain, C.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Strickberger, S. A.
Right arrow Articles by Cain, C.
Related Collections
Right arrow Arrythmias-basic studies
Right arrow Ablation/ICD/surgery
Right arrow Coronary imaging: angiography/ultrasound/Doppler/CC

(Circulation. 1999;100:203-208.)
© 1999 American Heart Association, Inc.


Basic Science Reports

Extracardiac Ablation of the Canine Atrioventricular Junction by Use of High-Intensity Focused Ultrasound

S. Adam Strickberger, MD; Takashi Tokano, MD; Jan-Ulco A. Kluiwstra, MS; Fred Morady, MD; Charles Cain, PhD

From the Division of Cardiology (S.A.S., T.T., F.M.) and the Department of Biomedical Engineering (J.-U.A.K., C.C.), University of Michigan, Ann Arbor.

Correspondence to S. Adam Strickberger, MD, University of Michigan Medical Center, 1500 E Medical Center Dr, Box 0022, Ann Arbor, MI 48109.

Background—High-intensity focused ultrasound has been applied to internal organs from outside the body to ablate tissue. No published study has assessed the feasibility of ablating cardiac tissue within the beating heart by use of this type of therapeutic ultrasound. The purpose of this study was to determine whether high-intensity focused ultrasound can be used to ablate the atrioventricular (AV) junction within the beating heart.

Methods and Results—Ten dogs were anesthetized and underwent a thoracotomy. The heart was covered with a polyvinyl chloride membrane. The thorax above the membrane was perfused with degassed water, which functioned as a coupling medium for the ultrasound. A 7.0-MHz diagnostic ultrasound probe was affixed to a spherically focused 1.4-MHz high-intensity focused ultrasound transducer with a 1.1x8.3-mm focal zone 63.5 mm from the ablation transducer. The diagnostic ultrasound probe was calibrated such that the location of the focal zone of the ablation transducer was identifiable on the 2-dimensional ultrasound image. Target sites were identified with the diagnostic ultrasound. The maximum ultrasound intensity for ablation (2.8 kW/cm2) was delivered to the AV junction only during electrical diastole and for a total of 30 seconds. Complete AV block was achieved in each of the 10 dogs with 6.5±5.6 (range, 3 to 21) 30-second applications of therapeutic ultrasound. Gross inspection showed that the mean lesion volume was 124±143 mm3, with a depth of 6.7±3.6 mm, a length of 5.7±2.5 mm, and a width of 4.7±1.8 mm. Four hours after the dogs were killed, histopathological study demonstrated a well-demarcated area of necrosis and early inflammation.

Conclusions—High-intensity focused ultrasound produces well-demarcated lesions and appears to be a feasible energy source to create complete AV block within the beating heart without damaging the overlying or underlying cardiac tissue. This energy source may allow for a noninvasive approach to ablation of cardiac arrhythmias.


Key Words: ablation • atrioventricular node • arrhythmias




This article has been cited by other articles:


Home page
Card Surg AdultHome page
R. K. Voeller, R. B. Schuessler, and R. J. Damiano Jr.
Surgical Treatment of Atrial Fibrillation
Card. Surg. Adult, January 1, 2008; 3(2008): 1375 - 1394.
[Full Text]


Home page
J Ultrasound MedHome page
K. Fujikura, R. Otsuka, A. Kalisz, J. A. Ketterling, Z. Jin, R. R. Sciacca, C. C. Marboe, J. Wang, R. Muratore, E. J. Feleppa, et al.
Effects of ultrasonic exposure parameters on myocardial lesions induced by high-intensity focused ultrasound.
J. Ultrasound Med., November 1, 2006; 25(11): 1375 - 1386.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
J. E. Burnes, B. Taccardi, and Y. Rudy
A Noninvasive Imaging Modality for Cardiac Arrhythmias
Circulation, October 24, 2000; 102(17): 2152 - 2158.
[Abstract] [Full Text] [PDF]