© American Heart Association, Inc.
Basic Science Reports |
From the Cardiovascular Institute (F.G.S.G.), Mt View, Calif; the Division of Cardiovascular Surgery (J.I.F., T.J.F.), Stanford University Medical Center, Stanford, Calif; Evalve, Inc (J.K.), Redwood City, Calif; the Division of Cardiology (E.F.), University of CaliforniaSan Francisco Medical Center, San Francisco, Calif; the Department of Cardiothoracic Surgery (M.C.O.), Columbia University, New York, NY; the Division of Cardiology (T.F.), Evanston Northwestern Healthcare, Evanston, Ill; and the Division of Cardiology (P.C.B.), Emory University, Atlanta, Ga.
Correspondence to Frederick St. Goar, MD, Cardiovascular Institute, 2660 Grant Rd, Mt View, CA 94040. E-mail fstgoar{at}aol.com
Received July 30, 2003; revision received August 21, 2003; accepted August 22, 2003.
Background The edge-to-edge technique is an accepted method for the surgical repair of a regurgitant mitral valve. This study reports the initial use of an endovascular technology that enables a double-orifice edge-to-edge mitral valve repair without cardiopulmonary bypass in an animal model.
Methods and Results Adult pigs (n=14) were anesthetized, and left thoracotomy was performed for epicardial echo imaging. Using femoral vein access, a steerable guide catheter was placed transseptally into the left atrium. An implantable clip designed to grasp and approximate the middle scallops of the anterior and posterior mitral leaflets was introduced through the guide catheter. The clip was opened in the left atrium, advanced through the mitral orifice, and retracted to grasp the leaflet edges. When a functional double-orifice valve was confirmed by echo, the clip was closed to coapt the leaflets and detached from the delivery catheter. Before final clip detachment, echo demonstrated a double orifice in all 14 animals. In 2 studies, the clip released from the anterior mitral leaflet. Retrospective analysis of echo images indicated an incomplete grasp of the anterior leaflet. Immediate postmortem examination revealed that the clip successfully approximated the middle scallops of the anterior and posterior leaflets in all 12 double-orifice studies.
Conclusions This study demonstrates for the first time that an endovascular system can be successfully used to perform the edge-to-edge repair technique in a nondiseased porcine model. This technique is potentially applicable as a percutaneous catheterization laboratory procedure for the treatment of mitral regurgitation in humans.
Key Words: mitral valve regurgitation catheterization
This article has been cited by other articles:
![]() |
R D Christofferson, S R Kapadia, V Rajagopal, and E M Tuzcu Emerging transcatheter therapies for aortic and mitral disease Heart, January 1, 2009; 95(2): 148 - 155. [Full Text] [PDF] |
||||
![]() |
A. Plass, I. Valenta, O. Gaemperli, P. Kaufmann, H. Alkadhi, G. Zund, J. Grunenfelder, and M. Genoni Assessment of coronary sinus anatomy between normal and insufficient mitral valves by multi-slice computertomography for mitral annuloplasty device implantation Eur. J. Cardiothorac. Surg., April 1, 2008; 33(4): 583 - 589. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Davidson and D. S. Baim Percutaneous Catheter-Based Mitral Valve Repair Card. Surg. Adult, January 1, 2008; 3(2008): 1101 - 1108. [Full Text] |
||||
![]() |
T. Feldman and M. B. Leon Prospects for Percutaneous Valve Therapies Circulation, December 11, 2007; 116(24): 2866 - 2877. [Full Text] [PDF] |
||||
![]() |
T. A. Timek, S. L. Nielsen, D. T. Lai, D. Liang, G. T. Daughters, N. B. Ingels Jr, and D. C. Miller Effect of Chronotropy and Inotropy on Stitch Tension in the Edge-to-Edge Mitral Repair Circulation, September 11, 2007; 116(11_suppl): I-276 - I-281. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Coats and P. Bonhoeffer NEW PERCUTANEOUS TREATMENTS FOR VALVE DISEASE Heart, May 1, 2007; 93(5): 639 - 644. [Full Text] [PDF] |
||||
![]() |
J. H. Rogers, J. A. Macoviak, D. A. Rahdert, P. A. Takeda, I. F. Palacios, and R. I. Low Percutaneous Septal Sinus Shortening: A Novel Procedure for the Treatment of Functional Mitral Regurgitation Circulation, May 16, 2006; 113(19): 2329 - 2334. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. K. Mishra, S. Mittal, P. Jaguri, and N. Trehan Coapsys Mitral Annuloplasty for Chronic Functional Ischemic Mitral Regurgitation: 1-Year Results Ann. Thorac. Surg., January 1, 2006; 81(1): 42 - 46. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Feldman, H. S. Wasserman, H. C. Herrmann, W. Gray, P. C. Block, P. Whitlow, F. St. Goar, L. Rodriguez, F. Silvestry, A. Schwartz, et al. Percutaneous Mitral Valve Repair Using the Edge-to-Edge Technique: Six-Month Results of the EVEREST Phase I Clinical Trial J. Am. Coll. Cardiol., December 6, 2005; 46(11): 2134 - 2140. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. C. Dang, M. S. Aboodi, T. Sakaguchi, H. S. Wasserman, M. Argenziano, D. M. Cosgrove, T. K. Rosengart, T. Feldman, P. C. Block, and M. C. Oz Surgical Revision After Percutaneous Mitral Valve Repair With a Clip: Initial Multicenter Experience Ann. Thorac. Surg., December 1, 2005; 80(6): 2338 - 2342. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Boudjemline, G. Agnoletti, D. Bonnet, L. Behr, N. Borenstein, D. Sidi, and P. Bonhoeffer Steps Toward the Percutaneous Replacement of Atrioventricular Valves: An Experimental Study J. Am. Coll. Cardiol., July 19, 2005; 46(2): 360 - 365. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. C. Block Percutaneous Mitral Valve Repair: Are They Changing the Guard? Circulation, May 3, 2005; 111(17): 2154 - 2156. [Full Text] [PDF] |
||||
![]() |
J. E. Sousa, M. A. Costa, E. M. Tuzcu, J. S. Yadav, and S. Ellis New Frontiers in Interventional Cardiology Circulation, February 8, 2005; 111(5): 671 - 681. [Full Text] [PDF] |
||||
![]() |
S. H. Rahimtoola The year in valvular heart disease J. Am. Coll. Cardiol., January 4, 2005; 45(1): 111 - 122. [Full Text] [PDF] |
||||
![]() |
G. Lutter, R. Ardehali, J. Cremer, and P. Bonhoeffer Percutaneous Valve Replacement: Current State and Future Prospects Ann. Thorac. Surg., December 1, 2004; 78(6): 2199 - 2206. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. V. Maniu, J. B. Patel, D. G. Reuter, D. M. Meyer, W. D. Edwards, C. S. Rihal, and M. M. Redfield Acute and chronic reduction of functional mitral regurgitation in experimental heart failure by percutaneous mitral annuloplasty J. Am. Coll. Cardiol., October 19, 2004; 44(8): 1652 - 1661. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Fukamachi, Z. B. Popovic, M. Inoue, K. Doi, S. Schenk, Y. Ootaki, M. W. Kopcak Jr., and P. M. McCarthy Changes in mitral annular and left ventricular dimensions and left ventricular pressure-volume relations after off-pump treatment of mitral regurgitation with the Coapsys device Eur. J. Cardiothorac. Surg., March 1, 2004; 25(3): 352 - 357. [Abstract] [Full Text] [PDF] |
||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2003 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |